Civco eTRAX Needle Sensor 12G recalled for incorrect needle tip positioning
Civco Medical Instruments Co. Inc. recalled the eTRAX Needle Sensor 12G (Part Number 667-156) used with Aurora Trackers due to an inspection and programming error. The sensor may incorrectly display the needle tip position on the user interface, which could affect clinical accuracy during procedures.
From the FDA notice · Read the official notice ↗
What to do
- Stop using eTRAX Needle Sensor 12G (Part Number 667-156, Lot A276634) immediately
- Contact Civco Medical Instruments for instructions on returning or replacing the sensor
- Verify your device's lot number matches A276634 before resuming use
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- eTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-156
- Sold in
- US States: OH, PA, WA. China.
- Made for
- Adults (under 100 years)
This recall covers 8 products — see all of them.
Details
- Units affected
- 5 units
- Illnesses reported
- None reported
- Manufactured in
- 102 1st St S, Kalona, IA, United States
- Classification
- Class II
- Recall number
- Z-1735-2026
In the FDA's words
- Description
- eTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-156 There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface. UDI-DI: 00841436111072; Lot Number: A276634 Distribution: US States: OH, PA, WA. China. Affected quantity: 5 units
- Hazard
- There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
8 products in this recall
Civco eTRAX needle systems recalled for tip position error
Civco is recalling 8 eTRAX needle system starter kits and needle sensors used with Aurora trackers. An error in inspection and programming may cause the needle tip position to be shown incorrectly on the user interface. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- eTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149Z-1731-2026 · 2 units
- eTRAX Needle System Starter Kit 14G (for Aurora Trackers), Part Number 667-150Z-1732-2026 · 6 units
- eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151Z-1733-2026 · 18 units
- eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152Z-1734-2026 · 20 units
- eTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-156 (this page)Z-1735-2026 · 5 units
- eTRAX Needle Sensor - 14G(for Aurora Trackers), Part Number 667-157Z-1736-2026 · 5 units
- eTRAX Needle Sensor - 16G(for Aurora Trackers), Part Number 667-158Z-1737-2026 · 10 units
- eTRAX Needle Sensor - 18G (for Aurora Trackers), Part Number 667-159Z-1738-2026 · 21 units