FDA-Device2026-10-21Class IIPROCESSING DEFECT

Medtronic MiniMed CareLink Clinic software recalled for insulin dosing calculation error

OTHERNationwide distributionMedical devices

What this means

Medtronic MiniMed CareLink Clinic software version 4.4A has a defect in its Settings Calculator feature. When healthcare providers enter custom insulin values, the calculator ignores them and continues using default values instead. This can result in incorrect insulin dose recommendations being sent to patients' insulin pumps, potentially causing patients to receive too much or too little insulin.

  • If you use CareLink Clinic software version 4.4A, contact Medtronic or your healthcare provider immediately
  • Do not rely on insulin calculations from the Settings Calculator in version 4.4A until you receive an update
  • Monitor your blood glucose closely and follow your healthcare provider's dosing instructions
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).

Intended for
Adults
Illnesses reported
None reported
Sold states
US Nationwide distribution in the state of KS.
Affected count
1
Manufactured in
18000 Devonshire St, Northridge, California, United States
Products
CareLink Clinic, REF: MMT-7350

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0203-2027