- 2020-07-08FDA-DeviceWestmed, Inc.Class IINeo-Vent Infant T-Piece Resuscitation Circuit With Neonatal Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0421, Qty:10, Length: 60" (1.5cm), Rx Only.
A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.
- 2020-07-08FDA-DeviceWestmed, Inc.Class IINeo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0420, Qty:10, Length: 60" (1.5cm), Rx Only.
A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.
- 2020-07-08FDA-DeviceWestmed, Inc.Class IINeo-Vent Infant T-Piece Resuscitation Circuit;Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0415, Qty:10, Length: 60" (1.5cm), Rx Only.
A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.
- 2020-07-08FDA-DeviceCodan US CorporationClass IICODAN SWAN-LOCK Swabable Needlefree Connector, Catalog #BC1000N, REF 16.7220, individually packaged, 504/case, Sterile. The firm name on the label is CODAN US CORPORATION, Santa Ana, CA.
Individual packages of connectors may have incomplete or bad seals which would compromise sterility.
- 2020-07-08FDA-DeviceArrow International IncClass IArrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- 2020-07-08FDA-DeviceArrow International IncClass IArrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/EMEA, Product Code IAP-0700
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- 2020-07-08FDA-DeviceArrow International IncClass IArrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE, Product Code IAP-0535
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- 2020-07-08FDA-DeviceArrow International IncClass IArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE DUTCH, Product Code IAP-0500NL
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- 2020-07-08FDA-DeviceArrow International IncClass IArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- 2020-07-08FDA-DeviceArrow International IncClass IArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE ITALIAN, Product Code IAP-0500I
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- 2020-07-08FDA-DeviceArrow International IncClass IArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE SPANISH, Product Code IAP-0500E
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- 2020-07-08FDA-DeviceArrow International IncClass IArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE, Product Code IAP-0500
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- 2020-07-08FDA-DeviceArrow International IncClass IArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 SPANISH, Product Code IAP-0400E
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- 2020-07-08FDA-DeviceArrow International IncClass IArrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product Code IAP-0400
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- 2020-07-01FDA-DeviceSiemens Healthcare Diagnostics IncClass IIRP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.
- 2020-07-01FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica CH Chemistry Calibrator
Calibration errors have been observed with the Total and Direct Bilirubin assays when using Chemistry Calibrator lot 534177 for ADVIA Chemistry and lot 534179 for Atellica CH.
- 2020-07-01FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIADVIA Chemistry Calibrator
Calibration errors have been observed with the Total and Direct Bilirubin assays when using Chemistry Calibrator lot 534177 for ADVIA Chemistry and lot 534179 for Atellica CH.
- 2020-07-01FDA-DeviceSuntech Medical, Inc.Class IIOscar 2, Model 250 System
The device display, when activated, can show an incorrect decimal point when the blood pressure reading is a three-digit number and the unit of measure is set for mmHg.
- 2020-07-01FDA-DevicePROVIDENCE MEDICAL TECHNOLOGIESClass IIAlly Bone Screw - Product Usage:Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation appropriate for the size of the device. Screws are intended for single use only. L Providence Medical Technology, Inc. 3875 Hopyard Road, Suite 300, Pleasanton, CA 94588, Rx Only.
There may be dimensional manufacturing error that represents a potential safety risk to patients.
- 2020-07-01FDA-DeviceArgon Medical Devices, IncClass IIBioPince Full Core Biopsy Instrument 16ga x 15cm.,Use with Optional CO-Axial Needle MCXS1615BP. QTY.: 5/CS - Product Usage: The disposable, automatic BioPince Full Core Biopsy Instrument is used to obtain multiple core samples from soft tissue such as liver, kidney, abdominal masses, etc.
Incorrect product labeling.
- 2020-07-01FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation standalone software treatment planning system (RayStation 8A, 8B, 9A, and 9B, including all service packs) model no. 8.0, 8.1, 9.0, 9.1, 9.2
It is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. In addition, there is an issue found with the display of planned and recorded Gantry Angle in the delivered beam list in RayTreat.
- 2020-07-01FDA-DeviceZest Anchors LLCClass IILOCATOR OVERDENTURE IMPLANT SYSTEM STERILE R Rx Only QTY 1 .
The abutment provided with the Implant System may be out of specification, which could lead to a small gap occurring between the implant and abutment on assembly in the patient.
- 2020-07-01FDA-DeviceGE Healthcare, LLCClass IICARESCAPE Patient Data Module (PDM) used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro, Model Numbers: 2042084-001 2094504-001 2042084-001-046050 2042084-001-208621 2042084-001-045915 2042084-001-215164 2042084-001-078474 2042084-001-053034 2042084-001-215165 2042084-001-079462 2042084-001-047389 2042084-001-112198 2042084-001-041673 2042084-001-227238 2042084-001-071564 2042084-001-503824 2042084-001-119839 2042084-001-569269 2042084-0
The CARESCAPE Patient Data Module (PDM) is used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro. If Pace detection is turned ON, on the monitor, and if an Automated External Defibrillator (AED) is used to perform defibrillation, the low frequency signals deployed by the AED, cause the PDM to issue a false flat line ECG waveform and alarm Asystole to be displayed on the monitor.
- 2020-07-01FDA-DeviceTosoh Bioscience IncClass IIAIA-900 Analyzer, Part no. 022930, UDI 04560189283992
The Substrate Monitoring System in the AIA-900 Analyzer did not detect an empty substrate bottle. The issue may lead to erroneous immunoassay test results.
- 2020-07-01FDA-DeviceColtene Whaledent IncClass IIHYGENIC Wingless Fiesta Kit, Model no. H02790 - Product Usage: The dental dam forceps are intended to be used in conjunction with the rubber dental dam and the dental dam clamps. The forceps are used to place and remove the dental dam clamps from the rubber dental dam.
Multiple Coltene batches of the product have a pin size larger than acceptable, which may make it difficult for the forceps to fit appropriately with dental dam clamps.
- 2020-07-01FDA-DeviceColtene Whaledent IncClass IIHYGENIC Winged Fiesta Kit, Model no. H02778 - Product Usage: The dental dam forceps are intended to be used in conjunction with the rubber dental dam and the dental dam clamps. The forceps are used to place and remove the dental dam clamps from the rubber dental dam.
Multiple Coltene batches of the product have a pin size larger than acceptable, which may make it difficult for the forceps to fit appropriately with dental dam clamps.
- 2020-07-01FDA-DeviceColtene Whaledent IncClass IIHYGENIC Simple Dam Kit, Model no. 60019066 - Product Usage: The dental dam forceps are intended to be used in conjunction with the rubber dental dam and the dental dam clamps. The forceps are used to place and remove the dental dam clamps from the rubber dental dam.
Multiple Coltene batches of the product have a pin size larger than acceptable, which may make it difficult for the forceps to fit appropriately with dental dam clamps.
- 2020-07-01FDA-DeviceColtene Whaledent IncClass IIHYGENIC Dental Dam Forceps, Model no. H01262 - Product Usage: The dental dam forceps are intended to be used in conjunction with the rubber dental dam and the dental dam clamps. The forceps are used to place and remove the dental dam clamps from the rubber dental dam.
Multiple Coltene batches of the product have a pin size larger than acceptable, which may make it difficult for the forceps to fit appropriately with dental dam clamps.
- 2020-07-01FDA-DeviceHocoma AGClass IILokomat Pro, Catalog Numbers 30003 LokomatPro L6.2 120V AD ; 30004 LokomatPro L6.2 120V Kombi; 30005 LokomatPro L6.2 120V PE; 30683 LokomatPro L6.2 110V FD PE; 30684 LokomatPro L6.2 110V FD Kombi; 32001 LokomatPro L6.2 110V FD AD Product Usage: The intended use of the Lokomat is to support treadmill training to treat patients with walking disabilities caused by neurological, muscular or bone-related disorders. isokinetic testing and evaluation system
The motor controller may fail, which can potentially lead to an error in functionality of the Body Weights Support Rope, which means that the rope can go up or down in an uncontrolled way.
- 2020-07-01FDA-DeviceUtak Laboratories IncClass IIIUtak NR Trace Elements Serum Control, 5 ml, REF 66816. The responsible firm on the label is UTAK LABORATORIES, INC., Valencia, CA. Control for in vitro diagnostics.
The product has an incorrect reference value for cadmium and an incorrect expected range.