MiniMed 640G loaner pumps recalled for possible cleaning damage

Certain loaner MiniMed 640G insulin pumps (models MMT-1711 and MMT-1712) were cleaned with products that can damage the pump case or keypad. This damage may allow moisture to enter the pump and cause it to deliver too much or too little insulin, potentially leading to dangerously high or low blood sugar levels.

From the FDA notice · Read the official notice ↗ · All Medtronic recalls

What to do

  1. If you received a loaner MiniMed 640G pump, check the model number (MMT-1711 or MMT-1712) and serial number against the FDA recall list.
  2. Contact Medtronic MiniMed immediately if your pump's serial number matches the recalled list.
  3. Do not use a recalled loaner pump; request a replacement from Medtronic or your healthcare provider.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 3 products — see all of them.

Details

Units affected
16
Illnesses reported
None reported
Manufactured in
18000 Devonshire St, Northridge, CA, United States
Classification
Class II
Recall number
Z-2980-2026

In the FDA's words

Description
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712; Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis. Model/UDI-DI/Serial Numbers: MMT-1711/00763000404550/NG2660048H, NG2663343H, NG3015306H; MMT-1712/00763000404567/NG2628486H, NG2758213H, NG2999651H, NG3349198H, NG3351719H, NG2678842H, NG2831821H, NG2831546H, NG2932027H, NG3049657H, NG3033094H, NG2831538H, NG3738739H Distribution: International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland. Affected quantity: 16
Hazard
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.

3 products in this recall

Medtronic MiniMed insulin pumps recalled for cleaning damage risk

Medtronic is recalling three MiniMed insulin pump models (640G, 670G, and 780G) that were loaned out and may have been cleaned with products that are incompatible with the pump. This could damage the case or keypad, let moisture in, and cause the pump to deliver too much or too little insulin. No injuries have been reported. The recall covers 280 affected pumps and was distributed in several European countries.

The agency lists each product separately. Check every one you might have.