Medline Hudson RCI neonatal breathing circuit recalled for heated wire connector failure risk
The heated wire connector in Medline Hudson RCI neonatal and infant breathing circuits may fail unpredictably, causing discoloration, melting, smoke, or burning. If the circuit fails, infants may experience airway obstruction, difficulty breathing, or fluid pooling in the airway.
From the FDA notice · Read the official notice ↗ · All Medline recalls
What to do
- If you use this breathing circuit (Medline HUD78019 or Teleflex 780-19), stop using it immediately.
- Contact Medline Industries or your equipment supplier for replacement or instructions.
- Inspect the heated wire connector for discoloration, melting, smoke residue, or burning odor before each use.
- Monitor patients closely for signs of respiratory distress or unusual odors during device operation.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Respiratory gas humidifier.
- Sold in
- Nationwide
- Made for
- Infants (under 12 months)
This recall covers 3 products — see all of them.
Details
- Units affected
- 20 units
- Illnesses reported
- None reported
- Manufactured in
- 3 Lakes Dr, N/A, Northfield, IL, United States
- Classification
- Class II
- Recall number
- Z-2991-2026
In the FDA's words
- Description
- Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Respiratory gas humidifier. Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress. All lot codes Distribution: Worldwide distribution - US Nationwide and the country of Panama. Affected quantity: 20 units
- Hazard
- Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
3 products in this recall
Medline Hudson RCI neonatal breathing circuits recalled for heat damage risk
Medline is recalling three neonatal and infant heated wire breathing circuit products, about 417,000 units in total. The heated wire connector at the humidifier can fail and show discoloration, melting, smoke, or a burning odor. Failure could cause breathing problems in infants, and no injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Re… (this page)Z-2991-2026 · 20 units
- Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-0…Z-2992-2026 · 145525 units
- Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Produc…Z-2993-2026 · 271406 units
More from Medline
See all- Medline surgical procedure kits recalled for reprocessed components
ModerateFDANov 20, 2026
- Medline Bubble CPAP kit recalled for possible heat damage risk
ModerateFDANov 20, 2026
- Medline catheter kits recalled for possible lubricant contamination
ModerateFDANov 10, 2026