Medtronic MiniMed 780G insulin pumps recalled for delivery stoppage
The FDA lists MiniMed 780G insulin pumps with software version 6.62 for a software anomaly that can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump can unexpectedly stop delivering insulin and show Pump Error 53 followed immediately by Pump Error 23. The notice says this may lead to delayed therapy and an increased risk of high blood sugar or diabetic ketoacidosis (DKA).
From the FDA notice · Read the official notice ↗ · All Medtronic recalls
What to do
- Do not stop using the pump on your own. Contact your diabetes care team and Medtronic MiniMed about this software issue.
- Check whether your pump is a listed model (REF MMT-1885, MMT-1886, MMT-1895 or MMT-1896) and runs software version 6.62, using the UDI/GTIN numbers in the notice.
- If the pump shows Pump Error 53 followed by Pump Error 23, contact your care team right away.
- The notice doesn't say how to get a software fix or replacement; ask Medtronic MiniMed or your care team how to get one.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
- Sold in
- Nationwide
- Made for
- All ages (under 100 years)
Details
- Units affected
- 778
- Illnesses reported
- None reported
- Manufactured in
- 18000 Devonshire St, N/A, Northridge, CA, United States
- Classification
- Class II
- Recall number
- Z-2698-2026
In the FDA's words
- Description
- MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896 A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA). UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62 Distribution: International distribution to the countries of Spain, Great Britain, Netherlands, Italy. Affected quantity: 778
- Hazard
- A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
More from Medtronic
See all- Medtronic MiniMed Go app recalled for failed long-acting dose alerts
ModerateFDAOct 22, 2026
- Medtronic MiniMed CareLink Clinic software recalled for insulin dosing calculation error
SeriousFDAOct 21, 2026
- Medtronic MiniMed Flex insulin pumps recalled for possible malfunction
SeriousFDAOct 14, 2026
- MiniMed 640G loaner pumps recalled for possible cleaning damage
ModerateFDAAug 26, 2026