Moderate riskFDA DeviceClass II

Olympus BiCOAG Hemostasis Probe recalled for ceramic tip detachment risk

Olympus has recalled certain BiCOAG Hemostasis Probe units (Model CD-B622LA) because the ceramic tip may detach during use. This can happen because some units were manufactured without the correct amount of adhesive securing the tip. No injuries have been reported so far.

From the FDA notice · Read the official notice ↗ · All Olympus recalls

What to do

  1. If you have this device, stop using it immediately.
  2. Contact your healthcare facility or Olympus customer service to verify if your unit is affected.
  3. Do not attempt to repair or reuse the device.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Olympus BiCOAG Hemostasis Probe 7Fr Fixed Pin Connector. Model/Catalog Number: CD-B622LA.
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 4 products — see all of them.

Details

Units affected
816 units
Illnesses reported
None reported
Manufactured in
3500 Corporate Pkwy, Center Valley, PA, United States
Classification
Class II
Recall number
Z-3247-2026

In the FDA's words

Description
Olympus BiCOAG Hemostasis Probe 7Fr Fixed Pin Connector. Model/Catalog Number: CD-B622LA. Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive. Lot Code: Model: CD-B622LA. UDI-DI: 00821925039452. Lot Numbers: All unexpired lots Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Latin America, Singapore, Hong Kong, and EU. Affected quantity: 816 units
Hazard
Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.

4 products in this recall

Olympus BiCOAG probes recalled for ceramic tip detachment risk

Olympus is recalling four BiCOAG Hemostasis Probe models because they may have been made without the full amount of adhesive, so the ceramic tip could detach. The recall covers 3,467 units sold in the US and several other countries. No injuries have been reported.

The agency lists each product separately. Check every one you might have.