- 2026-07-22FDA-DeviceClass IIDaavlin Photera 600 3-D phototherapy device recalled for incorrect UV calibration
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- 2026-07-22FDA-DeviceClass IIDaavlin Photera 600 phototherapy device recalled for incorrect UV output
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- 2026-07-22FDA-DeviceClass IIDaavlin Photera 400 phototherapy device recalled for incorrect UV calibration
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).