- 2020-05-20FDA-DeviceMerit Medical Systems, Inc.Class IICustom Sheath Introducer Kit REF K21-00031 - Product Usage: intended use is to provide access and facilitate the percutaneous introduction of various devices into vein.
labeling error: Due to a manufacturing issue, product package was incorrectly labeled with the incorrect sheath size and the incorrect sheath assembly was included in the product packaging. The use of the incorrect kit components may result in vasoconstriction, moderate hemorrhage, and/or embolism.
- 2020-05-20FDA-DeviceCardiac Assist, IncClass IITandemHeart pump is assembled into kits: TandemLife Kit Product: 5740-0000 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
Failure to prime due to an assembly error
- 2020-05-20FDA-DeviceCardiac Assist, IncClass IITandemHeart pump is assembled into kits: TandemLung Kit - DL31, Product: 5730-3118 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
Failure to prime due to an assembly error
- 2020-05-20FDA-DeviceCardiac Assist, IncClass IITandemHeart pump is assembled into kits: ProtekDuo Kit - DL31 RD, Product: 5720-3631 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
Failure to prime due to an assembly error
- 2020-05-20FDA-DeviceCardiac Assist, IncClass IITandemHeart pump is assembled into kits: ProtekDuo Kit - DL3, Product: 5720-3118 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
Failure to prime due to an assembly error
- 2020-05-20FDA-DeviceCardiac Assist, IncClass IITandemHeart pump is assembled into kits: ProtekDuo Kit - DL29, Product: 5720-2916 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
Failure to prime due to an assembly error
- 2020-05-20FDA-DeviceCardiac Assist, IncClass IITandemHeart pump is assembled into kits: TandemHeart Kit - TS62/A17, Product: 5710-6217 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
Failure to prime due to an assembly error
- 2020-05-20FDA-DeviceCardiac Assist, IncClass IITandemHeart pump is assembled into kits: TandemHeart Kit - TS62/A15, Product: 5710-6215 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
Failure to prime due to an assembly error
- 2020-05-20FDA-DeviceCardiac Assist, IncClass IITandemHeart Pump Kit, Product: 5120-0000 - Product Usage: provides temporary extracorporeal circulatory system support with the unique ability to locally deliver an anticoagulant through the infusion system.
Failure to prime due to an assembly error
- 2020-05-20FDA-DeviceSunrise Medical (US) LLCClass IIQuickie/Zippie powered wheelchair. Q700M wheelchair. A battery operated powered wheelchair, which is controlled by Power Module RNET 120A Expandable (Item No.: 250ER10, Part No.: 123966).
Due to programming errors in the Controller, an increase in the set motor parameters can be made by the end user beyond tested safe limits.
- 2020-05-20FDA-DeviceBecton Dickinson & CompanyClass IIPowerGlide ST Midline Catheter Maximal Barrier Kit-18 gauge,10 cm length Catalog Number: ST018101D
Iincorrectly packaged without the required lidocaine as labeled
- 2020-05-20FDA-DeviceBeckman Coulter Inc.Class IIiQ200 Series Urine Microscopy Analyzer- All part numbers
A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.
- 2020-05-20FDA-DeviceBeckman Coulter Inc.Class IIiChem¿VELOCITY Urine Chemistry System-
A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.
- 2020-05-20FDA-DeviceHandicare ABClass IIEVA Floor Lifts, Model Numbers: 400, 450, 600
After feedback from customers and internal suspicion, Handicare has become aware that abnormal use of Eva-lifts could lead to a deterioation in the joint of the liftarm. This could lead to injury if the joint breaks. All lifts must therefore be checked and reported back to the manuafacturer.
- 2020-05-20FDA-DeviceKarl Storz EndoscopyClass IIIInsufflation tubing to be used with varies insufflator units. REF 031200-10; REF 031322-10; REF 031532-10; REF 031622-10
The sterile single-use tubing may contain higher ethylene oxide (EO) residues than the acceptable level.
- 2020-05-20FDA-DeviceMedtronic Sofamor Danek USA IncClass IIVERTEX SELECT MULTI AXIAL SCREW. 4.0 X 38MM MULTI AXIAL SCREW. Lot H5517628
Nonconforming product; length of the product measured shorter than the labeled length.
- 2020-05-20FDA-DeviceCook Inc.Class IIFormula 418 Renal Balloon-Expandable Stent, RPN FOR418-18-80-6-12, Order Number (GPN) G56611 Balloon-expandable stent made of 316L stainless steel with a slotted tube configuration. It is pre-mounted on an over-the-wire balloon catheter, which serves as the delivery system. The stent is positioned between two radiopaque marker bands, which are located inside the balloon at the proximal and distal tapers of the balloon.
The firm received complaints on five devices from one specific lot where the stent dislodged from the balloon delivery system prior to deployment. Use of affected product may lead to increased procedural time and additional intervention to remove a stent that dislodged within an unintended location.
- 2020-05-20FDA-DeviceRemel Inc.Class IISWIN 2017 database, Model YSW4000
Potential for out of range microbial results
- 2020-05-20FDA-DeviceArgon Medical Devices, IncClass IIPAC Tray/5 Catheter Introducer Tray 8F, Model 497424, Rx Only, STERILE EO, UDI: 00886333214440 - Product Usage: The introducers are indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices. Curved tip Introducer allows for contralateral access to the iliac artery and selected vasculature. Case Label: PAC Tray/5 Catheter Introducer Tray 8F, QTY 10/Bx, Rx Only, STERILE EO
Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.
- 2020-05-20FDA-DeviceArgon Medical Devices, IncClass IIArterial Line Kit 20ga X 3", Model 498106, Rx Only, Sterile EO, UDI: 00886333215577 - Product Usage: Arterial Catheters are indicated for use when in instances where continuous invasive blood pressure readings and/or frequent blood gas measurements are required. Case Label: Arterial Line Kit/1 20ga x 3", Qty: 10/CS, CE 2797
Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.
- 2020-05-20FDA-DeviceArgon Medical Devices, IncClass IIArterial Line Kit 20ga X 3", Model 498411B, Rx Only, Sterile EO, UDI: 00886333215652 - Product Usage: Arterial Catheters are indicated for use when in instances where continuous invasive blood pressure readings and/or frequent blood gas measurements are required. Case Label: Arterial Line Kit 20ga x 3", Ref 498411B, Qty: 10/CS
Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.
- 2020-05-20FDA-DeviceStryker Medical Division of Stryker CorporationClass IIStryker Power-PRO TL cot-powered ambulance cot Model Number 6550 - Product Usage: is a powered wheeled stretcher that is intended to support and transport the entire body of a traumatized, ambulatory or non-ambulatory human patient (includes infants and adults).
Ambulance Cots may not meet crash test standard BS EN 1789
- 2020-05-20FDA-DeviceCuattro LLCClass IICuattroDR System Software, REF: SWR-00069-MD, UDI: B269SWR00069MD0
There is a potential that an image from a previous patient study to show up in a later patient study. This may result in misdiagnosis of the correct patient.
- 2020-05-20FDA-DevicePhilips North America, LLCClass IIPhilips Zenition 70, System code 718133 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.
When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.
- 2020-05-20FDA-DevicePhilips North America, LLCClass IIPhilips Zenition 50, System code 718096 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.
When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.
- 2020-05-20FDA-DeviceOrtho-Clinical DiagnosticsClass IIVITROS Chemistry Products CKMB Slides, Creatine Kinase MB, Product Code 8001133-90 slides, UDI Number 10758750004201 - Product Usage: The VITROS Chemistry Product Slides are used to measure creatine kinase MB (CK-MB) activity in serum which used to be ordered to detect Myocardial Infarction.
The Spanish and Swedish language versions of the IFU state that the assay can be tested with Plasma which is false and could result in biased results.
- 2020-05-20FDA-DeviceOrtho-Clinical DiagnosticsClass IIVITROS Chemistry Products CKMB Slides, Creatine Kinase MB, Product Code 805 8232-300 slides, UDI Number 10758750004294 - Product Usage: The VITROS Chemistry Product Slides are used to measure creatine kinase MB (CK-MB) activity in serum which used to be ordered to detect Myocardial Infarction.
The Spanish and Swedish language versions of the IFU state that the assay can be tested with Plasma which is false and could result in biased results.
- 2020-05-20FDA-DeviceEncore Medical, LPClass IIEMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029 803-15-030 Product Usage: The EMPOWR Acetabular Impactor Balls is intended for use in joint replacement in patients suffering from disability due to noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; rheumatoid arthritis, correction of functional deformity, femoral fracture, and in the salvage of previously fai
The firm received complaints that the EMPOWR Acetabular Impactor Balls were fracturing/cracking during impaction which could cause debris to migrate or be left in the joint space.
- 2020-05-20FDA-DeviceEdwards Lifesciences, LLCClass IIOxymetry Cable, Rxonly. REF/UDI: HEMOXSC100/(01)00690103196986 - Product Usage: The HemoSphere oximetry cable is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure venous oxygen saturation.
Intermittent communication due to inconsistent crimping of connectors onto cables.
- 2020-05-13FDA-DeviceICU Medical, Inc.Class IIThe Cogent Hemodynamic Monitoring System. REF58400-000, GTIN 00840619099459; REF 58400-000R, GTIN 0087709094314; REF 58403, GTIN 00840619079505; REF 58401, GTIN 00840619079499 - Product Usage: intended for patients for whom the monitoring of continuous cardiac output and calculated hemodynamic parameters is indicated for diagnostic and prognostic evaluation by a clinician. The target population includes Critical Care Patients, Trauma Patients, and Cardiac Surgery Patients.
Firm identified software issues which leads to the patient ID and patient information not being stored in the system.