- 2025-10-08FDA-DeviceClass IIGetinge 88-Series Washer- Disinfector. Model Number: 88-5.
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
- 2025-05-14FDA-DeviceClass IIWasher Disinfector Aquadis 56; Model Number: 56M
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
- 2025-05-14FDA-DeviceClass IIWasher Disinfector Aquadis 56; Model Number: 56A
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.