Serious riskFDA DeviceClass II

Medline Hudson RCI heated breathing circuits recalled for connector failure risk in neonates and infants

Medline's Hudson RCI heated breathing circuits for newborns and infants may experience sporadic failure in the heated wire connector at the humidifier. The connector can show discoloration, melting, smoke, or burning odor. This failure could reduce oxygen delivery and increase risk of airway problems in infants using the device.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

  1. Check your device against the product numbers listed (HUD78001KIT, HUD78007KIT, HUD78008KIT, HUD78009KIT, HUD78017KIT, HUD78018KIT, HUD88001KIT, HUD88013KIT, HUD88014KIT, HUD88015KIT, or HUD88017KIT).
  2. Inspect the heated wire connector (pigtail) for discoloration, melting, smoke, or burning odor before each use.
  3. Contact Medline or your equipment provider for a replacement or instructions.
  4. Report any observed connector damage or device malfunction to Medline and the FDA.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT); 3) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT); 4) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT); 5) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT); 6) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL HTD, DUAL LIMB,COLUMN, Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT); 7) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD SINGLE LIMB,INFANT, Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT); 8) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD,INFANT,HTD,COLUMN,DRN, Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT); 9) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT); 10) MEDLINE HUDSON RCI,ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT); 11) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT).
Sold in
Nationwide
Made for
Infants (under 12 months)

This recall covers 3 products — see all of them.

Details

Units affected
271406 units
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, N/A, Northfield, IL, United States
Classification
Class II
Recall number
Z-2993-2026

In the FDA's words

Description
Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT); 3) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT); 4) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT); 5) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT); 6) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL HTD, DUAL LIMB,COLUMN, Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT); 7) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD SINGLE LIMB,INFANT, Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT); 8) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD,INFANT,HTD,COLUMN,DRN, Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT); 9) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT); 10) MEDLINE HUDSON RCI,ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT); 11) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT). Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress. 1) Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT), UDI/DI 10197344010752 (each) 20197344010759 (case), All lot numbers; 2) Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT), UDI/DI 10197344010738 (each) 20197344010735 (case), All lot numbers; 3) Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT), UDI/DI 10197344010721 (each) 20197344010728 (case), All lot numbers; 4) Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT), UDI/DI 10197344010714 (each) 20197344010711 (case), All lot numbers; 5) Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT), UDI/DI 10197344010677 (each) 20197344010674 (case), All lot numbers; 6) Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT), UDI/DI 10197344010905 (each) 20197344010902 (case), All lot numbers; 7) Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT), UDI/DI 10197344010929 (each) 20197344010926 (case), All lot numbers; 8) Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT), UDI/DI 10197344010493 (each) 20197344010490 (case), All lot numbers; 9) Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT), UDI/DI 10197344010486 (each) 20197344010483 (case), All lot numbers; 10) Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT), UDI/DI 10197344010479 (each) 20197344010476 (case), All lot numbers; 11) Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT), UDI/DI 10197344010462 (each) 20197344010469 (case), All lot numbers; Distribution: Worldwide distribution - US Nationwide and the country of Panama. Affected quantity: 271406 units
Hazard
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

3 products in this recall

Medline Hudson RCI neonatal breathing circuits recalled for heat damage risk

Medline is recalling three neonatal and infant heated wire breathing circuit products, about 417,000 units in total. The heated wire connector at the humidifier can fail and show discoloration, melting, smoke, or a burning odor. Failure could cause breathing problems in infants, and no injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from Medline

See all