Medical device recalls
38685 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2022-11-16FDA-DeviceBeaver Visitec International, Inc.Class IIBeaver Arthro-Lok Pointed Tip Curve L 4mm Blade-an orthopedic blade, which may be used in various types of orthopedic surgery (i.e. foot, ankle, hand, knee, hip or spine surgery). Catalog#: 376780
Incorrect configuration of the blades in the package. The blades curve to the right instead of curving to the left. If used may cause cause unintended tissue damage during the procedure or if noticed prior to use may cause a delay in surgery
- 2022-11-16FDA-DeviceAlphatec Spine, Inc.Class IIDisc Preparation instruments contained in the following products: CI-11050-08102515 Longer IdentiTi PO Trial Tip, 11A X 8P X 10 X 25mm, 15¿ TC11050 00840180510810 CI-11050-09102515 Longer IdentiTi PO Trial Tip, 12A X 9P X 10 X 25mm, 15¿ TC11050 00840180510827 CI-11050-10102515 Longer IdentiTi PO Trial Tip, 13A X 10P X 10 X 25mm, 15¿ TC11050 00840180510834 CI-11050-11102515 Longer IdentiTi PO Trial Tip, 14A X 11P X 10 X 25mm, 15¿ TC11050 00840180510841 CI-11050-12102515
Due to manufacturing instruments which included additional features and adequate consideration was not given to the need for a new Regulatory filing.
- 2022-11-16FDA-DeviceAvanos Medical, Inc.Class IIY-CONNECTOR: 6" Small Kink Resistant Tubing Bifurcated Connector /2 Male Luer Locks, 1 Female Luer Lock, Labeled as: SUMMIT MEDICAL PRODUCTS, REF 220227; MPS Acacia CAT NO: BC-006-04; MEDICAL PRODUCTS SPECIALISTS EXTENSION SET, Catalog Number BC00604, pain control pump accessory
Sterile extension sets were distributed without an expiration date
- 2022-11-16FDA-DeviceMedtronic MiniMedClass IIInsulin Pump/Model: MiniMed 620G/ MMT-1750 MiniMed 640G/ MMT-1711, MMT-1712, MMT-1751, MMT-1752
Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.
- 2022-11-16FDA-DeviceMedtronic MiniMedClass IIInsulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.
- 2022-11-16FDA-DeviceCardinal Health 200, LLCClass IICardinal Health Reusable Gel Pack, Hot/Cold Pack, 4.5 in. x 10 in., REF 70304A
Users are being informed that the product should not be used on infants/neonates. A caution statement will be added to the labeling stating: Not for use on Infants/Neonates .
- 2022-11-16FDA-DeviceCardinal Health 200, LLCClass IICardinal Health INSULATED GEL PACK REUSABLE, EXTRA LARGE, Cat. 80600
Users are being informed that the product should not be used on infants/neonates. A caution statement will be added to the labeling stating: Not for use on Infants/Neonates .
- 2022-11-16FDA-DeviceCardinal Health 200, LLCClass IICardinal Health INSULATED GEL PACK REUSABLE, MEDIUM, Cat. 80304A
Users are being informed that the product should not be used on infants/neonates. A caution statement will be added to the labeling stating: Not for use on Infants/Neonates .
- 2022-11-16FDA-DeviceCardinal Health 200, LLCClass IICardinal Health INSULATED GEL PACK REUSABLE, SMALL, Cat. 80204A
Users are being informed that the product should not be used on infants/neonates. A caution statement will be added to the labeling stating: Not for use on Infants/Neonates .
- 2022-11-16FDA-DeviceCardinal Health 200, LLCClass IICardinal Health INSULATED GEL PACK REUSABLE, LARGE, Cat. 80104
Users are being informed that the product should not be used on infants/neonates. A caution statement will be added to the labeling stating: Not for use on Infants/Neonates .
- 2022-11-16FDA-DeviceCardinal Health 200, LLCClass IICardinal Health HOT AND COLD GEL PACK REUSABLE, M, Cat. 70304
Users are being informed that the product should not be used on infants/neonates. A caution statement will be added to the labeling stating: Not for use on Infants/Neonates .
- 2022-11-16FDA-DeviceCardinal Health 200, LLCClass IICardinal Health GEL PACK REUSABLE, SMALL, Cat. 70204
Users are being informed that the product should not be used on infants/neonates. A caution statement will be added to the labeling stating: Not for use on Infants/Neonates .
- 2022-11-16FDA-DeviceCardinal Health 200, LLCClass IICardinal Health GEL PACK REUSABLE, EXTRA SMALL, Cat. 61115
Users are being informed that the product should not be used on infants/neonates. A caution statement will be added to the labeling stating: Not for use on Infants/Neonates .
- 2022-11-16FDA-DeviceFerno-Washington IncClass IIFERNO POWER X1 AMBULANCE COT- Item #0015807/UDI-DI 00190790002183 with 36V Li-Ion Batteries Model Number #0822484/0822483 Note: Transcend Stair Chair Powertraxx (Item #0731403/UDI-DI 00190790004064) not included in this action).
Batteries could become unstable when not properly maintained and/or subjected to repeated drops, power washing directly into battery contacts or abuse, may result in a fire and/or injury
- 2022-11-16FDA-DeviceiRhythm Technologies, Inc.Class IIZio AT Clinical Reference Manual, Part: ALB0031, and Zio AT Important Information Pamphlet, Part: ALB0034. For Zio Ambulatory ECG monitoring system
Ambulatory ECG monitoring system labeling update: 1) Clinically actionable arrhythmia notifications delayed until patient registration is complete; 2) Patches limited to 100 symptomatic and 500 asymptomatic wireless transmissions; once the maximum transmission limit is reached for either type, any further transmissions for that type will cease. Risk of delayed reporting to health providers.
- 2022-11-16FDA-DeviceSEDECAL SAClass IISedecal Soltus Battery Mobile X-Ray System REF Soltus 450 Battery Mobile X-Ray System REF Soltus 500
Due to un-commanded movement of mobile x-ray unit.
- 2022-11-16FDA-DeviceFresenius Medical Care Holdings, Inc.Class IINaturaLyte Liquid Acid concentrate, Part Number 08-0231-4, four (4) one-gallon bottles per bottle case. For dialysis.
The "Nominal Chemical Composition" section of the label incorrectly lists 5.62g/L KCL (potassium chloride). 5.62g/L actually corresponds to the CaCl (calcium chloride) content. The product remains potassium free.
- 2022-11-09FDA-DeviceBaxter Healthcare CorporationClass IISimple Clamp, US/EURO Rails, UPPER EXTREMITY DISTRACTION CMPT, REF F-40018, #800101. To position, support and/or distract the patients hand, wrist, forearm, arm, and shoulder in surgical procedures.
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
- 2022-11-09FDA-DeviceBaxter Healthcare CorporationClass IIAllen Life Assist Beach Chair, BCSSP-BEACH CHAIR , REF A-91501, #104002. For use with orthopaedic procedures using the Fowlers or beach chair position.
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
- 2022-11-09FDA-DeviceBaxter Healthcare CorporationClass IIBeach Chair Clamp, BEACH CHAIR SHLDER POSITIONER CMPT, REF A-90013, #800104. Component to position and support the patients head in surgical procedures.
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
- 2022-11-09FDA-DeviceBaxter Healthcare CorporationClass IIAllen Standard Starburst, Mayfield/Doro, STANDARD SPINE POSITIONERS CMPT, REF A-70710, #800103. Component to position and support the patients head in surgical procedures.
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
- 2022-11-09FDA-DeviceBaxter Healthcare CorporationClass IIC-Flex Polar Head Positnr-Device Only, SSP-NEUROSURGICAL HEADREST, REF A-70701, #103001,
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
- 2022-11-09FDA-DeviceBaxter Healthcare CorporationClass IIAllen Advance Table, RSP-ADVANCE TABLE: a) REF A-71100-US, #107003, b) A-71101-EU, #107003. operating room table
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.
- 2022-11-09FDA-DeviceBecton Dickinson Infusion Therapy Systems Inc.Class IIBD Nexiva Closed IV Catheter System - Single Port REF 383516 20 GA X1.0o in (1.1 x 25 mm) 61 mL/min (3660 mL/hr)
During needle withdrawal, the tip shield may prematurely separate from the catheter adapter and preventing the needle tip from being in the safely shielded (leaving the needle tip exposed).
- 2022-11-09FDA-DeviceBecton, Dickinson and Company, BD BiosciencesClass IIBD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, BD FACSCanto II, and BD FACSCanto flow cytometers to determine the percentages and absolute counts of human lymphocyte subsets in peripheral whole blood for immunophenotyping.
Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant.
- 2022-11-09FDA-DeviceBecton, Dickinson and Company, BD BiosciencesClass IIBD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount Tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount Tubes can be used with the BD FACS Loader.
Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.
- 2022-11-09FDA-DeviceBecton, Dickinson and Company, BD BiosciencesClass IIBD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.
Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.
- 2022-11-09FDA-DeviceMICROVENTION INC.Class IIWEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.
Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels
- 2022-11-09FDA-DeviceJohnson & Johnson Surgical Vision, Inc.Class IITHERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and recording) and, in conjunction with an RF generator, for cardiac ablation.
Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.
- 2022-11-09FDA-DeviceRebiscan, Inc.Class IIBlinq Pediatric Vision Scanner-Pediatric Vision Scanner-intended for automated detection of misalignment of the visual axes using polarized light. Model Number BQ830
Unable to charge the battery pack- battery packs will allow the batteries to be charged if the voltage of any cell is below 2.5V. When this occurs, the battery pack will not accept the charge, could vent and possibly catch fire