Moderate riskFDA DeviceClass II

GE Healthcare MAC VU360 ECG recalled for patient data mix-up risk

GE Healthcare recalled certain MAC VU360 electrocardiograph units running software v3.0. After the order refresh icon is selected, a preview or review screen may not close automatically after 2 minutes of inactivity, so data from a previous patient could stay on screen and be linked to the next patient. No injuries have been reported.

From the FDA notice · Read the official notice ↗ · All GE recalls

What to do

  1. Check whether your MAC VU360 serial number appears on the recall list.
  2. Contact GE Healthcare for guidance on the software issue, since this notice lists no specific remedy.
  3. Manually close any open preview or review screen before using the device with a new patient.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
163 units
Illnesses reported
None reported
Manufactured in
9900 W Innovation Dr, Milwaukee, WI, United States
Classification
Class II
Recall number
Z-2876-2026

In the FDA's words

Description
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient. GTIN 00198953107263, Serial Numbers: VU221150078SA, VU222400121SA, VU223230048SA, VU222520132SA, VU219400033SA, VU219400034SA, VU219400045SA, VU219400055SA, VU220270027SA, VU220510048SA, VU220510078SA, VU221100161SA, VU221100181SA, VU221150116SA, VU222120212SA, VU222120220SA, VU222120239SA, VU222120243SA, VU222120249SA, VU222120250SA, VU222120253SA, VU222120258SA, VU222120260SA, VU222120261SA, VU222120266SA, VU222120269SA, VU222140004SA, VU222200296SA, VU222520114SA, VU223120014SA, VU223120041SA, VU223420119SA, VU224050158SA, VU224070091SA, VU224200028SA, VU224420159SA, VU224520140SA, VU225170260SA, VU225420050SA, VU225520037SA, VU226040025SA, VU419320042SA, VU421490135SA, VU421490151SA, VU421490153SA, VU421490155SA, VU421490159SA, VU422460150SA, VU422460194SA, VU422460203SA, VU422460303SA, VU422480164SA, VU422500239SA, VU422500240SA, VU422500241SA, VU422500242SA, VU422500244SA, VU422500245SA, VU422500246SA, VU422500247SA, VU422500248SA, VU422500249SA, VU422500250SA, VU422500251SA, VU422500254SA, VU422500260SA, VU422500261SA, VU422500264SA, VU422500265SA, VU422500266SA, VU422500267SA, VU422500269SA, VU422510038SA, VU422510049SA, VU422510060SA, VU422510061SA, VU423050080SA, VU423050178SA, VU423220064SA, VU423220225SA, VU425130020SA, VU425130022SA, VU425130024SA, VU425130084SA, VU425130085SA, VU421490060SA, VU421490063SA, VU421490065SA, VU421490068SA, VU421490069SA, VU421490097SA, VU421490122SA, VU422460206SA, VU422460209SA, VU422460215SA, VU422460218SA, VU422460219SA, VU422460302SA, VU422460304SA, VU422460269SA, VU422460273SA, VU422460274SA, VU422460278SA, VU422460282SA, VU422460284SA, VU422480071SA, VU422460306SA, VU422460285SA, VU422480070SA, VU422460287SA, VU422460301SA, VU422480066SA, VU422480078SA, VU422480083SA, VU422480084SA, VU422480085SA, VU422480086SA, VU422480087SA, VU422480088SA, VU422480099SA, VU422480110SA, VU422480111SA, VU422480120SA, VU422480121SA, VU422480125SA, VU422480127SA, VU422480128SA, VU422480129SA, VU425130087SA, VU224200105SA, VU222120088SA, VU223130071SA, VU223220135SA, VU223220174SA, VU219490019SA, VU226050047SA, VU225460222SA, VU225460237SA, VU225470012SA, VU225470018SA, VU225470022SA, VU225470023SA, VU225470024SA, VU225470030SA, VU225470034SA, VU225470035SA, VU225470038SA, VU221050014SA, VU223310039SA, VU223450002SA, VU223310012SA, VU222190155SA, VU225060039SA, VU225060040SA, VU225060046SA, VU225060038SA, VU226154224SA, VU225500096SA, VU225500106SA, VU225500109SA, VU225500019SA, VU225500031SA, VU225500076SA. Distribution: Worldwide distribution - US Nationwide and the countries of Belgium, Hong Kong, Netherlands, Sweden, UK. Affected quantity: 163 units
Hazard
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.