- 2021-02-24FDA-DeviceCombat Medical Systems, LLCClass IFresh Whole Blood Transfusion Set (P/N 80-801 NSN 6515016574750) . Blood convenience kit. The Fresh Whole Blood Transfusion Set contains everything needed to test for ABO/Rh compatibility and perform a fresh whole blood transfusion in a compact sealed package.
There is the potential of a bent or disconnected needle in one of the components of the convenience kits.
- 2021-02-24FDA-DeviceCombat Medical Systems, LLCClass ILow Titer Type O FWB Transfusion Set (P/N 80-875 NSN 651501684994). Blood convenience kit. The Transfusion Set includes the items necessary to administer the collected blood to a patient.
There is the potential of a bent or disconnected needle in one of the components of the convenience kits.
- 2021-02-24FDA-DeviceCombat Medical Systems, LLCClass IValkyrie LTOWB Administration (P/N 80-821 NSN 6515016862401). Blood convenience kit. The Valkyrie LTOWB Administration Set contains the items needed to administer the blood to the patient.
There is the potential of a bent or disconnected needle in one of the components of the convenience kits.
- 2021-02-24FDA-DeviceCombat Medical Systems, LLCClass IValkyrie LTOWB Collection (P/N 80-820, NSN 6515016862423). Blood collection convenience kit. The Valkyrie LTOWB Collection Set includes the items needed to allow for pre-screening to identify Low Titer Type O Donors as well as collect the blood from the donor.
There is the potential of a bent or disconnected needle in one of the components of the convenience kits.
- 2021-02-24FDA-DeviceBecton Dickinson & CompanyClass IIBD Hypodermic Syringe 10 ml, eccentric tip. Catalog (Ref No.) 305482
The product packaging for the 10 ml eccentric tip syringe may exhibit open seals. An open seal impacts package integrity and potentially compromises sterility of the syringe.
- 2021-02-17FDA-DeviceHelena Laboratories, Corp.Class IICascade Abrazo aPTT Test Card - Product Usage: are to be used with the Cascade Abrazo analyzer and are intended for the determination of the activated Partial Thromboplastin Time (aPTT) of citrated whole blood or plasma, Model 5722, For Export Only.
Intermittent potential for shorter than expected clot times.
- 2021-02-17FDA-DevicePhadia US IncClass IIEliA dsDNA Well - Product Usage: intended for the In-Vitro quantitative measurement of IgG antibodies directed to dsDNA in human serum and plasma as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE). Material Number: 14-5500-01
Decreased values for EliA ANA Positive Control when using the EliA dsDNA Well lot BFA37/0142 may potentially cause false positive results (increase in patient test results).
- 2021-02-17FDA-DeviceArrow International IncClass IIArrow AGB+ Multi-Lumen CVC Kit 7Fr 3 Lumen- Intended for venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. Product Code: CDC-45703-1A
The lidstock / information card on the kits correctly states the product code as CDC-45703-1A. The banner card on the side of the kit may incorrectly state the product code as CDC-15703-1A and not identify the product as a coated ARROWg+ard Blue PLUS (AGB+) antimicrobial catheter which may cause a minor delay during use and placement of a coated catheter in patients with known sensitivity to chlorohexidine, silver sulfadiazine or sulfa could result in an allergic reaction
- 2021-02-17FDA-DeviceBEMER INT. AGClass IIBemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set), REF. 420201 (B.BOX Professional), REF. 420100 (B.BOX CLASSIC), REF. 410100 (BEMER Classic-Set), REF. 420200 (B.BOX Professional), and REF. 410200 (BEMER Pro - Set)
There potential that electromagnetic fields generated in the therapy system may cause technical defects in a patient using an insulin pumps or a malfunction of a pacemaker.
- 2021-02-17FDA-DeviceDraegar Medical Systems, Inc.Class IIInfinity Acute Care System (IACS) Monitoring Solution: The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals.
The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and the M540 reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring.
- 2021-02-17FDA-DeviceDePuy Orthopaedics, Inc.Class IIUniversal Femoral Sleeves - Product Usage: used primarily in revision surgeries with the S-ROM Noiles, LPS, SIGMA, and LCS Knee products.
The taper dimensions may be out of specification due to a production issue during the manufacturing of these lots.
- 2021-02-17FDA-DeviceBaxter Healthcare CorporationClass IIBaxter ONE-LINK Non-DEHP Y-Type Microbore Catheter Extension Set, Bonded Needle-free IV Connector, Neutral Fluid Displacement, 5.7 " (14 cm), REF 7N8370K - Product Usage: intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use.
There were customer reports of separation between the tubing and Male Luer Lock Adaptor.
- 2021-02-17FDA-DeviceGE Healthcare, LLCClass IIGE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System - Product Usage: Intended for use by a qualified/trained doctor or technologist on both adult and pediatric subjects taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities and other body parts.
GE Healthcare has become aware of a potential safety issue associated with the improper installation of the Precision 600FP monitor suspension. A partial detachment of the monitor suspension could result in potential injury to a patient or operator.
- 2021-02-17FDA-DeviceArjohuntleigh Magog, Inc.Class IIArjo Flat Dynamic Positioning System (DPS) - Product Usage: used in conjunction with passive clip slings is designed for resident transfer in a homecare setting, at nursing homes and other assisted living centers. Please note that the DPS can be utilized only if affixed to ceiling lift or passive floor lift.
Risk of Sling Clip breakage while in use with obsolete Flat Dynamic Positioning System (DPS)
- 2021-02-17FDA-DeviceCytocell Ltd.Class IIAquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric) to the CRLF2 gene
Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.
- 2021-02-17FDA-DeviceRoche Diagnostics Operations, Inc.Class II9180 Electrolyte Analyzer - Product Usage: intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and lithium in the whole blood, serum or plasma, urine, dialysate solutions, or QC materials. catalog number 03157334001 . in vitro diagnostic use.
May display Calcium results on the screen with wrong arrow direction which may lead to misinterpretation of results and incorrect medical decision. If the unit for iCa++ is set to mg/dL (configured as Service Code MGL) and a measured iCa++ value is lower than the normal range, an upward arrow is shown on the display instead of a downward arrow. The display is set per default to mmol/L; therefore, the arrow indicating an alarm reflects normal ranges of iCa++ in mmol/L. If mg/dL is chosen, the arrow warnings on screen still reflect the ranges from mmol/L, rather than mg/dL. The numeric result is displayed correctly.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IIGown Kendall Blue XXL, SKU CT5102 Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IIGown Kendall Blue L, SKU CT5100, CT5101 Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IIGown Kendall Blue RG, SKU CT5000 Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IILab Coats SMS 2X-Large White, SKU 2304LC Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IILab Coats SMS Large White, SKU 2302LC Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IILab Coats SMS Medium White, SKU 2301LC Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IILab Coats SMS 2X-Large Blue, SKU 2204LC Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IILab Coats SMS X-Large Blue, SKU 2203LC Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IILab Coats SMS Large Blue, SKU 2202LC Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IIGown Isolation W/Ties Xl Blue, SKU 2201PG Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IIGown Isolation W/Ties Univ Blu, SKU 1200PG, 2200PG, 552200PG Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IIGown Isolation W/Ties Xl Ylw, SKU 1101PG Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceCardinal Health 200, LLCClass IIGown Isolation W/Ties Univ Ylw, SKU 1100PG, V1100PG, 2100PG, V2100PG, 2150PG, V2100PG Cardinal Health Isolation Gown, Lab Coat, and Kendall" Gowns are intended to provide minimal barrier protection to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate material where minimal fluid settings are expected.
There is a potential for rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns and lab coats.
- 2021-02-17FDA-DeviceWahl Clipper CorpClass IIWAHL 4212 HEATED MASSAGER, SKU F800000004212 - Product Usage: intended to relieve minor muscle aches and pains.
The connection between the massager and heat element can overheat causing smoke or spark, which may pose a fire hazard