RAYSEARCH LABORATORIES AB
36 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for RAYSEARCH LABORATORIES AB and translate it into plain English — free weekly digest, optional same-day alerts.
- 2024-05-08FDA-DeviceClass IIRayStation 8.0.0.61, 8.0.1.10. Radiation Therapy Treatment Planning System.
Potential for reported SSD to be too high.
- 2024-05-08FDA-DeviceClass IIRayStation 7.0.0.19. Radiation Therapy Treatment Planning System.
Potential for reported SSD to be too high.
- 2024-04-17FDA-DeviceClass IIRayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.
A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
- 2023-11-08FDA-DeviceClass IIRayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care Product name (build number): (1)RayCare 5B SP1 (5.1.1.60246); (2)RayCare 5B SP3 (5.1.3.60023); (3) RayCare 6A (6.0.0.60553)
An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with single- or multi-selectable values.
- 2023-09-06FDA-DeviceClass IIRayCare software, include version number 5A, 5B, 6A, including service packs-An oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care. Model Numbers: 5.0.0, 5.0.1, 5.1.1,5.1.2, 5.1.3, 6.0.0
An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstances
- 2023-03-29FDA-DeviceClass IIRaystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0. Software Version Numbers 9A, 9B, 10A, 10B, 11A, 11B and 12A including service packs
It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.