- 2020-09-30FDA-DeviceIrvine Biomedical Inc, a St. Jude Medical Co.Class IIInquiry Steerable Diagnostic Catheter, REF: IBI-81104 - Product Usage: INDICATIONS FOR USE The Inquiry fixed curve and steerable electrophysiology catheters are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies. The catheters are commonly placed at the high right atrium, right ventricular apex, and HIS bundle.
Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307 incorrectly contained Model IBI-81104, batch 7397352.If physicians fail to observe the catheter difference, the intracardiac electrograms will not be representative of the desired signals, could result in procedure delay
- 2020-09-30FDA-DeviceBeckman Coulter Inc.Class IIBECKMAN COULTER ACCESS Immunoassay Systems - Access Unconjugated Estriol, REF 33570.
Due to alkaline phosphatase (ALP) associated interference causing falsely elevated uE3 (Unconjugated Estriol assay) results.
- 2020-09-30FDA-DeviceSiemens Medical Solutions USA, IncClass IISoftware: Syngo.via RT Image Suite software version VA30_SP02 with license T_RT_IMAGE_SUITE_PAT_MARK Intended Use: used as an advanced application on a SOM10 scanner SMN: 10496180
Calibration offset between the central CT image plane and the external RT lasers has been set automatically to the factory default of 50.00 cm, if the misadjustment of the offset value is not recognized during the subsequent treatment planning process or remains undetected during the position verification at the treatment system (e.g. linear accelerator), this can potentially result in an incorrect dose or dose applied to the wrong location
- 2020-09-30FDA-DeviceDNA Genotek Inc.Class IIIPrepIT Q2A Kit, Model PT-Q2A-96 containing component reagent AG (PT-QAG-96) 1.4mL and ST (PT-QST-96) 2.9mL - Product Usage: used for the removal of SDS and other inhibitors from samples collected using DNA Genotek oral sample collection kits.
Reagents were shipped to customers after the Use by data indicated on its labeling.
- 2020-09-30FDA-DeviceOraSure Technologies, Inc.Class IIIPointts sistema de ellmlnact!;o de verrugas OTC - Brazil Catalog Number: 1001-0280 - Product Usage: indicated for over-the-counter a cryosurgical system for the treatment of skin tags.
Unit boxes not properly sealed
- 2020-09-30FDA-DeviceOraSure Technologies, Inc.Class IIIPersonelle OTC Skin Tag Remover, Canada Catalog Number: 1001-0314 - Product Usage: indicated for over-the-counter a cryosurgical system for the treatment of skin tags.
Unit boxes not properly sealed
- 2020-09-30FDA-DeviceOraSure Technologies, Inc.Class IIICVS Advanced Wart Remover Kit, Catalog Number: 1001-0407 - Product Usage: indicated for over-the-counter a cryosurgical system for the treatment of common warts and plantar warts.
Unit boxes not properly sealed
- 2020-09-30FDA-DeviceArthroCare CorporationClass IISmith&Nephew SmartStitch PerfectPasser Connector, REF # OM-8010, Rx Only, CE 0123, UDI: 00817470006971
The voluntary recall is being initiated due to a manufacturing error resulting in incorrect upper jaw dimensions, and the upper jaw could potentially disengage or detach during use.
- 2020-09-30FDA-DeviceArthroCare CorporationClass IISpeed Stitch Needle Cassette
Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.
- 2020-09-30FDA-DeviceGE Healthcare, LLCClass IINM/CT 870 DR Model # H3100AS
Rotor bearing screws were found loose on detector.
- 2020-09-30FDA-DeviceGE Healthcare, LLCClass IIDiscovery NM670 Pro Model # H3100XB
Rotor bearing screws were found loose on detector.
- 2020-09-30FDA-DeviceGE Healthcare, LLCClass IIDiscovery 670 DR Model # H3100BT
Rotor bearing screws were found loose on detector.
- 2020-09-30FDA-DeviceGE Healthcare, LLCClass IINM 830 Model # H3910AC
Rotor bearing screws were found loose on detector.
- 2020-09-30FDA-DeviceGE Healthcare, LLCClass IIDiscovery NM 630 Model # H3101RH
Rotor bearing screws were found loose on detector.
- 2020-09-30FDA-DeviceGE Healthcare, LLCClass IINM/CT 870 CZT Model # H3906CW
Rotor bearing screws were found loose on detector.
- 2020-09-30FDA-DeviceGE Healthcare, LLCClass IINM/CT 860, SPECT/CT
Rotor bearing screws were found loose on detector.
- 2020-09-30FDA-DeviceGE Healthcare, LLCClass IINM/CT 850 Model Number: H3907AD
Rotor bearing screws were found loose on detector.
- 2020-09-30FDA-DeviceHologic, IncClass IIPanther Fusion Tube Tray, Part: FAB-15004, REF: PRD-04000
Tube trays may leak which could invalidate patient results or assay worklists, causing delayed results for any assays with which they are used. A possibility exists of false negative result with flu A/B/RSV assay or paraflu assay.
- 2020-09-30FDA-DeviceWright Medical Technology IncClass IIEVOLVE STEM 8.5MM Model # 496S085
Lack of sterility assurance
- 2020-09-30FDA-DeviceSorin Group Italia S.r.l.Class IIPerceval Dual Collapser, Size S/M, REF: ICV1235, Sterile EO, Rx Only, CE, UDI: 08022057014832 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
- 2020-09-30FDA-DeviceSorin Group Italia S.r.l.Class IIPerceval Single Use Accessory Kit, Size M, REF: ICV1350, Sterile EO, Rx Only, CE, UDI: 8022057015259 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
- 2020-09-30FDA-DeviceSorin Group Italia S.r.l.Class IIPerceval Single Use Accessory Kit, Size S, REF: ICV1349, Sterile EO, Rx Only, CE, UDI: 8022057015211 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
- 2020-09-30FDA-DeviceSorin Group Italia S.r.l.Class IIPerceval Single Use Accessory Kit, Size M, REF: ICV1346, Sterile EO, Rx Only, CE, UDI: 8022057015228 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
- 2020-09-30FDA-DeviceSorin Group Italia S.r.l.Class IIPerceval Single Use Accessory Kit, Size S, REF: ICV1345, Sterile EO, Rx Only, CE, UDI: 8022057015204 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
- 2020-09-30FDA-DeviceBeckman Coulter Biomedical GmbHClass IIDxA 5000; Catalog Number:B50516; UDI:15099590707118 - Product Usage: The DxA 5000 is a high-speed, modular, automated sample handling system that performs pre-analytical and post-analytical sample processing and storage. The automation system also sorts, routes, and presents sample tubes to analyzers for analysis. The DxA 5000 also consolidates a variety of analytical instruments, such as an Immunoassay analyzer, into a unified workstation on a track system.
The DxA 5000 instructions for use may be confusing regarding the re-use of sample IDs. Beckman Coulter has received complaints where the re-use of sample barcodes were not being detected by the system .
- 2020-09-30FDA-DeviceBio-Rad Laboratories, Inc.Class IIBioPlex 2200 Syphilis Total & RPR Linked Immunoabsorption Assay, Treponema pallidum:UDI:03610520653012 - Product Usage: intended for the qualitative detection of total (IgG/IgM) antibodies to Treponema pallidum and the qualitative detection and/or titer determination of non-treponemal reagin antibodies in human serum or plasma.
The increased level of RPR reactivity in the BioPlex 2200 Syphilis Total & RPR kit material number 12000650.
- 2020-09-30FDA-DeviceMicroPort Orthopedics Inc.Class IPROFEMUR Neck Extra Long Titanium, Catalog Numbers: PHA01206, PHA01236, PHA01256. Hip prosthesis component
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.
- 2020-09-30FDA-DeviceMicroPort Orthopedics Inc.Class IPROFEMUR Neck Long Titanium, Catalog Numbers: PHA01204, PHA01214, PHA01224, PHA01234, PHA01244, PHA01254, PHA01264. Hip prosthesis component.
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR¿ Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.
- 2020-09-23FDA-DeviceRoche Diagnostics Operations, Inc.Class IIElecsys Anti-TPO Assay on cobas e analyzers 411, 601, 602, Catalog Number 06368590190
Recent internal verification studies found the current RF interference claim to no longer be met for the assay on the cobas e 411, 601, and 602 analyzers. Therefore, the claim has been updated.
- 2020-09-23FDA-DeviceMedtronic Inc.Class IIThe Guardian Connect App CSS7200 iOS - Product Usage: intended for use by patients with a compatible consumer mobile device.
As a result of the release of new software version to CareLink Personal website, the IOS app for the Continuous Monitoring Glucose system stopped automatically uploading data to the website.