- 2024-09-18FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMedline Semi-rigid suction liners labeled as follows: a) SEMI RIGID SUCTION LINER, REF OR515; b) SEMI RIGID SUCTION LINER WITH TUBING, REF OR526K; c) SEMI RIGID SUCTION LINER WITH TUBING, REF OR527K; d) SEMI RIGID SUCTION LINER WITH TUBING, REF OR529K
Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- 2024-09-18FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIADVIA Centaur CA 19-9 (250 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10491244
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
- 2024-09-18FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIADVIA Centaur CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10491379 SMN: 11206239
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
- 2024-09-18FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica IM CA 19-9 (250 Test - REF)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 11206239
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
- 2024-09-18FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica IM CA 19-9 (250 Test)- In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN:10995489
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
- 2024-09-18FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10995490
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
- 2024-09-18FDA-DeviceZoe Medical IncorporatedClass IISpacelabs Ultraview DM4 Vital Signs Monitor (all variations -M, -N, -MF, -NF, -MW, -NW, -MFW, -NFW), Model Number 91331
This product labeling does not adequately provide warning about the risks associated with the use of damaged batteries or batteries past the intended use life.
- 2024-09-18FDA-DeviceZoe Medical IncorporatedClass IISpacelabs Ultraview DM3 Vital Signs Monitor (all variations -M, -MT, -N, -NT), Model Number 91330
This product labeling does not adequately provide warning about the risks associated with the use of damaged batteries or batteries past the intended use life.
- 2024-09-18FDA-DeviceSmith & Nephew Inc.Class IIROTATION MEDICAL TENDON STAPLES (8) (narrow indication)-Intended for fixation of prosthetic material to soft tissues. Product Number: 72205201
Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch, compromising product sterility
- 2024-09-18FDA-DeviceSmith & Nephew Inc.Class IIROTATION MEDICAL TENDON STAPLES (8) (broad indication)-Intended for fixation of prosthetic material to soft tissues. Product Number: 2504-1
Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch, compromising product sterility
- 2024-09-18FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMedline ReNewal Reprocessed Ethicon Harmonic ACE+7 Shears w/ Adv. Hemostasis Use Only w/GEN11 SW v2018-1 and Lower 5mm x 36cm. Item Number: HARH36R.
Potential for an incomplete seal on the packaging tray.
- 2024-09-18FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMedline ReNewal Reprocessed Covidien LigaSure Impact, Sealer/Divider, Nano-Coated, Compatible w/ FT10 18cm FT10 SW Version 4.0.1.15 or lower. Item Number: LF4418CR.
Potential for an incomplete seal on the packaging tray.
- 2024-09-18FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMedline ReNewal Reprocessed Covidien LigaSure Maryland Jaw Sealer/Divider, Nano-Coated, Compatible w/ FT10 37 cm FT10 SW Version 4.0.1.15 or lower. Item Number: LF1937CR.
Potential for an incomplete seal on the packaging tray.
- 2024-09-18FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMedline ReNewal Reprocessed Covidien LigaSure Blunt Tip, Sealer/Divider, Nano-Coated, Compatible w/ FT10 37cm FT10 SW Version 4.0.1.15 or lower. Item Number: LF1837CR.
Potential for an incomplete seal on the packaging tray.
- 2024-09-18FDA-DeviceIon Beam Applications S.A.Class IIIBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11 versions, and PTS-12 versions
Safety Parameters Verification Mechanisms can be deactivated in clinical runtime
- 2024-09-18FDA-DeviceSmiths Medical ASD Inc.Class Ismiths medical portex BLUperc Percutaneous Diliation Procedural Trays with BLUselect tracheostomy tubes, labeled as follows: 1) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 7.0 mm BLUselect trach tube, Item Number 101/561/070; 2) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 8.0 mm BLUselect trach tube, Item Number 101/561/080; 3) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 9.0 mm BLUselect trach tub
There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.
- 2024-09-18FDA-DeviceSmiths Medical ASD Inc.Class Ismiths medical portex BLUgriggs Percutaneous Dilation Procedural Trays with BLUselect tracheotomy tubes, labeled as follows: 1) BLUgriggs Percutaneous Dilation Procedural Tray with forceps and 8.0mm BLUselect trach tube (drugs, US ONLY), Item Number 101/540/080; 2) BLUgriggs Percutaneous Dilation Procedural Tray with forceps and 9.0mm BLUselect trach tube (drugs, US ONLY), Item Number 101/540/090; 3) BLUgriggs Percutaneous Dilation Procedural Kit with forceps and 7.0mm BLUselect trach tub
There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.
- 2024-09-18FDA-DeviceSmiths Medical ASD Inc.Class Ismiths medical portex BLUselect, tracheostomy tube, labeled as follows: 1) BLUSELECT 6.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/060CZ; 2) BLUSELECT 7.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/070CZ; 3) BLUSELECT 7.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/075CZ; 4) BLUSELECT 8.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/080CZ; 5) BLUSELECT 8.5, CUFFED, NON-FEN, 10/BX, Item Number 101/800/085CZ; 6) BLUSELECT 9.0, CUFFED, NON-FEN, 10/BX, Item Number 101/800/090CZ; 7) BLU
There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.
- 2024-09-18FDA-DeviceSol-Millennium Medical Inc.Class IISyringe Tip Caps labeled as: 1) SOL-M Syringe Tip Cap, Blue, Model No. STCB01, UPC N/A; 2) SOL-M Syringe Tip Cap, Red, Model No. STCB02, UPC N/A; 3) SOL-M Red Cap, Model No. 110101150020, UPC N/A 4) SOL-M Syringe Tip Cap Tray, Black, Model Number STCT01, UPC N/A
Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.
- 2024-09-18FDA-DeviceSol-Millennium Medical Inc.Class IISyringes with Low Dead Space labeled as: a) SOL-M 1ml Slip Tip Syringe w/o needle (low dead space), Model No. 180011LDS, UPC 840368511899; b) SOL-M 1ml Slip Tip Syringe w/o needle (low dead space), Model No. P180011LDS, UPC 810018097554; c) SOL-M 1ml Luer Lock Syringe w/Exch Needle 23G*1" (low dead space), Model No. 181023L, UPC Sold in Canada Only; d) SOL-CARE 1ml Luer Lock Syringe w/Safety Needle 25G*5/8" low dead space), Model No. 12558LDS, UPC 810062435319; e) SOL-M 1ml Luer Lock
Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.
- 2024-09-18FDA-DeviceSol-Millennium Medical Inc.Class IISyringes with Eccentric Tips labeled as: 1) SOL-M 10ml Eccentric Tip Syringe w/o Needle, Model No. P180010ET, UPC 81001809551; 2) SOL-M 20ml Eccentric Tip Syringe w/o Needle, Model No. P180020ET, UPC 810018098605; 3) SOL-M 60ml Eccentric Tip Syringe w/o Needle, Model No. P180060ET, UPC 810018098698;
Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.
- 2024-09-18FDA-DeviceSol-Millennium Medical Inc.Class IISyringes with Luer Slip Tips labeled as: 1) SOL-M 1ml Slip Tip Syringe w/o Needle, Model No. 180011, UPC 8183923010138; 2) SOL-M 1ml Slip Tip Syringe w/o Needle, Model No. P180011, UPC 810018097523; 3) SOL-M 20ml Slip Tip Syringe w/o Needle, Model No. 180020ST, UPC 840368511998; 4) SOL-M 3ml Slip Tip Syringe w/o Needle, Model No. P180003ST, UPC 810018097493; 5) SOL-M 5ml Slip Tip Syringe w/o Needle, Model No. P180005ST, UPC 810018097462; 6) SOL-M 10ml Slip Tip Syringe w/o Needle, Mo
Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.
- 2024-09-18FDA-DeviceSol-Millennium Medical Inc.Class IISyringes with exchangeable needles labeled as: SOL-CARE: 1) SOL-CARE 10ml Luer Lock Syringe w/Safety Needle 21G*1 1/2 (needle aside), Model No. 102115SN, UPC 818392019858; 2) SOL-CARE 10ml Luer Lock Syringe w/Safety Needle 22G*1 1/2 (needle aside), Model No. 102215SN, UPC 818392019865; 3) SOL-CARE 1ml Luer Lock Syringe w/Safety Needle 25G*1 (needle aside), Model No. 12510SN, UPC 818392018851; 4) SOL-CARE 1ml Luer Lock Syringe w/Safety Needle 25G*5/8 (needle aside), Model No. 12558SN, U
Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.
- 2024-09-18FDA-DeviceSol-Millennium Medical Inc.Class IISyringes with Fixed Needles labeled as: a) SOL-M: 1) SOL-M 1ml TB Syringe w/Fixed Needle 25G*5/8'', Model No. 181025, UPC 818392014884; 2) SOL-M 3ml Luer Lock Syringe w/Blunt Fill Needle 18G*1 1/2'', Model No. 1831815B, UPC 818392011593; 3) SOL-M 5ml Luer Lock Syringe w/Blunt Fill Needle 18G*1 1/2'', Model No. 1851815B, UPC 818392011609; 4) SOL-M 10ml Luer Lock Syringe w/Blunt Fill Needle 18G*1 1/2'', Model No. 1811815B, UPC 818392011616;
Products in distribution were found by FDA to be to be substantially different from the current cleared 510(k)s with a major change or modification to the device s intended use.
- 2024-09-18FDA-DeviceMedtronic NeuromodulationClass IIMedtronic Handset with Communicator, Model TH90Q01, InterStim Smart Programmer and Communicator Kits, containing the Model HH90 Handset and the TM90 Communicator.
A subset of the handsets within the kit may not be able to complete the pairing process with the communicator upon initial setup.
- 2024-09-18FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIANESTHESIA TURNOVER KIT, Pack Number DYNJAA0377A
A non-sterile Mastisol component, which was not labeled as being sterile or non-sterile, was placed in the non-sterile portion of certain sterile convenience kits and distributed. In the event that the user assumes sterility during its use, there is an increased risk of patient infection.
- 2024-09-18FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIANESTHESIA TURNOVER KIT, Pack Number DYNJAA0377
A non-sterile Mastisol component, which was not labeled as being sterile or non-sterile, was placed in the non-sterile portion of certain sterile convenience kits and distributed. In the event that the user assumes sterility during its use, there is an increased risk of patient infection.
- 2024-09-18FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIKIT PEDI CRANI LAMI, Pack Number DYNJ907372A
A non-sterile Mastisol component, which was not labeled as being sterile or non-sterile, was placed in the non-sterile portion of certain sterile convenience kits and distributed. In the event that the user assumes sterility during its use, there is an increased risk of patient infection.
- 2024-09-18FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IICRANIOTOMY PACK, Pack Number DYNJ46552O
A non-sterile Mastisol component, which was not labeled as being sterile or non-sterile, was placed in the non-sterile portion of certain sterile convenience kits and distributed. In the event that the user assumes sterility during its use, there is an increased risk of patient infection.
- 2024-09-18FDA-DeviceUltradent Products, Inc.Class IIBrand Name: Transcend Product Name: Transcend Singles Universal Body (UB) Refill 10pk Model/Catalog Number: 4757- Software Version: N/A Product Description: Transcend universal composite is a light-cured, tooth shade resin composite material (Bis-GMA-based) to be used for direct and indirect posterior and anterior tooth restorations. Transcend universal composite is radiopaque and available in a range of dentin, enamel, and body shades. It is 79% filled by weight and 60-61% filled by v
Due to a manufacturing issue, red particulates present in the expressed composite.