Moderate riskFDA DeviceClass II

Olympus BiCOAG Hemostasis Probe recalled for ceramic tip detachment risk

Olympus identified that certain BiCOAG Hemostasis Probe units (Model CD-B620LA) may have been manufactured with insufficient adhesive, creating a risk that the ceramic tip could detach during use. This affects all unexpired lots of this model worldwide.

From the FDA notice · Read the official notice ↗ · All Olympus recalls

What to do

  1. If you use or have this device in your facility, contact Olympus for instructions on affected lot verification
  2. Do not use units from affected lots until you receive guidance from Olympus
  3. Report any incidents of ceramic tip detachment to Olympus and the FDA

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Olympus BiCOAG Hemostasis Probe 7Fr Coaxial Connector. Model/Catalog Number: CD-B620LA.
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 4 products — see all of them.

Details

Units affected
2,142 units
Illnesses reported
None reported
Manufactured in
3500 Corporate Pkwy, Center Valley, PA, United States
Classification
Class II
Recall number
Z-3246-2026

In the FDA's words

Description
Olympus BiCOAG Hemostasis Probe 7Fr Coaxial Connector. Model/Catalog Number: CD-B620LA. Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive. Lot Code: Model: CD-B620LA. UDI-DI: 00821925039483. Lot Numbers: All unexpired lots. Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Latin America, Singapore, Hong Kong, and EU. Affected quantity: 2,142 units
Hazard
Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.