Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2024-06-19FDA-DeviceB Braun Medical IncClass IIStimuplex A, 30 DEG, 24GX1", 0.55x25mm, Catalogue Number: 4894251.
The devices are labeled "not made with DEHP" or the DEHP free symbol; however, the glue used contains traces of DEHP.
- 2024-06-19FDA-DeviceRygel Advanced Machines d/b/a Omtech LaserClass IIOMTech Desktop Laser Engraver
Lack of redundant interlocks, missing required labels and certification labels as required by 21 1010.2.
- 2024-06-19FDA-DeviceCanon Medical System, USA, INC.Class IIMRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020)
The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
- 2024-06-19FDA-DeviceGE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGINGClass IIMyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicated cardiac SPECT (single photon emission computerized tomography system).
There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.
- 2024-06-19FDA-DeviceMckesson Medical-Surgical Inc. Corporate OfficeClass IICoviden Monoject 60ml Syringe Luer-Lock Tip-used to inject fluid into or withdraw fluids from the body MMS Catalog Number: 1159851 Cardinal Catalog Number: 8881560125
Quarantined product was inadvertently distributed
- 2024-06-19FDA-DeviceCovidien, LLCClass IIEsoFLIP, 30mm BALLOON DILATION CATHETER, REF ES-330, Rx Only
Due to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead to various harms.
- 2024-06-19FDA-DeviceAvanos Medical, Inc.Class IIINEOMED* Pharmacy Syringes - Nonsterile (Non ENFit) with below product descriptions: a.0.5mL Oral Syringe Orange NonSterile, Product Code BC-S05EO; b. 12mL Oral Syringe Orange NonSterile, Product Code BC-S12EO; c. 1mL Oral Syringe Orange NonSterile, Product Code BC-S1EO; d. 20mL Oral Syringe Orange NonSterile, Product Code BC-S20EO; e. 35mL Oral Syringe Orange NonSterile, Product Code BC-S35EO; f. 3mL Oral Syringe Orange NonSterile, Product Code BC-S3EO; g. 60mL Oral Syringe Orang
The expiration date was not printed on the outer shipping label, leading to expired syringes being shipped to customers.
- 2024-06-19FDA-DeviceAvanos Medical, Inc.Class IIINEOMED* Pharmacy Syringes - Nonsterile (Non ENFit) with below product descriptions: a. 0.5mL Oral Syringe Blue NonSterile, Product Code BB-S05EO; b. 12mL Oral Syringe Blue NonSterile, Product Code BB-S12EO; c. 1mL Oral Syringe Blue NonSterile, Product Code BB-S1EO; d. 20mL Oral Syringe Blue NonSterile, Product Code BB-S20EO; e. 35mL Oral Syringe Blue NonSterile, Product Code BB-S35EO; f. 3mL Oral Syringe Blue NonSterile, Product Code BB-S3EO; g. 60mL Oral Syringe Blue NonSterile,
The expiration date was not printed on the outer shipping label, leading to expired syringes being shipped to customers.
- 2024-06-19FDA-DeviceAvanos Medical, Inc.Class IIINEOMED* Pharmacy Syringes - Nonsterile (Non ENFit) with below product descriptions: a. 0.5mL Oral Syringe Amber NonSterile, Product Code BA-S05EO; b. 12mL Oral Syringe Amber NonSterile, Product Code BA-S12EO; c. 1mL Oral Syringe Amber NonSterile, Product Code BA-S1EO; d. 20mL Oral Syringe Amber NonSterile, Product Code BA-S20EO; e. 35mL Oral Syringe Amber NonSterile, Product Code BA-S35EO; f. 3mL Oral Syringe Amber NonSterile, Product Code BA-S3EO; g. 60mL Oral Syringe Amber NonS
The expiration date was not printed on the outer shipping label, leading to expired syringes being shipped to customers.
- 2024-06-19FDA-DeviceWondfo USA Co LtdClass IIStreptococcal A rapid test kits, labeled as: Wondfo Biotech Co., Ltd, WondView Strep A Test Kit, Guangzhou Wondfo Biotech Co., Ltd, Wondfo One Step Strep A Swab Test, Dealmed Medical Supplies LLC, Strep A Rapid Test Kit, Jant Pharmacal Corporation, Accustrip VALUPAK Strep A Test Strip, SMC Direct, LLC, RefuAH Strep A Rapid,
Wondfo USA has identified unauthorized distribution of the Preview Strep A Test Kit. Unauthorized distribution of this device would include sales and distribution of this test for at-home use, non-prescription use, over-the-counter use, or direct-to-consumer customers.
- 2024-06-19FDA-DeviceWondfo USA Co LtdClass IIIndependent Medical Co-op Inc., Streptococcal A rapid test kits, labeled as: IMCO Strep A Rapid Test Strip and IMCO Strep A Rapid Test, Part Number STP25-IMC
Wondfo USA has identified unauthorized distribution of the Preview Strep A Test Kit. Unauthorized distribution of this device would include sales and distribution of this test for at-home use, non-prescription use, over-the-counter use, or direct-to-consumer customers.
- 2024-06-19FDA-DeviceWondfo USA Co LtdClass IIStreptococcal A rapid test kits, labeled as: Easy Healthcare Corporation, Areta Strep A Swab Test and EASY at Home Medical LLC, Areta One Step Strep A Swab Test, Part Number ARST-100
Wondfo USA has identified unauthorized distribution of the Preview Strep A Test Kit. Unauthorized distribution of this device would include sales and distribution of this test for at-home use, non-prescription use, over-the-counter use, or direct-to-consumer customers.
- 2024-06-19FDA-DeviceWondfo USA Co LtdClass IIClarity and Clarity Diagnostics LLC, Streptococcal A rapid test kits, Labeled as: a) Clarity Strep A Rapid Test, Part Number CD-STP25, b) Clarity Diagnostics LLC, Clarity Strep A Rapid Test, Part Number CD-STP25V
Wondfo USA has identified unauthorized distribution of the Preview Strep A Test Kit. Unauthorized distribution of this device would include sales and distribution of this test for at-home use, non-prescription use, over-the-counter use, or direct-to-consumer customers.
- 2024-06-19FDA-DeviceWondfo USA Co LtdClass IIPreview Strep A, Streptococcal A rapid test kits, Part Numbers: a) PRE-STA-25, b) PRE-STA-A-10
Wondfo USA has identified unauthorized distribution of the Preview Strep A Test Kit. Unauthorized distribution of this device would include sales and distribution of this test for at-home use, non-prescription use, over-the-counter use, or direct-to-consumer customers.
- 2024-06-19FDA-DeviceThoratec Corp.Class IThoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN, 1286INT and L1286INT. Left ventricular assist device
System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to anxiety /inconvenience, hemodynamic compromise, extended surgical time, and Pump Stop command may be selected
- 2024-06-12FDA-DeviceOlympus Corporation of the AmericasClass IIOES 4000 Hysteroscope, Model/Catalog Number: A4674A
The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
- 2024-06-12FDA-DeviceRemote Diagnostic Technologies Ltd.Class IIPhilips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual
Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.
- 2024-06-12FDA-DeviceGE Medical Systems Information Technologies IncClass IIGE HealthCare CRITIKON ONE-CUF blood pressure cuffs, Model numbers: 1. ONE-A1-1B, ONE-CUF, SMALL ADULT, 1 TB BAYONET, 17 - 25 CM, 20/BOX; 2. ONE-A1-2A, ONE-CUF, SMALL ADULT, DINACLICK 80369-5, 17 - 25 CM, 20/BOX; 3. ONE-A2-1B, ONE-CUF, ADULT, 1 TB BAYONET, 23 - 33 CM, 20/BOX; 4. ONE-A1-1B-L, ONE-CUF, ADULT LONG, 1 TB BAYONET, 23 - 33 CM, 20/BOX; 5. ONEA2-2A, ONE-CUF, ADULT, DINACLICK 80369-5, 23 - 33 CM, 20/BOX; 6. ONE-A2-2A-L, ONE-CUF, ADULT LONG, DINACLICK 80369-5, 23 - 33 CM, 20
Use of the blood pressure cuffs could result in inaccurate non-invasive blood pressure measurements.
- 2024-06-12FDA-DeviceMedtronic Navigation, Inc.-LittletonClass IIO-arm O2 Imaging System-Mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients. Product Number: BI70002000
Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or completion of surgery without the benefit of image guidance.
- 2024-06-12FDA-DeviceNihon Kohden America IncClass IIBSM-3000 Series Bedside Monitor REF BSM-3572A The Life Scope BSM-3000 Series Bedside Monitoring System is intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility.
Due to software issue the device may give false "SpO2 Probe Failure" alarm
- 2024-06-12FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMEDLINE STERILE IRIS SCISSORS CVD/STD REF DYNJ04049 (case) and DYNJ04049H (individual unit)
There is the potential of the tip protector to fall off 4.5" sterile, curved, Iris Scissors
- 2024-06-12FDA-DeviceBecton Dickinson & Co.Class IIBD SARS-CoV-2 Reagents for BD MAX" Systems. RT-PCR test to detect SARS-CoV-2 in nasal, nasopharyngeal, oropharyngeal swab specimens, and saliva specimens.
Assays may produce false negative results due to decreased activity of the enzyme utilized in manufacturing.
- 2024-06-12FDA-DeviceVerathon, Inc.Class IIBrand Name: GlideScope Core 15-inch FHD Model/Catalog Number: 0570-0437 Software Version: Core 15 FHD v1.7 and earlier Product Description: Serialized Monitor
Due to software issues, their is the potential for loss of image or a degradation of the image when using the monitor with other connectable devices.
- 2024-06-12FDA-DeviceVerathon, Inc.Class IIBrand Name: GlideScope Core 15-inch Monitor Model/Catalog Number: 0570-0404 Product Description: Serialized Monitor
Due to software issues, their is the potential for loss of image or a degradation of the image when using the monitor with other connectable devices.
- 2024-06-12FDA-DeviceOrtho-Clinical Diagnostics, Inc.Class IIIVITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in verifying the calibration range of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems when used for the quantitative measurement of vitamin B12 - Product Code: 6800872
The results generated by the VITROS Immunodiagnostic Products Vitamin B12 Range Verifier Lot# 1200 RV Level 2 may cause range verification failure. The issue is detectable by the user when the product fails to meet the required range during processing of RV Level 2 of this lot. There is a potential for a delay in patient test results should the user have a need to switch to an alternate method of range verification including use of an alternate lot. As this issue affects the upper limit range verification, there is no impact to VITROS Immunodiagnostic Products Vitamin B12 assay performance. Were the assay to not perform as expected, the failure would be detectable by Quality Control. Past patient sample results are unaffected, and no review of reported results is required.
- 2024-06-12FDA-DeviceAtrium Medical CorporationClass IIFLIXENE, 4-7X30, 2GDS, GWT-GW. Double-Ended Slider GDS. Vascular graft
Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.
- 2024-06-12FDA-DeviceAtrium Medical CorporationClass IIFLIXENE, 4-7X45, 2GDS, GWT-GW. Double-Ended Slider GDS. Vascular graft
Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.
- 2024-06-12FDA-DeviceAtrium Medical CorporationClass IIFLIXENE, 4-6X45, 2GDS, GWT-GW. Double-Ended Slider GDS. Vascular graft
Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.
- 2024-06-12FDA-DeviceAtrium Medical CorporationClass IIFLIXENE, 4-7X45, 2GDS, GWT. Double-Ended Slider GDS. Vascular graft
Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.
- 2024-06-12FDA-DeviceAtrium Medical CorporationClass IIFLIXENE, 4-6X45, 2GDS, GWT. Double-Ended Slider GDS. Vascular graft
Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.