Brand
HemoCue AB
1 federal recall on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for HemoCue AB and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-07-09FDA-DeviceClass IIThe HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201 (or DM) Analyzer with plasma conversion that multiplies the measured whole blood glucose value by a factor of 1.112 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. The box contains 4 vials with 25 microcuvettes in each vial.
Glucose microcuvettes experienced transit time outside of limits, so stability through the whole microcuvette lifetime can no longer be guaranteed. The affected products may not have been stored at the recommended labeled storage conditions which may have impacted the safety, quality, identity, potency, and purity of the product. This failure could increase the probability of an incorrect test result or a delayed test result.
Get notified about new HemoCue AB recalls
Free weekly digest. We email when a new federal recall affects HemoCue AB or anything else in your household.
Get the Sunday BriefExplore other recall topics
By hazard