Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2020-09-09FDA-DeviceCardinal Health 200, LLCClass IIPROTEXIS Latex Micro Surgical Gloves Size 9.0 - Product Usage: This surgeon s glove is a disposable device made of natural rubber latex or synthetic rubber intended to be worn by operating room personnel to be a barrier protection from contamination. Various thickness for sensitivity or durability depending on need.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
- 2020-09-09FDA-DeviceCardinal Health 200, LLCClass IIPROTEXIS Latex Micro Surgical Gloves Size 8.5 - Product Usage: This surgeon s glove is a disposable device made of natural rubber latex or synthetic rubber intended to be worn by operating room personnel to be a barrier protection from contamination. Various thickness for sensitivity or durability depending on need.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
- 2020-09-09FDA-DeviceCardinal Health 200, LLCClass IIPROTEXIS Latex Micro Surgical Gloves Size 8.0 - Product Usage: This surgeon s glove is a disposable device made of natural rubber latex or synthetic rubber intended to be worn by operating room personnel to be a barrier protection from contamination. Various thickness for sensitivity or durability depending on need.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
- 2020-09-09FDA-DeviceCardinal Health 200, LLCClass IIPROTEXIS Latex Micro Surgical Gloves Size 7.5 - Product Usage: This surgeon s glove is a disposable device made of natural rubber latex or synthetic rubber intended to be worn by operating room personnel to be a barrier protection from contamination. Various thickness for sensitivity or durability depending on need.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
- 2020-09-09FDA-DeviceCardinal Health 200, LLCClass IIPROTEXIS Latex Micro Surgical Gloves Size 7.0 - Product Usage: This surgeon s glove is a disposable device made of natural rubber latex or synthetic rubber intended to be worn by operating room personnel to be a barrier protection from contamination. Various thickness for sensitivity or durability depending on need.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
- 2020-09-09FDA-DeviceCardinal Health 200, LLCClass IIPROTEXIS Latex Micro Surgical Gloves Size 6.5 - Product Usage: This surgeon s glove is a disposable device made of natural rubber latex or synthetic rubber intended to be worn by operating room personnel to be a barrier protection from contamination. Various thickness for sensitivity or durability depending on need.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
- 2020-09-09FDA-DeviceCardinal Health 200, LLCClass IIPROTEXIS Latex Micro Surgical Gloves Size 6.0 - Product Usage: This surgeon s glove is a disposable device made of natural rubber latex or synthetic rubber intended to be worn by operating room personnel to be a barrier protection from contamination. Various thickness for sensitivity or durability depending on need.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
- 2020-09-09FDA-DeviceCardinal Health 200, LLCClass IIPROTEXIS Latex Micro Surgical Gloves Size 5.5 - Product Usage: This surgeon s glove is a disposable device made of natural rubber latex or synthetic rubber intended to be worn by operating room personnel to be a barrier protection from contamination. Various thickness for sensitivity or durability depending on need.
A degradation defect was found that could lead to holes and donning tears at the folding area around the cuffs.
- 2020-09-09FDA-DeviceC.A. Greiner & Sohne GesellschaftmbhClass IIPlastic Cannula HOLDEX Item # 450216 - Product Usage: - Product Usage: It is intended for use in routine specimen withdrawal from specimen collection bags or bottles through needleless cannula ports.
Sample leaking from the sleeve covering the needle inside the cannula
- 2020-09-09FDA-DeviceGe HealthcareClass IIThe OEC 9900 C-Arm is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical-care and emergency room procedures. The GE OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The system includes features
GEHC Surgery became aware that some users of the 9800 and 9900 fluoroscopic mobile C-Arms (referred to as the system here on) within the United States had requested that a Field Service Engineer (FSE) change the dose value on the live update screen from AKR / CAK to DAP. All systems sold to customers in the US that leave GEHC Surgery (the place of manufacture) are configured to display AKR / CAK when they leave the factory. An FSE changing the dose display, per the customer request, to display DAP instead does not comply to 21CFR 1020.32(k), which requires fluoroscopic equipment manufactured on or after June 10, 2006 to display continuous updates of AKR and CAK values.
- 2020-09-02FDA-DeviceNeurovision Medical Products IncClass IICobra x5 2-Ch EMG ET Tube, 8.0mm, REF: LTE700DCL-5, Rx Only, Sterile EO UDI: B006LTE700DCL52 - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
- 2020-09-02FDA-DeviceNeurovision Medical Products IncClass IICobra x5 2-Ch EMG ET Tube, 7.0mm, REF: LTE700DCM-5, Rx Only, Sterile EO UDI: B006LTE700DCM52 - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
- 2020-09-02FDA-DeviceNeurovision Medical Products IncClass IICobra x5 2-Ch EMG ET Tube, 6.0mm, REF: LTE700DCS-5, Rx Only, Sterile EO UDI: B006LTE700DCS52/$$7062618C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
- 2020-09-02FDA-DeviceNeurovision Medical Products IncClass IICobra x5 1-Ch EMG ET Tube, 8.0mm, REF: LTE700L-5, Rx Only, Sterile EO UDI: B006LTE700L52$$7061318B - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
- 2020-09-02FDA-DeviceNeurovision Medical Products IncClass IICobra x5 1-Ch EMG ET Tube, 7.0mm, REF LTE700M-5, Rx Only, Sterile EO CE UDI: B006LTE700M52/$$7061418C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
- 2020-09-02FDA-DeviceNeurovision Medical Products IncClass IICobra x5 1-CH EMG ET Tube, 6.0mm, REF: LTE700S-5, Rx Only, Sterile EO UDI: B006:TE700S52$$7050918B - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
- 2020-09-02FDA-DeviceNeurovision Medical Products IncClass IINerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMG ET Tube, 8.0 mm, REF NVTKIT-C8-5, Rx Only, Sterile UDIB006NVTKITC852/$$7062118C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
- 2020-09-02FDA-DeviceNeurovision Medical Products IncClass IINerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMT ET Tube, 7.0mm REF: NVTKIT-C7-5, RxOnly, Sterile, UDI: B006NvtKITc752/$$7062818C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
- 2020-09-02FDA-DeviceNeurovision Medical Products IncClass IINerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMG ET Tube, 6.0mm REF NVTKIT-C6-5, Rx Only, Sterile, UDI: B006NVTKITC652/$$7061918A - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.
The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.
- 2020-09-02FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayCare, device is stand-alone software - Product Usage: used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care.
When performing offline image review in RayCare 2C, RayCare 3A and RayCare 3B an offline image review task will guide you to the image viewer with the treatment image related to the fraction and the planning image selected. When starting the task, an error message might be displayed stating: Review files not available.
- 2020-09-02FDA-DeviceSiemens Medical Solutions USA, Inc.Class IIE.cam or Symbia systems that use foresight detectors - Product Usage: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.
E. CAM and Symbia system with foresight detectors performing gated or dynamic acquisition may lose some detector time information. Our risk analysis indicates that the probability of occurrence is remote. SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. CT: The CT component is intended to produce cross sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial place taken at different angles or spiral planes taken at different angles. SPECT+CT: Perform CT scans and nuclear imaging studies within the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.
- 2020-09-02FDA-DeviceKarl Storz EndoscopyClass IIFlexible Intubation Video Endoscope, Part: 11303BNX, with Instruction Manual: Z20827US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
- 2020-09-02FDA-DeviceKarl Storz EndoscopyClass IIFlexible Intubation Video Endoscope, Part: 11302BDX, with Instruction Manual: Z18443US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
- 2020-09-02FDA-DeviceKarl Storz EndoscopyClass IIFlexible DCI Intubation Fiberscope, Part: 11302BDD2, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
- 2020-09-02FDA-DeviceKarl Storz EndoscopyClass IIFlexible DCI Intubation Fiberscope, Part: 11301BND1, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
- 2020-09-02FDA-DeviceKarl Storz EndoscopyClass IIFlexible Intubation Fiberscope, Part: 11302BD2, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
- 2020-09-02FDA-DeviceKarl Storz EndoscopyClass IIFlexible Intubation Fiberscope, Part: 11301BN1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
- 2020-09-02FDA-DeviceKarl Storz EndoscopyClass IIFlexible Intubation Video Endoscope, Part: 11301BNX, with Instruction Manual: Z18444US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
- 2020-09-02FDA-DeviceKarl Storz EndoscopyClass IIFlexible Intubation Fiberscope, Part: 11301AA1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
- 2020-09-02FDA-DeviceBecton Dickinson & CompanyClass IIBD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack
Printed scale markings on the 3mL SafetyGlide syringes, with catalog # 305924, Lot# 8289599, are printed incorrectly. The scale is skewed to different degrees, resulting in missing and/or partially printed scale numbers and scale lines.