Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2024-03-06FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IMEDLINE ARTERIAL BUNDLE NO CATHETER, REF ART995
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- 2024-03-06FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IMEDLINE BRONCHOSCOPY TRAY, REF DYNDL1350A
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- 2024-03-06FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IMEDLINE kits and trays, labeled as: a) SUCTION CATH 10FR KIT W/SALINE, REF DYND41470SA, DYND41470SAH; b) TRACH CARE TRAY, REF DYNDA1931B, DYNDA1931BH
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- 2024-03-06FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IMEDLINE TRAY care and cleaning kits and trays for Tracheostomy and Stomas, labeled as: a) STOMA & TRACH CLEANING KIT, REF DYNDJ1129; b) TRACH CARE TRAY, REF DYNDJ1057A; c) TRACHEOSTOMY CARE KIT, REF DYNDA1893A; d) TRACH CARE TRAY, REF DYNDA1482
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- 2024-03-06FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IMEDLINE TRAY TRACH BASIC SOLUTION, REF DYND40511
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
- 2024-03-06FDA-DeviceVentec Life Systems, Inc.Class IVOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
- 2024-03-06FDA-DeviceVyaire MedicalClass IAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8004
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
- 2024-03-06FDA-DeviceVyaire MedicalClass IAirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
- 2024-03-06FDA-DeviceVyaire MedicalClass IAirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
- 2024-03-06FDA-DeviceVyaire MedicalClass IAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REF 2K8005F
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
- 2024-03-06FDA-DeviceVyaire MedicalClass IAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8005
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
- 2024-03-06FDA-DeviceVyaire MedicalClass IAirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Adult Mask, CO2 Detector, REF 2K8004C2
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
- 2024-03-06FDA-DeviceVyaire MedicalClass IAirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Expiratory Filter, REF 2K8004F
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
- 2024-03-06FDA-DeviceVyaire MedicalClass IAirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, without Mask, REF 2K8001
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
- 2024-03-06FDA-DeviceVyaire MedicalClass IAirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, without Mask, REF 2K8000,
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
- 2024-02-28FDA-DevicePTW-FREIBURGClass IISoftware VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology.
Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.
- 2024-02-28FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIThese wipes are saturated with 0.65% bleach solution. These wipes are antimicrobial, disinfectant with sporicidal, virucidal, bactericidal, and fungicidal activites. MEDLINE MICRO-KILL Bleach GERMICIDAL BLEACH WIPES: a) MSC351400AN/Z (7 X 8 , 150 ct canister), b) MSC351410AN/Z (6 x 5 , 150ct canister), c) MSC3560012 (private-labeled 6 x 5 , 150ct canister), d) MSC351450AN (12" x 12", 110ct bucket), e) MSC351450BG4 (12" x 12", 110ct bag)
There is stability concern for products dated with a 24-month shelf life. Product is safe and effective up to 12 months shelf life.
- 2024-02-28FDA-DeviceW.L. Gore & Associates, Inc.Class IIGORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2) Catalog Number VSWVN1008, 10 mm x 8 cm; no holes
Devices without transmural drainage holes may be mislabeled as devices with transmural drainage holes, or vice versa. Potential health impact includes extending procedure time, unplanned device removal, secondary procedure, cholangitis, cholecystitis, and/or pain. The GORE¿ VIABIL¿ Short Wire Biliary Endoprosthesis is a flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed. The endoprosthesis is radially compressed and secured onto the distal end of a catheter-based delivery system for deployment. The catheter provides a means for accessing and implanting the GORE¿ VIABIL¿ Short Wire Biliary Endoprosthesis at the target site in the biliary tract.
- 2024-02-28FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAzurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722223, (2) 722224, (3) 722225, (4) 722226
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
- 2024-02-28FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAzurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722227 (2) 722228
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
- 2024-02-28FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAzurion 3 with a Certeray generator -To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722221 (2) 722222 (3) 722280
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
- 2024-02-28FDA-DeviceAbiomed, Inc.Class IIAutomated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)
Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
- 2024-02-28FDA-DeviceMRP, LLC dba AquabilitiClass IIAQUABILITI AQUASTAT 0.9% Sodium Chloride Injection syringe, USP, 5mL in 12mL, REF 2T0805; and 3mL in 12mL, REF 2T0807, intermediate cartons containing 100 syringes, shipping cases containing 400 syringes, Rx, Sterile.
The saline flush syringes may fail to meet USP criteria for stated sodium chloride content after 12 months.
- 2024-02-28FDA-DeviceIMPULSE DYNAMICS (USA) INCClass IIOPTIMIZER model CCM X11 implantable pulse generator (IPG) devices - Smart Mini and Lite
OPTIMIZER devices may cease to deliver CCM therapy if the device incorrectly detects a charging error. This may cause patients to experience heart failure symptoms similar to before implantation of the device.
- 2024-02-28FDA-DeviceFUJIFILM Healthcare Americas CorporationClass IISynapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
- 2024-02-28FDA-DeviceFUJIFILM Healthcare Americas CorporationClass IISynapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.
The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
- 2024-02-28FDA-Device3M Company - Health Care BusinessClass II3M Specialty Blade Assembly, REF 9660,70-2011-8491-1, 7100213026, used with used with 3M Surgical Clipper Catalog Number 9661L
3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41¿C (105.8¿F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes without applying to skin, the Model 9690 blade maximum temperature may reach 60¿C (140¿F) and the Model 9660 blade maximum temperature may reach 77¿C (170.6¿F)."
- 2024-02-28FDA-Device3M Company - Health Care BusinessClass II3M Specialty Blade Assembly, REF 9660,70-2011-8462-2, 7100213114, used with used with 3M Surgical Clipper Catalog Number 9661L
3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41¿C (105.8¿F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes without applying to skin, the Model 9690 blade maximum temperature may reach 60¿C (140¿F) and the Model 9660 blade maximum temperature may reach 77¿C (170.6¿F)."
- 2024-02-28FDA-DeviceW&H DENTALWERK BUERMOOS GMBHClass IIW&H Surgical handpiece S-12. For Surgical treatment of dental hard tissue Model Number: 30061000
1:2 speed increasing surgical handpieces have been incorrectly laser marked during production. The specific laser marking should read 1:2, but instead it reads 1:1 on the body of the handpiece
- 2024-02-28FDA-DeviceOlympus Corporation of the AmericasClass IIESG PK CUTTING FORCEPS, 5MM, 33CM
The jaw may fracture prior to the procedure during the inspection instructed per the IFU, or during the procedure.