FDA-Device2026-09-02Class IIPROCESSING DEFECT

Medline Hudson RCI neonatal heated breathing circuits recalled for connector failure risk

BURNFIREOTHERNationwide distributionMedical devicesBaby & kids

What this means

Medline's heated wire breathing circuits for newborns and infants may fail unexpectedly. The heated connector at the humidifier can discolor, melt, or show signs of heat damage, which could cause the circuit to stop working or allow fluid to enter the airway.

  • Check your neonatal breathing circuit for the product numbers listed: HUD78007, HUD78009, 780-13, 780-14, 780-15, or HUD78018
  • If you have one of these circuits, stop using it and contact Medline or your equipment supplier
  • Contact your hospital or clinic's biomedical or respiratory equipment team to report and replace affected units
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

Intended for
Infants (under 12 months)
Illnesses reported
None reported
Sold states
Worldwide distribution - US Nationwide and the country of Panama.
Affected count
145525 units
Manufactured in
3 Lakes Dr, N/A, Northfield, IL, United States
Products
Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD78018) (Teleflex Product Number/SKU 780-18).

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2992-2026