Philips Ingenia Ambition MRI systems recalled for software crash during image processing
What this means
Philips Ingenia Ambition S and X MRI systems contain a software issue that can cause an examination to terminate unexpectedly during image reconstruction if scan parameters are manually adjusted in certain ways. This interruption occurs when the reconstruction requirements exceed the system's available memory.
- Contact Philips or your service representative for a software update or patch
- Review scan parameter settings to avoid exceeding system memory limits during reconstruction
- Report any unexpected examination terminations to Philips technical support
The agency's original wording
Hazard
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia;
Affected count
922
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Products
Ingenia Ambition S and Ingenia Ambition X; Model Numbers (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 781356 (X), 5. 782109 (X), 6. 782138 (X);
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3237-2026More Philips recalls
- FDA-Device2026-09-30Philips Ingenia 3.0T MRI systems recalled for software image reconstruction failure
- FDA-Device2026-09-30Philips SmartPath MRI software recalled for exam termination risk
- FDA-Device2026-09-30Philips Achieva 1.5T MRI systems recalled for software issue causing exam termination
- FDA-Device2026-09-30Philips Intera 1.5T MRI systems recalled for software examination termination risk
- FDA-Device2026-09-30Philips Evolution upgrade MRI systems recalled for software crash risk during image reconstruction
Want us to check the next one for you?
We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.