FDA-Device2026-10-07Class IIMISBRANDING

Olympus M3 Gold Autoclavable Lateral Telescope 70 recalled for sterilization parameter discrepancies

OTHERNationwide distributionMedical devices

What this means

Olympus has identified differences between the sterilization instructions in the U.S. Instructions for Use and those in the Japanese package insert for the M3 Gold Autoclavable Lateral Telescope 70 (model M3-70A). This discrepancy could lead to improper sterilization if the wrong parameters are followed. The device was distributed internationally, with 50 units affected.

  • If you use or manage an M3-70A telescope, verify you are following the correct sterilization parameters for your region.
  • Contact Olympus Corporation of the Americas for clarification on the correct sterilization instructions for your device.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.

Intended for
All ages
Illnesses reported
None reported
Sold states
International distribution to the country of Japan.
Affected count
50 units
Manufactured in
3500 Corporate Pkwy, Center Valley, PA, United States
Products
Olympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3300-2026