AB Sciex, LLC
4 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for AB Sciex, LLC and translate it into plain English — free weekly digest, optional same-day alerts.
- 2023-10-11FDA-DeviceClass IICitrine Triple Quad Mass Spectrometer, Part Number 5063684
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
- 2023-10-11FDA-DeviceClass IICitrine QTRAP Mass Spectrometer, Part Number 5063685
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
- 2023-10-11FDA-DeviceClass II4500MD Triple Quad Mass Spectrometer, Part Number 5032522
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
- 2023-10-11FDA-DeviceClass IIAPI3200MD Mass Spectrometer, Part Number 4466230
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.