GE Healthcare
46 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for GE Healthcare and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-04-09FDA-DeviceClass IIGE Healthcare Centricity Cardiology CA1000 (CA1000), Model Numbers: 1) 2033901-001; 2) 2038437-001; 3) 2038437-009; 4) 2038437-014; 5) 2038437-015 ; 6) 2038437-0XX; 7) 2109571-007; 8) 2109571-010; 9) 2109571-011; 10) 2109571-012; 11) MANLEGACY111; 12) MANLEGACY112; 13) MANLEGACY47; 14) P00011TS.
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
- 2025-04-09FDA-DeviceClass IIGE Healthcare Centricity Radiology RA600 (RA600), Model Numbers: 1) 2022296-001; 2) 2033901-001; 3) 2038437-001; 4) 2038437-011; 5) 2038437-012; 6) 2038437-013; 7) 2038437-014; 8) 2038437-0XX; 9) 2098071-004; 10) 2109571-002; 11) 2109571-006; 12) 2109571-010; 13) 2109594-009; 14) 2109594-011; 15) K2090PS; 16) K2090PTED; 17) MANLEGACY114; 18) MANLEGACY64.
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
- 2025-04-09FDA-DeviceClass IIGE Healthcare Centricity PACS-IW (PACS-IW), Model Numbers: 1) 2042988-001; 2) 2049587-009; 3) 2049587-011; 4) 2049587-012; 5) 2049587-015; 6) 2049588-012 ; 7) 2052829-001.
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
- 2025-04-09FDA-DeviceClass IIGE Healthcare Centricity Universal Viewer (UV), Model Numbers: 1) 2066908-077; 2) 2066908-136; 3) 2066908-150; 4) 2088026-003; 5) 2088026-026; 6) 2088026-032; 7) 2088026-043; 8) 2088026-110; 9) 2088026-115; 10) 2088026-132; 11) 2088026-306; 12) 2088026-308; 13) 2088026-406; 14) 2088026-506; 15) 2088026-614; 16) 2088026-713; 17) 2088026-714; 18) 2088026-715; 19) 2088026-723; 20) 2088026-806; 21) 2088026-906; 22) 2088026-936; 23) 2088026-940; 24) 20880
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
- 2024-11-13FDA-DeviceClass IIGE Healthcare Centricity Universal Viewer Zero Footprint, Model Numbers: 1) 2089507-083, GDZH-JINWAN-Centricity 2) 2089507-093, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and potentially manipulate patient data.
- 2024-10-16FDA-DeviceClass IIGE Healthcare Centricity Universal Viewer Zero Footprint Client, Model/Catalog/Code Numbers: a)¿¿¿¿¿ 2089507-120; b)¿¿¿¿ 2089507-129; c)¿¿¿¿¿ 2089507-131; d)¿¿¿¿ 2089507-135; e)¿¿¿¿ 2089507-136; f)¿¿¿¿¿¿ 2089507-140; g)¿¿¿¿¿ 2089507-144; h)¿¿¿¿ 2089507-145; i)¿¿¿¿¿¿ 2089507-149; j)¿¿¿¿¿¿ 2089507-153; k)¿¿¿¿¿ 2089507-154; l)¿¿¿¿¿¿ 2089507-155; m)¿¿¿ 2089507-156; n)¿¿¿¿ 2089507-157; o)¿¿¿¿ K2037DAED; p)¿¿¿¿ K2037DBED; q)¿¿¿¿ K2037DED; r)¿¿¿¿¿ K2037DJED; s)¿¿¿
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) version v6.0 SP11.x where the latest report is not shown by default to the user. The issue may occur when a study has multiple DICOM structured reports (i.e. preliminary, final, and/or addendum) stored in different series.
- 2024-10-02FDA-DeviceClass IIGE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.
GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.
- 2024-09-25FDA-DeviceClass IIGE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008; b) 2110344-009; c) 2110344-010; d) 2110344-011; e) 2110344-013; f) 2110344-019; g) 2110344-025; h) 2110344-029; i) 2110344-030; j) 2110344-032 ; Radiological image processing system
Footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a different time, and only when launched on the following browsers. " Internet Explorer 10 (IE10) " Internet Explorer 11 (IE11) " Firefox.
- 2023-11-01FDA-DeviceClass IIUniversal Viewer Workflow Manager, Image processing radiological system
When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.
- 2023-11-01FDA-DeviceClass IICentricity Universal Viewer Workflow Manager, Image processing radiological system
When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.
- 2023-11-01FDA-DeviceClass IICentricity Universal Viewer, software versions 6.0 SP0 through 6.0 SP7.1; Radiological Image Processing System
GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW with Universal Viewer V5.0 and Centricity Universal Viewer V6.0 with PACS-IW foundation.
- 2023-11-01FDA-DeviceClass IICentricity PACS-IW with Universal Viewer, versions 5.0 SPx with PACS-IW foundation; Radiological Image Processing System
GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW with Universal Viewer V5.0 and Centricity Universal Viewer V6.0 with PACS-IW foundation.
- 2023-11-01FDA-DeviceClass IICentricity PACS-IW, software versions V3.7.x through 3.7.3.9 SP3; Radiological Image Processing System
GE HealthCare has become aware of an issue where images from two different patients may be contained in a single study, when stored in Centricity PACS-IW V3.7.x, Centricity PACS-IW with Universal Viewer V5.0 and Centricity Universal Viewer V6.0 with PACS-IW foundation.
- 2023-05-10FDA-DeviceClass IICentricity PACS-IW, Product ID Numbers: a) 2041037-001, b) 2042988-001, c) 2049587-009, d) 2049587-011, e) 2049587-012, f) 2049587-015, g) 2052829-001, h) Not applicable; Radiological Image Processing System
GE HealthCare has become aware of an issue where information from two different patients can be mismatched when correcting patient or study information.
- 2023-05-10FDA-DeviceClass IICentricity Universal Viewer, Centricity Universal Viewer 6.0, Model/Catalog Numbers: a) 2088026-003, b) 2088026-026, c) 2088026-032, d) 2088026-043, e) 2088026-044, f) 2088026-115, g) 2088026-132, h) 2088026-306, i) 2088026-308, j) 2088026-406, k) 2088026-506, l) 2088026-613, m) 2088026-713, n) 2088026-714, o) 2088026-723, p) 2099714-001, q) 2100169-001, r) 2100491-006, s) 2102675-010, t) 2102676-001, u) 2104867-044, v) 2104867-045; Radiological Image Processing System
GE HealthCare has become aware of an issue where information from two different patients can be mismatched when correcting patient or study information.
- 2023-05-10FDA-DeviceClass IICentricity PACS-IW with Universal Viewer, Model/Catalog Numbers: a) 2066908-061, b) 2066908-077, c) 2066908-086, d) 2066908-111, e) 2066908-123, f) 2066908-136, g) 2066908-150; Radiological Image Processing System
GE HealthCare has become aware of an issue where information from two different patients can be mismatched when correcting patient or study information.