Moderate riskFDA DeviceClass II

Philips Achieva 1.5T MRI systems recalled for software issue causing exam termination

A software problem in certain Philips Achieva 1.5T MRI machines can cause a scan to stop unexpectedly during image processing if technicians manually adjust scan settings in a way that requires more memory than the system has available. This affects specific serial numbers of five model variants distributed worldwide.

From the FDA notice · Read the official notice ↗ · All Philips recalls

What to do

  1. Check if your facility operates an affected Achieva 1.5T model by serial number (available in the full recall notice)
  2. Contact Philips for a software update or workaround instructions
  3. Review scan protocols to avoid parameter combinations that exceed available memory until the issue is resolved

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Achieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (Achieva 1.5T), 2. 781343 (Achieva 1.5T), 3. 781283 (Achieva 1.5T Conversion), 4. 781346 (Achieva 1.5 Conversion), 5. 781178 (Achieva 1.5T Initial System);
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 12 products — see all of them.

Details

Units affected
387
Illnesses reported
None reported
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Classification
Class II
Recall number
Z-3232-2026

In the FDA's words

Description
Achieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (Achieva 1.5T), 2. 781343 (Achieva 1.5T), 3. 781283 (Achieva 1.5T Conversion), 4. 781346 (Achieva 1.5 Conversion), 5. 781178 (Achieva 1.5T Initial System); A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. 1. Model Numbers (REF): 781196 (Achieva 1.5T); UDI-DI: N/A; Serial Numbers: 22135, 21807, 22149, 22162, 22186, 22264, 21329, 21787, 21989, 22022, 22150, 22180, 22270, 21674, 22080, 21190, 21254, 21315, 21530, 21592, 21903, 21913, 21914, 22072, 22076, 22161, 22176, 22217, 22278, 21796, 21996, 21306, 21499, 21525, 21556, 21626, 21897, 21985, 22213, 22056, 21342, 21571, 21594, 21810, 21924, 22335, 21235, 21492, 22198, 21243, 21449, 22029, 22069, 21731, 22095, 22185, 22239, 22108, 22079, 21984, 21073, 21211, 21269, 21487, 21488, 21550, 21598, 21624, 21808, 22071, 22223, 21911, 21225, 21954, 21215, 21397, 21431, 21689, 21986, 22021, 22243, 21427, 22117, 22326, 21076, 21104, 21183, 21264, 21272, 21284, 21321, 21334, 21351, 21405, 21455, 21479, 21494, 21508, 21572, 21574, 21589, 21666, 21776, 21805, 21868, 21879, 21935, 21957, 21968, 21993, 22001, 22004, 22066, 22130, 22136, 22155, 22207, 22258; 2. Model Numbers (REF): 781343 (Achieva 1.5T); UDI-DI: 884838055292; Serial Numbers: 72020, 72104, 72137, 72138, 72146, 72157, 72160, 72170, 72171, 72179, 72183, 72185, 72200, 72223, 72233, 72238, 72251, 72269, 72013, 72111, 72113, 72231, 72235, 72270, 72271, 72273, 72286, 72079, 72083, 72109, 72120, 72122, 72123, 72124, 72126, 72139, 72141, 72142, 72147, 72150, 72247, 72007, 72156, 72101, 72116, 72117, 72180, 72197, 72267, 72189, 72008, 72080, 72036, 72052, 72057, 72103, 72134, 72143, 72144, 72145, 72151, 72186, 72199, 72217, 72224, 72227, 72228, 72236, 72253, 72254, 72255, 72256, 72257, 72258, 72259, 72260, 72264, 72278, 72279, 72296, 72297, 72298, 72299, 72304, 72029, 72129, 72153, 72175, 72194, 72207, 72281, 72290, 72301, 72302, 72310, 72311, 72316, 72320, 72321, 72326, 72329, 72075, 72076, 72261, 72022, 72047, 72100, 72154, 72159, 72240, 72285, 72287, 72317, 72324, 72015, 72016, 72023, 72071, 72074, 72205, 72038, 72322, 72053, 72063, 72077, 72118, 72121, 72177, 72188, 72192, 72202, 72203, 72265, 72274, 72288, 72289, 72308, 72313, 72048, 72001, 72058, 72230, 72234, 72102, 72201, 72003, 72040, 72114, 72010, 72049, 72055, 72136, 72148, 72266, 72319, 72222, 72034, 72198, 72220, 72226, 72119, 72168, 72303, 72309, 72312, 72318, 72025, 72066, 72193, 72325, 72327, 72331, 72174, 72082, 72249; 3. Model Numbers (REF): 781283 (Achieva 1.5T Conversion); UDI-DI: N/A; Serial Numbers: 39056, 8989, 10126, 39098, 39108, 39004, 39010, 39028, 39032, 39086, 39094, 39095, 39058, 39091, 39105, 39053, 39073, 39067, 39016, 39025; 4. Model Numbers (REF): 781346 (Achieva 1.5 Conversion); UDI-DI: N/A; Serial Numbers: 18052, 8361, 8415, 10015, 75008, 5243, 8727, 10172, 10189, 10545, 18737, 75002, 75003, 75004, 75015, 8059, 8757, 8767, 8842, 8883, 8886, 8897, 8946, 8974, 8981, 8990, 10060, 18001, 18710, 75009, 75092, 75095, 75099, 75100, 8069, 8312, 8520, 10006, 10048, 18588, 75000, 75001, 75005, 75031, 75049, 75087, 8371, 8289, 8367, 8740, 10539, 75011, 8351, 8782, 8756, 75010, 75007, 8124, 8020; 5. Model Numbers (REF): 781178 (Achieva 1.5T Initial System); UDI-DI: N/A; Serial Numbers: 20095, 20056, 20230, 20258, 20085, 20094, 20113, 20050, 20072, 20088, 20096, 20252, 20281, 20297, 20307; Distribution: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbeki
Hazard
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

12 products in this recall

Philips MRI systems recalled for software issue ending exams

Philips is recalling 12 groups of MRI systems and related upgrade products because of a software issue. If scan settings are manually adjusted so the image processing needs more memory than the system has, the examination can end partway through. No injuries have been reported.