Brand
TAS Medical Inc
1 federal recall on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for TAS Medical Inc and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-03-12FDA-DeviceClass IITISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
Get notified about new TAS Medical Inc recalls
Free weekly digest. We email when a new federal recall affects TAS Medical Inc or anything else in your household.
Get the Sunday BriefExplore other recall topics