Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2024-01-03FDA-DeviceAcumed LLCClass IIRibLoc U Plus 90 Instrument PRIMARY GUIDE ASSY, LOW PROFILE REF RBL2320 The RBL2320, Low Profile Primary Guide is an instrument used with a power system to compress the RibLoc plates onto the bone, determine the length of the screw to install, locate the position of the holes to drill, and provide proper orientation for the installation of the screws.
Due to potential breakage during use.
- 2024-01-03FDA-DeviceSteris CorporationClass IIHookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope Reprocessors Model # Model Description 2-8-001 HU ZERO CHANNEL NO LT 2-8-010 ASSY, HOOK-UP, OLYMPUS 2-8-014 HOOKUP, OPTIM LT 2-8-030 ASSY, HOOKUP, PENTAX-0K 2-8-040 ASSY, HKUP STORZ 0K 2-8-070 ASSY, HU VISION SCIENCE 2-8-112 ASSY, HOOKUP OLMPUS 1K 2-8-120 ASSY, HOOKUP FUJI 1K 2-8-210 ASSY, HOOK-UP OLYMPUS 2K 2-8-213 ASSY, HOOKUP, OLYMPUS-2K 2-8-218 ASSY HOOKUP OLY 2K 2-8-236 AS
Identification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement of patient procedures.
- 2024-01-03FDA-DeviceLandauerClass IILandauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.
- 2024-01-03FDA-DeviceMedtronic NeuromodulationClass IIMedtronic SynchroMed II, Model 8637-40, Programmable pump
Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.
- 2024-01-03FDA-DeviceMedtronic NeuromodulationClass IIMedtronic SynchroMed II, Model 8637-20, Programmable pump
Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.
- 2024-01-03FDA-DeviceAbbott LaboratoriesClass IIAlinity ci-series System Control Module, REF: 03R70-01
There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3) Customer requests mechanism to prevent user error when creating new lots of value assigned Calibrators; 4) Error in processing tests after CCCintController encountered an error with the assay information downloaded from SCC; 5) The assay editor does not correctly check sample dilution total volume limits; 6) User-Applied Labels not adhering to Reagent Bottles; 7) ICT Module did not expire after warranty was exceeded. These performance issues could lead to erroneous results for multiple analytes.
- 2024-01-03FDA-DeviceARJOHUNTLEIGH POLSKA Sp. z.o.o.Class IIArjo medical beds, Models Enterprise 5000X, Enterprise 8000X, Enterprise 9000X and Citadel, assembled with IndiGo Drive Assistance module and retrofitted with IndiGo Drive Assistance modules
Unintended movement of bed wheels
- 2024-01-03FDA-DeviceARJOHUNTLEIGH POLSKA Sp. z.o.o.Class IIArjo medical beds, Models Enterprise 5000X, Enterprise 8000X, Enterprise 9000X and Citadel, assembled with IndiGo Drive Assistance module and retrofitted with IndiGo Drive Assistance modules
Unintended movement of bed wheels
- 2024-01-03FDA-DeviceMegadyne Medical Products, Inc.Class IProduct Code 0848, MEGADYNE MEGA SOFT Universal Plus Dual Patient Return Electrode, Reusable Patient Return Electrode, Mega Soft Universal Dual Plus
Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
- 2024-01-03FDA-DeviceMegadyne Medical Products, Inc.Class IProduct Code 0847, MEGADYNE MEGA SOFT Universal Plus Patient Return Electrode, Reusable Patient Return Electrode, Mega Soft Universal Plus
Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
- 2024-01-03FDA-DeviceMegadyne Medical Products, Inc.Class IProduct Code 0846, MEGADYNE MEGA SOFT Universal Dual Patient Return Electrode, Reusable Patient Return Electrode, Mega Soft Universal Dual Cord
Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
- 2024-01-03FDA-DeviceMegadyne Medical Products, Inc.Class IProduct Code 0845, MEGADYNE MEGA SOFT Universal Patient Return Electrode, Reusable Patient Return Electrode
Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
- 2024-01-03FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass I(1) Nurse Kit, SKU #EDUC1025, Case UPC 40195327196647, Non-sterile, For Training Purposes Only, Not For Human Use.; and (2) Nursing Tote Kit, SKU #EDUC5056A, Case UPC 40195327153695, Non-sterile, For Training Purposes Only, Not For Human Use.
The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
- 2024-01-03FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass I(1) Nursing Skills kit, SKU #EDUC05038B, Case UPC 40195327019878, Non-sterile, For Training Purposes Only, Not For Human Use.
The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
- 2024-01-03FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass ITracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A, Sterile; (5) Medline Mini Tracheostomy Care Kit, REF CC3T3030A, Sterile; and (6) Medline Mini Tracheostomy Care Kit, REF DYND40622, Non-sterile.
The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
- 2023-12-27FDA-DeviceStryker SpineClass IILITe Decompression Snake Arm, REF 48080230
Product arm was manufactured with the incorrect clamp component. As a result, the affected product arm cannot attach to the arm post shaft.
- 2023-12-27FDA-DeviceSiemens Healthcare Diagnostics IncClass IIRAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 750-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491449
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
- 2023-12-27FDA-DeviceSiemens Healthcare Diagnostics IncClass IIRAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 400-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491448
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
- 2023-12-27FDA-DeviceSiemens Healthcare Diagnostics IncClass IIRAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 250-Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10491447
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
- 2023-12-27FDA-DeviceSiemens Healthcare Diagnostics IncClass IIRAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 100 -Intended for in vitro diagnostic use and is designed to provide the determination in whole blood. SMN: 10844813
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
- 2023-12-27FDA-DeviceMaquet Medical Systems USAClass IIVenous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i System, part numbers 701072780 and 701041802
Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be exposed to inappropriate blood flow, blood temperature, or hematocrit/hemoglobin levels as a result of the CARDIOHELP-i not receiving information from the venous probe.
- 2023-12-27FDA-DeviceAbbott LaboratoriesClass IIAbbott CELL-DYN Ruby, Model CD-Ruby, List number 08H67-01, and Abbott CELL-DYN Sapphire, Model CD Sapphire, list number 08H00-01. Multi-parameter, automated hematology analyzer.
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
- 2023-12-27FDA-DeviceZeus Scientific, Inc.Class IIZEUS ELISA Parvovirus B19 IgM Test System, Product Number SM9Z7701G
The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
- 2023-12-27FDA-DeviceZeus Scientific, Inc.Class IIZEUS ELISA Parvovirus B19 IgG Test System, Product Number SM9Z7701M
The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
- 2023-12-27FDA-DeviceSiemens Healthcare Diagnostics IncClass IIepoc NXS Host Blood Analysis System, Siemens Material Numbers 11413497 (EU), 11413475 (US), 11413506 (CA), 11413517 (MX), 11413518 (ROW), 11413498 (JP), 11413583 (CN, KR), 11413879 (IN) with software versions from 4.9.4 to the current version 4.11.11
Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions, a software problem is causing erroneous results to be printed. When the results are run on the epoc NXS Host and Delete Blood Tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. The worst possible outcome may result in a failure to diagnose hyperkalemia, due to an erroneous depressed potassium level or inappropriate treatment to hyperglycemia due to erroneous glucose reading.
- 2023-12-27FDA-DeviceMaquet Medical Systems USAClass IIICARDIOHELP-i Heart Lung Support System: Blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass (up to six hours). Product REF Numbers: 701072780; 701048012
CARDIOHELP-i System was not properly tested to measure leakage current
- 2023-12-27FDA-DeviceMaquet Cardiovascular, LLCClass IIAcrobat-i Vacuum Positioner System, Model Number XP-5000Z
The Positioner Arm may not tighten or lock during normal use.
- 2023-12-27FDA-DeviceExactech, Inc.Class IIExactech Vantage Mobile Bearing Tibial Inserts, Catalog Numbers: a) 350-41-23, b) 350-42-03
Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
- 2023-12-27FDA-DeviceExactech, Inc.Class IIExactech Vantage Ankle System, Catalog Numbers: a) 350-21-04, b) 350-22-03
Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
- 2023-12-27FDA-DeviceExactech, Inc.Class IIExactech Truliant, Activit-E Knee Insert, Catalog Numbers: a) 02-023-02-0029, b) 02-023-02-0032, c) 02-023-02-0035, d) 02-023-02-0038, e) 02-024-35-1512, f) 02-024-35-1515, g) 02-024-35-1519, h) 02-024-35-2010, i) 02-024-35-2013, j) 02-024-35-2015, k) 02-024-35-2509, l) 02-024-35-2510, m) 02-024-35-2511, n) 02-024-35-2512, o) 02-024-35-3010, p) 02-024-35-3011, q) 02-024-35-3012, r) 02-024-35-3013, s) 02-024-35-3513, t) 02-024-35-3515, u) 02-024-35-4012,
Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.