Alphatec Spine, Inc.
3 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Alphatec Spine, Inc. and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-02-05FDA-DeviceClass IICalibrate CCX Interbody System for spinal fusion procedures: Implants: (1) Calibrate CCX-O, 7 x 10 x 25mm 5 degrees, Part #325-07102505, nonsterile; (2) Calibrate CCX-O, 7 x 10 x 25mm 10 degrees, Part #325-07102510, nonsterile; (3) Calibrate CCX-O, 7 x 10 x 25mm 15 degrees, Part #325-07102515, nonsterile; (4) Calibrate CCX-O, 7 X 10 X 30mm 5 degrees, Part #325-07103005, nonsterile; (5) Calibrate CCX-O, 7 x 10 x 30mm 10 degrees, Part #325-07103010, nonsterile; (6) Calibrate CCX-O,
Complaints have been received regarding post-operative implant collapse.
- 2024-03-06FDA-DeviceClass IIIdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 132-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm 132-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm 132-10-90-250 / IdentiTi ALIF SA Graft Bolt, 09 x 25 mm 132-10-90-300 / IdentiTi ALIF SA Graft Bolt, 09 x 30 mm 132-10-90-350 / IdentiTi ALIF SA Graft Bolt, 09 x 35 mm 232-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 232-10-85-300 / IdentiTi
Due to reports of intraoperative graft bolt implantation breakages.
- 2023-09-06FDA-DeviceClass IILIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems, IdentiTi LIF, Transcend LIF and Battalion LLIF,
Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.