- 2025-01-01FDA-DeviceClass IMonitoring Service Application (MSA). Software to process, analyze, display, and report cardiac events in ECG data received from compatible devices.
Not all Electrocardiogram (ECG) events received July 2022-July 2024 were not properly routed and subsequently reviewed due to an analysis step being disabled with the monitoring service application software.