Argon Medical L-Cath PICC catheters recalled for possible hub leaks
Argon Medical Devices is recalling certain L-Cath PICC catheters because they may leak or crack at the hub. The agency says this may lead to catheter replacement, inability to flush or aspirate, blockage, therapy interruption, blood loss, tissue irritation, or bloodstream infection. No injuries have been reported.
From the FDA notice · Read the official notice ↗ · All Argon Medical Devices recalls
What to do
- Set aside and stop using affected lots of these catheters.
- Check REF numbers 384539, 384464, 384516 and 384463 against the lot numbers and expiration dates listed in the notice.
- The notice doesn't say how to get a replacement or refund; contact Argon Medical Devices or your supplier.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm, REF: 384539; 18ga (3.5F) 1.15mm x 60cm, REF: 384464; Basic Kit 28GA (1.2F) x 25cm 1 Lumen), REF: 384516; 20ga (3F) 1.00mm x 60cm, REF: 384463
- Sold in
- US: AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OR, TN, TX, UT, VA, WA. OUS: MYS, CAN, SAU
- Made for
- All ages (under 100 years)
This recall covers 2 products — see all of them.
Details
- Units affected
- 6928
- Illnesses reported
- None reported
- Manufactured in
- 1445 Flat Creek Rd, N/A, Athens, TX, United States
- Classification
- Class II
- Recall number
- Z-2803-2026
In the FDA's words
- Description
- L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm, REF: 384539; 18ga (3.5F) 1.15mm x 60cm, REF: 384464; Basic Kit 28GA (1.2F) x 25cm 1 Lumen), REF: 384516; 20ga (3F) 1.00mm x 60cm, REF: 384463 Catheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection. REF/UDI-DI/Lot(Expiration): 384539/00886333209934/11488167(7/14/2026), 11488168(7/17/2026), 11494535E(8/1/2026), 11507766(11/13/2026), 11550780(12/12/2026), 11551487(12/15/2026), 11551488(12/15/2026); 384464/00886333209835/11490593(7/28/2026), 11494886(8/28/2026)11501150(10/6/2026); 384516/00886333209910/11500410(8/14/2026), 11501856(8/14/2026), 11504222(9/9/2026), 11506952(9/9/2026), 11523868(9/18/2026), 11551789(9/18/2026), 11553909(11/13/2026), 11559226(11/20/2026); 384463/00886333209828/11490797(8/1/2026), 11490959(8/11/2026) Distribution: US: AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OR, TN, TX, UT, VA, WA. OUS: MYS, CAN, SAU Affected quantity: 6928
- Hazard
- Catheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection.
2 products in this recall
Argon L-Cath PICC and midline catheters recalled for possible leaks
Argon Medical Devices is recalling 2 L-Cath catheter products, the L-Cath PICC S/L Peripherally Inserted Central Catheter and the L-Cath Midline S/L Catheter, with 6,948 affected units in total. The catheters may leak or crack at the hub, which may lead to replacement, inability to flush or aspirate the catheter, occlusion, therapy interruption, blood loss, tissue irritation, or bloodstream infection. The notice reports no injuries.
The agency lists each product separately. Check every one you might have.
- L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm, REF: 384539; 18ga (3.5F) 1.15mm x 60cm, REF: 384464; Basic Kit 28GA (1.2F)… (this page)Z-2803-2026 · 6928
- L-Cath Midline S/L Catheter 20ga (3F) 1mm x 20cm, REF: 384620Z-2804-2026 · 20