- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIUltrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding organs Model Number: TGF-UC180J.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IITransnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Number: PEF-V.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IISmall intestinal videoscope-For endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach, duodenum, and small intestine) by oral insertion or the lower digestive tract (including the anus, rectum, sigmoid colon, colon, ileocecal valve, and small intestine) by anal insertion. - Model Number: SIF-Q180.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IISignmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum and sigmoid colon). Model Number: PCF-S.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIColonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-H190DI, PCF-H190DL, PCF-H190I, PCF-H190L, PCF-H190TI, PCF-H190TL, PCF-HQ190I, PCF-HQ190L, PCF-PH190I, PCF-PH190L, PCF-Q180AI, PCF-Q180AL, PCF-S.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIMobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-TM2.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IITracheal videoscope- For airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Number: LF-V.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IITracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: LF-DP, LF-GP, LF-TP.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIHysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIHysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIHysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIGastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q160Z, GIF-Q180, GIF-XP180N, GIF-XP190N, GIF-XTQ160. 510(k) Numbers: K112680, K954451, K051645, K100584, K111756, K011151, K123317.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIGastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for endoscopic surgery within the upper digestive tract. Model Numbers: GF-UC140P-AL5, GF-UCT180, GF-UE160-AL5.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIRhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-VT3.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IICystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIICholeodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIColonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I, CF-2T160L, CF-H170I, CF-H170L, CF-H180AI, CF-H180AL, CF-H180DI, CF-H180DL, CF-H190I, CF-H190L, CF-HQ190I, CF-HQ190L, CF-LV1I, CF-LV1L, CF-Q150I, CF-Q150L, CF-Q160S, CF-Q180AI, CF-Q180AL. 510(k) Numbers: K954451, K112680, K100584, K051645, K111756, K001241, K911278.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIBronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceAizu Olympus Co., Ltd.Class IIBronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190, BF-MP190F, BF-MP60, BF-P150, BF-P190, BF-P60, BF-Q170, BF-Q190, BF-XP190, BF-XP60, BF-XT160, BF-XT190. 510(k)s: K023984, K121959, K963033, K172726, K201758, K033225, K183419.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
- 2023-11-29FDA-DeviceSentinel CH SpAClass IIAlinity c Iron Reagent, Reference Number 08P3920
A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
- 2023-11-29FDA-DeviceSentinel CH SpAClass IIIRON assay, Reference Numbers 6K95-41 and 6K95-30
A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
- 2023-11-29FDA-DeviceConvergent DentalClass IISolea Models 2.0 and 3.0 Laser Surgical Instrument, Foot Pedal assembly SUB-00434 with foot pedal CUS-00226 Rev. K reworked
It has been found that potential unintended laser activation can occur without foot pedal depression.
- 2023-11-29FDA-DeviceMusculoskeletal Transplant Foundation, Inc.Class IIDBX Demineralized Bone Matrix Putty 10cc-Indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury. Model Number: 038100
Units are labeled as 10cc; however the units have a fill volume of 1cc units. This may result in prolongation of surgery.
- 2023-11-29FDA-DeviceMedtronic IncClass IIMedtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130068X5
The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
- 2023-11-29FDA-DeviceMedtronic IncClass IIMedtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130069X6
The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
- 2023-11-29FDA-DeviceMedtronic IncClass IIMedtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6
The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
- 2023-11-29FDA-DeviceUnomedical A/SClass IVariSoft infusion set, Model Number 1002830
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
- 2023-11-29FDA-DeviceUnomedical A/SClass IVariSoft infusion set, Model Number 1002828
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.hyperglycemia). Diabetic ketoacidosis.
- 2023-11-29FDA-DeviceUnomedical A/SClass IVariSoft infusion set, Model Number 1002827
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
- 2023-11-22FDA-DevicePhilips North AmericaClass IISpectral CT on Rails, Software Version 5.1.0, Model Number 728334
This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID93363).