FDA-Device2026-10-07Class IIPROCESSING DEFECT

Philips Spectral CT 7500 software recalled for excessive radiation risk from bolus tracking malfunction

OTHERNationwide distributionMedical devices

What this means

The bolus tracking application on Philips Spectral CT 7500 scanners (Software V5.2.x) can malfunction and continue taking imaging shots even after reaching the contrast threshold. This can cause patients to receive unintended additional radiation during CT scans.

  • Contact Philips Medical Systems for a software update or patch.
  • Do not use the bolus tracking feature until the issue is resolved.
  • Review recent imaging records for potentially affected scans and consult with your radiologist if needed.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.

Intended for
Adults
Illnesses reported
None reported
Sold states
U.S. and U.S. Territories.
Affected count
62
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Products
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3155-2026

More Philips recalls

Want us to check the next one for you?

We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Free · One email a week · Unsubscribe anytime
By signing up you agree to our Terms and Privacy Policy.