Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2023-08-09FDA-DeviceScican Ltd.Class IISTATCLAVE G4 Chamber Autoclave -Tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. Model: G4 622303
The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and / or damage to the autoclave.
- 2023-08-09FDA-DeviceScican Ltd.Class IISTATCLAVE G4 Chamber Autoclave -Tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. Model: G4 621103
The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and / or damage to the autoclave.
- 2023-08-09FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAzurion 7 M20. X-ray Image guided Therapy System.
The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
- 2023-08-09FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD10 and FD20. X-ray Imaging System.
The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
- 2023-08-09FDA-DeviceCovidien, LLCClass IICovidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM
Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration.
- 2023-08-09FDA-DeviceAbiomed, Inc.Class IImpella RP Flex with SmartAssist intracardiac microaxial blood pump, Product Number 1000323
A higher than expected rate of thrombus formation or deposition has been observed. The risk is for thrombus formations or deposits on indwelling central venous lines or cannulas may break free and enter into the Impella RP Flex, resulting in reduced flow, loss of support, or hemolysis
- 2023-08-09FDA-DeviceHamamatsu CorporationClass IINanoZoomer S360MD Slide scanner system-automated system for creating, viewing, and managing digital slides. The NanoZoomer System creates diagnostic-quality digital images of glass slides containing formalin-fixed paraffin-embedded (FFPE) tissue. Catalog Number: C13220-01MD
If the wrong barcode information is recorded in the NDPi file, a pathologist may refer to the whole slide image (WSI) of a different patient.
- 2023-08-09FDA-DeviceGE OEC Medical Systems, IncClass IIField Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE INTENSIFIER, 9 INCH - used with OEC Flexiview 8800, OEC 9800, and OEC 9900 Systems
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
- 2023-08-09FDA-DeviceGE OEC Medical Systems, IncClass IIOEC Elite Systems with 9-inch Image Intensifier
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
- 2023-08-09FDA-DeviceGE OEC Medical Systems, IncClass IIOEC 9900 Systems with 9-inch Image Intensifier
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
- 2023-08-09FDA-DeviceGE OEC Medical Systems, IncClass IIOEC 9800 Systems with 9-inch Image Intensifier
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
- 2023-08-09FDA-DeviceGE OEC Medical Systems, IncClass IIOEC Flexiview 8800 Systems with 9-inch Image Intensifier
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
- 2023-08-09FDA-DeviceMaquet Cardiovascular, LLCClass IIACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization system for cardiac and chest surgery.
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
- 2023-08-09FDA-DeviceMaquet Cardiovascular, LLCClass IIACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery.
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
- 2023-08-09FDA-DeviceMaquet Cardiovascular, LLCClass IIACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization system for cardiac and chest surgery.
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
- 2023-08-09FDA-DeviceMaquet Cardiovascular, LLCClass IIACCESSRAIL Platform (Standard Blade) Part Number SB-1000, and (Deep Blades) Part Number C-DB-1000. Component of a chest stabilization system for cardiac and chest surgery.
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
- 2023-08-09FDA-DeviceSHIMADZU CORPORATION MEDICAL SYSTEMS DIVISIONClass IIOver-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Vision safire, Versa and G4 units, FluoroSpeed 300 and X1 units
Eight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label behind the bracket, indicating when service is required, to eliminate the risk of the device falling.
- 2023-08-09FDA-DeviceOlympus Corporation of the AmericasClass IIAlways-On Tip Tracked Instruments (SPiN Drive instruments) SPiN Xtend 2.0mm OD, 21ga Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-5450
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
- 2023-08-09FDA-DeviceOlympus Corporation of the AmericasClass IIAlways-On Tip Tracked Instruments (SPiN Drive instruments) 22ga SPiN Flex Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-5410
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
- 2023-08-09FDA-DeviceOlympus Corporation of the AmericasClass IIAlways-On Tip Tracked Instruments (SPiN Drive instruments) Triple Needle Brush, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-5300
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
- 2023-08-09FDA-DeviceOlympus Corporation of the AmericasClass IIAlways-On Tip Tracked Instruments (SPiN Drive instruments) 21ga ANSO Cytology Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-0392
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
- 2023-08-09FDA-DeviceOlympus Corporation of the AmericasClass IIAlways-On Tip Tracked Instruments (SPiN Drive instruments) 19ga ANSO Histology Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-0382
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
- 2023-08-09FDA-DeviceOlympus Corporation of the AmericasClass IIAlways-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8mm OD, Serrated Cup-Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-0372
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
- 2023-08-09FDA-DeviceOlympus Corporation of the AmericasClass IIAlways-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8 mm OD, Oval Cup-Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-0362
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
- 2023-08-09FDA-DeviceOlympus Corporation of the AmericasClass IIAlways-On Tip Tracked Instruments (SPiN Drive¿ instruments) Brush, 15mm L, 1.8mm OD Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic bronchoscopy are currently used for visualizing Model: INS-0352
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
- 2023-08-09FDA-DevicePhilips North AmericaClass IIIncisive CT Plus-Computed Tomography X-Ray System Model: 728149
Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
- 2023-08-09FDA-DevicePhilips North AmericaClass IIIncisive CT Power (China)-Computed Tomography X-Ray System Model: 728148
Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
- 2023-08-09FDA-DevicePhilips North AmericaClass IIIncisive CT for Brazil SKD-Computed Tomography X-Ray System Model: 728146
Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
- 2023-08-09FDA-DevicePhilips North AmericaClass IIIncisive CT-Computed Tomography X-Ray System Model: 728143
Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
- 2023-08-09FDA-DeviceMegadyne Medical Products, Inc.Class IIMEGADYNE Suction Coagulators, Product Code 004125 (Suction Coagulator, Handswitching, 10 Fr, 6 inch (15.24cm)), Product Code 004225 (Suction Coagulator, Handswitching, 8 Fr, 6 inch (15.24cm), Product Code 004325 (Suction Coagulator, Handswitching, 12 Fr, 6 inch (15.24cm)), and Product Code 0041BN (Suction Coagulator, Handswitching, 10 Fr, 6 inch (15.24cm), Bulk Non Sterile)
An issue with the Suction Coagulator product family may lead to fluid ingress into the handpiece. Fluid ingress can lead to intermittent device activation, non-activation, or self-activation when plugged into the electrosurgery unit (ESU).