Philips Evolution upgrade MRI systems recalled for software crash risk during image reconstruction

Certain Philips Evolution upgrade 1.5T and 3.0T MRI systems contain a software issue that can cause an examination to terminate unexpectedly during image reconstruction. This happens when scan parameters are manually adjusted in a way that exceeds the system's available memory.

From the FDA notice · Read the official notice ↗ · All Philips recalls

What to do

  1. Check if your facility operates an affected Philips Evolution upgrade MRI system using the model and serial numbers listed in the recall notice.
  2. Contact Philips for a software update or corrective instructions.
  3. Do not manually adjust scan parameters beyond system memory capacity until the issue is resolved.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Evolution upgrade 1.5T and Evolution upgrade 3.0T; Model Numbers (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782117 (3.0T), 4. 782143 (3.0T);
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 12 products — see all of them.

Details

Units affected
91
Illnesses reported
None reported
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Classification
Class II
Recall number
Z-3235-2026

In the FDA's words

Description
Evolution upgrade 1.5T and Evolution upgrade 3.0T; Model Numbers (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782117 (3.0T), 4. 782143 (3.0T); A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. 1. Model Numbers (REF): 782116 (1.5T); UDI-DI: 884838099722; Serial Numbers: 41130, 70284, 41491, 41196, 41520, 70068, 41024, 41025, 41051, 41139, 41144, 41286, 41327, 41349, 41424, 41432, 41433, 70099, 70202, 81073, 41344, 41229, 41457, 41497, 41413, 41080, 41527, 70205, 70589, 81021, 41179, 41189, 70949; 2. Model Numbers (REF): 782148 (1.5T); UDI-DI: 884838108714; Serial Numbers: 41523, 70630, 70966, 84102, 84473, 84825, 41031, 41043, 41085, 41284, 41392, 41553, 70596, 81006, 70644, 41488, 41534, 70454, 41167, 41375, 41446, 70652, 41011, 70590, 70195, 84628; 3. Model Numbers (REF): 782117 (3.0T); UDI-DI: 884838099739; Serial Numbers: 42246, 85713, 42029, 71774, 85668, 85680, 85685, 85694, 42023, 42024, 42078, 42217, 42402, 71348, 71044, 85698, 85146, 42003, 85663, 85677; 4. Model Numbers (REF): 782143 (3.0T); UDI-DI: N/A; Serial Numbers: 71815, 42440, 71389, 85688, 85690, 85649, 42193, 42366, 85683, 85697, 85701, 85708; Distribution: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia; Affected quantity: 91
Hazard
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

12 products in this recall

Philips MRI systems recalled for software issue ending exams

Philips is recalling 12 groups of MRI systems and related upgrade products because of a software issue. If scan settings are manually adjusted so the image processing needs more memory than the system has, the examination can end partway through. No injuries have been reported.

The agency lists each product separately. Check every one you might have.