Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2023-09-27FDA-DeviceETHICON, LLCClass IIPDS Plus Antibacterial (polydioxanone) Suture, Product Numbers PDP072H40, PDP259H50, PDP416H50, PDP441H50, PDP442H50, PDP489G50, PDP489H40, PDP497G51, PDP497H16, PDP497H40, PDP498H40, PDP502V05, PDP507V05, PDP510V05, PDP515V05, PDP662H50, PDP663H50, PDP684H40, PDP9615H40, PDP9625H40, PDP9626H40, PDP9631H40, PDP9715H15, PDP9715H40, PDP9733H16, PDP9861H40, PDP9865H40, XWPDP421H40, XWPDP423H40, XWPDP442H40, XWPDP443H16
Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.
- 2023-09-27FDA-DeviceAbbott MedicalClass IIGallant HF Cardiac Resynchronization Therapy Defibrillator: HF CRT-D REF CDHFA500Q Neutrino HF Cardiac Resynchronization Therapy Defibrillator: NxT HF CRT-D REF CDHFA600Q Entrant HF Cardiac Resynchronization Therapy Defibrillator : HF CRT-D REF CDHFA300Q The primary function of ICDs and CRT-Ds are to treat life-threatening ventricular tachyarrhythmias via anti-tachycardia pacing (ATP), or cardioversion/defibrillation. Implantable Cardioverter Defibrillators (ICDs) can either be a dua
Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.
- 2023-09-27FDA-DeviceAbbott MedicalClass IIGallant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA500Q Gallant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA500Q Neutrino VR Implantable Cardioverter Defibrillator: NxT VR ICD REF CDVRA600Q Neutrino DR Implantable Cardioverter Defibrillator: NxT DR ICD REF CDDRA600Q Entrant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA300Q Entrant DR Implantable Cardioverter Defibrillator: DR ICD REF CDDRA300Q The primary function of ICDs and CRT-Ds
Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.
- 2023-09-27FDA-DeviceGE Medical Systems Information Technologies IncClass IIPrucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX and ComboLab AltiX.
A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.
- 2023-09-27FDA-DeviceTytek Medical IncClass IITyTek PneumoDart- Intended for introduction into the body to facilitate removal of air from the pleural cavity in patients with a tension pneumothorax Model Number: TM-317
Updated IFU to inlclude: (1) identify the risks associated with the hazard of needle decompression and (2) improve the artwork showing placement site location
- 2023-09-27FDA-DeviceTytek Medical IncClass IITyTek Tension Pneumothorax Access Kit TPAK10: 10 gauge, 3.25 inch- Intended for introduction into the body to facilitate removal of air from the pleural cavity in patients with a tension pneumothorax Model Number: TM-310
Updated IFU to inlclude: (1) identify the risks associated with the hazard of needle decompression and (2) improve the artwork showing placement site location
- 2023-09-27FDA-DeviceTytek Medical IncClass IITyTek Tension Pneumothorax Access Kit TPAK 14 gauge 3.25 inch- Intended for introduction into the body to facilitate removal of air from the pleural cavity in patients with a tension pneumothorax Model Number: TM-303
Updated IFU to inlclude: (1) identify the risks associated with the hazard of needle decompression and (2) improve the artwork showing placement site location
- 2023-09-27FDA-DeviceRandox Laboratories Ltd.Class IILiquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215
there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet are incorrect, which may lead to misreporting patient results. Target and range values are correct, and risk is mitigated; most clinical labs will use the target and range values in clinical practice.
- 2023-09-27FDA-DevicePhilips North America LlcClass IIPatient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number: 866390; (3) PIC iX Essentials Product Number: 867093; (4) PIC iX Essentials Expand Product Number: 867154
Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients
- 2023-09-27FDA-DeviceEncore Medical, LPClass IIDJO Surgical: Altivate Reverse, Humeral Stem, Small Shell, Size 6 x 108mm, REF: 533-06-108
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
- 2023-09-27FDA-DeviceEncore Medical, LPClass IIDJO Surgical: Reverse Shoulder System (RSP) Semi Humeral Socket Insert, 32mm, REF: 508-01-032; Baseplate, Glenoid HA-Coat, RSP, 6.5mm X 30mm, REF: 508-32-104; Reverse Shoulder System (RSP) Glenoid, Neutral Head W/ Retaining Screw, 36mm, REF: 508-36-101
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
- 2023-09-27FDA-DeviceEncore Medical, LPClass IIDJO Surgical: CLP Hip, Stem, Polished Neck, Lateral Offset Implant, Size#3, REF: 426-11-030
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
- 2023-09-27FDA-DeviceEncore Medical, LPClass IIDJO Surgical: Delta Ceramic Femoral Head, 44 mm + 4.0 mm, REF: 400-03-443
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
- 2023-09-27FDA-DeviceEncore Medical, LPClass IIDJO Surgical: Empowr Knee, Universal Tibia Baseplate, 6R, REF: 352-02-106
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery
- 2023-09-27FDA-DeviceStryker CorporationClass IIStryker ACM (Advanced Cement Mixing)System: ACM Kit w/ Femoral Breakaway Nozzle, Restrictors & Prox. Med. Press- Allows the user to mix bone cement with a constant high vacuum through the process. Catalog Number: 0306-705-000
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
- 2023-09-27FDA-DeviceStryker CorporationClass IIStryker ACM (Advanced Cement Mixing)System: ACM Kit w/ Femoral Breakaway Nozzle, Restrictors & Prox. Press-Allows the user to mix bone cement with a constant high vacuum through the process Catalog Number: 0306-703-000
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
- 2023-09-27FDA-DeviceStryker CorporationClass IIStryker ACM (Advanced Cement Mixing) System: ACM Kit w/ Femoral Breakaway Nozzle & Prox. Med. Press-Allows the user to mix bone cement with a constant high vacuum through the process Catalog Number: 0306-573-000
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
- 2023-09-27FDA-DeviceStryker CorporationClass IIStryker ACM (Advanced Cement Mixing) System: ACM Kit w/ Femoral Breakaway Nozzle & Solid Blades-Allows the user to mix bone cement with a constant high vacuum through the process Catalog Number: 0306-564-000
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
- 2023-09-27FDA-DeviceStryker CorporationClass IIStryker ACM (Advanced Cement Mixing) System: ACM Kit w/ Femoral Breakaway Nozzle-Allows the user to mix bone cement witha constant high vacuum through the process Catalog Number: 0306-563-000
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
- 2023-09-27FDA-DeviceStryker CorporationClass IIStryker ACM (Advanced Cement Mixing): 180-Gram Cement Cartridge with Breakaway Femoral Nozzle-Allows the user to mix bone cement with a constant high vacuum through the process Catalog Number: 0206-530-000
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
- 2023-09-27FDA-DeviceStryker CorporationClass IIStryker ACM (Advanced Cement Mixing) System: Break-Away Femoral Nozzle-, Allows the user to mix bone cement with a constant high vacuum through the process. Catalog number: 0206-512-000.
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
- 2023-09-27FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMEDLINE STERILE IRIS SCISSORS CVD/STD, REF DYNJ04049
4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user.
- 2023-09-27FDA-DeviceAquila CorporationClass IISofTech Pressure-Sensing Wheelchair Cushions
Defective battery pack in wheelchair cushions can overheat resulting in property damage and one possible injury.
- 2023-09-27FDA-DeviceAquila CorporationClass IIAPK2 Pressure-Sensing Wheelchair Cushions
Defective battery pack in wheelchair cushions can overheat resulting in property damage and one possible injury.
- 2023-09-27FDA-DeviceSenTec AGClass IIMembrane Changer Single-Use, REF MC, each package contains 3 changers; Membrane Changer Reloadable, REF MC-R, each package contains 3 changers; and Membrane Changer Insert, REF MC-I, each package contains 5 inserts. All are components of the Sentec Digital Monitoring System (SDMS).
During rework of components during the manufacturing process, an additional piece of material may have been left in the device, which may result in an impact on measurement performance.
- 2023-09-27FDA-DeviceARROW INTERNATIONAL Inc.Class IIArrow ErgoPack, Pressure Injectable Two-Lumen PICC pre-loaded with Arrow VPS Stylet, REF CDC-35552-VPS; Catheter: Percutaneous, implanted, long-term intravascular catheter
This recall has been initiated due to reports that the product code and product name were incorrectly listed on the lidstock of affected kits.
- 2023-09-27FDA-DeviceARROW INTERNATIONAL Inc.Class IIArrow ErgoPack Pressure Injectable One-Lumen PICC, REF CDC-05541-HPK1A; catheter, intravascular
This recall has been initiated due to reports that the product code and product name were incorrectly listed on the lidstock of affected kits.
- 2023-09-27FDA-DeviceKarl Storz EndoscopyClass IIFive S 5 3x65, REF: 0915612-06, sterile for single use, Sterile EO, Rx Only
Flexible intubation endoscopes sterility assurance can not be confirmed.
- 2023-09-27FDA-DeviceHamilton Medical, Inc.Class IHamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
- 2023-09-27FDA-DeviceHamilton Medical, Inc.Class IHamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.