Moderate riskFDA DeviceClass II

Medtronic MiniMed 780G insulin pumps recalled for delivery stoppage

The FDA lists MiniMed 780G insulin pumps with software version 6.62 for a software anomaly that can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump can unexpectedly stop delivering insulin and show Pump Error 53 followed immediately by Pump Error 23. The notice says this may lead to delayed therapy and an increased risk of high blood sugar or diabetic ketoacidosis (DKA).

From the FDA notice · Read the official notice ↗ · All Medtronic recalls

What to do

  1. Do not stop using the pump on your own. Contact your diabetes care team and Medtronic MiniMed about this software issue.
  2. Check whether your pump is a listed model (REF MMT-1885, MMT-1886, MMT-1895 or MMT-1896) and runs software version 6.62, using the UDI/GTIN numbers in the notice.
  3. If the pump shows Pump Error 53 followed by Pump Error 23, contact your care team right away.
  4. The notice doesn't say how to get a software fix or replacement; ask Medtronic MiniMed or your care team how to get one.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
Sold in
Nationwide
Made for
All ages (under 100 years)

Details

Units affected
778
Illnesses reported
None reported
Manufactured in
18000 Devonshire St, N/A, Northridge, CA, United States
Classification
Class II
Recall number
Z-2698-2026

In the FDA's words

Description
MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896 A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA). UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62 Distribution: International distribution to the countries of Spain, Great Britain, Netherlands, Italy. Affected quantity: 778
Hazard
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).