- 2025-03-05FDA-DeviceClass IISIGNA MR380, NMRI system, Non-US Only
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
- 2025-03-05FDA-DeviceClass IISIGNA MR355, SIGNA MR360, NMRI system
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
- 2025-03-05FDA-DeviceClass IIOptima MR360, NMRI system
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
- 2025-03-05FDA-DeviceClass IIBrivo MR355, NMRI system
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
- 2025-03-05FDA-DeviceClass IISIGNA Creator, SIGNA Explorer, NMRI systems
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80662DA: China Only: Discovery CT750HD
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80522DA: China Only: Revolution Discovery CT
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80212DA: EU: Operator Console (for BrightSpeed, Elite/BrightSpeed Elite Select), Non-EU: BrightSpeed Elite, BrightSpeed Elite Select; X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B79632DA: EU: RT Refresh Package, Non-EU: Discovery RT, Optima CT580, or Discovery CT590 RT; X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Revolution Maxima Select, Model Number, 6670000-55; X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Revolution Maxima M, Model Number, 6670000-2; X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Revolution Maxima, Model Numbers: 1) 6665000-2, 2) 6670000-2, 3) 6670000-200, 4) 6670000-6; X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Revolution Frontier, Model Numbers: 1) 5232085-213, 2) 5873009; X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Revolution EVO, Model Number 5454001-260; X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Revolution Eagle, Model Number 5849504-3, X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Revolution Ascend Elite, Model Number 6967000-100; X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Revolution Ascend, Model number 6969000-100; X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Revolution Advance, Model number 5849504; X-ray/computed tomography systems
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Revolution ACE, Model number 6670000-3; X-ray/computed tomography systems
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Optima 680 Expert, Model number 6670000; X-ray/computed tomography system
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Optima 520, Catalog number 5439126; X-ray/computed tomography systems
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-11FDA-DeviceClass IIGE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as:
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- 2024-12-04FDA-DeviceClass IIGE Proteus XR/A radiographic system
Steel cables in the Wall Stand for certain Proteus XR/a systems serviced by GE HealthCare have not been replaced at the intervals detailed in the Preventative Maintenance Action List.
- 2024-09-04FDA-DeviceClass IIGE Healthcare Proteus XR/a, stationary X-Ray System, Model/Catalog Numbers: 1) 2259973; 2) 2259988; 3) 2266999; 4) 2373519; 5) 2379827; 6) 5729352; 7) 2259973-1; 8) 2259973-2; 9) 2259973-3; 10) 2259973-4; 11) Not Available
The XR system, Proteus XR/a, does not have a de-installation manual describing process steps for de-installing the devices. De-installation instruction process steps must be followed when de-installing these devices to avoid personnel injury during de-install.