Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2020-04-01FDA-DeviceLivaNova USA IncClass IIVNS Therapy Programmer, Model 3000 used with Models 103, 104, 105, 106, or 8103 VNS Therapy generators.
False positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and widen pulse width. Only system diagnostic testing evaluates output current. Users may conclude device malfunction, could lead to medical/surgical intervention.
- 2020-04-01FDA-DeviceDexcom IncClass IIDexcom G6 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.
The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.
- 2020-04-01FDA-DeviceDexcom IncClass IIDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.
The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.
- 2020-04-01FDA-DeviceDexcom IncClass IIDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.
The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.
- 2020-04-01FDA-DeviceIntellijoint Surgical, Inc.Class IIIntellijoint Navigation System, Software Application, IJK1 (P/N 000-0547) - Product Usage: The Intellijoint Navigation System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified.
During total knee arthroplasty (TKA), the software may incorrectly calculate femur resection depth.
- 2020-04-01FDA-DeviceAbbott Diabetes Care, Inc.Class IIFreeStyle Precision H Blood Glucose and Ketone Monitoring System, UDI: 00093815711810
Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate meter disinfection or improper handling of contaminated meters, which may cause infection or exacerbation of an existing infection.
- 2020-04-01FDA-DeviceAbbott Diabetes Care, Inc.Class IIOptium EZ Blood Glucose Monitoring System, UDI: 00093815710653
Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate meter disinfection or improper handling of contaminated meters, which may cause infection or exacerbation of an existing infection.
- 2020-04-01FDA-DeviceAbbott Diabetes Care, Inc.Class IIPrecision Xtra Blood Glucose & Ketone Monitoring System, UDI: 00093815715023, 00093815988144
Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate meter disinfection or improper handling of contaminated meters, which may cause infection or exacerbation of an existing infection.
- 2020-04-01FDA-DeviceBoston Scientific CorporationClass IImager II 5F Angiographic Catheters, 5 units per package.
Potential for tip detachment of Imager II 5F Angiographic Catheters
- 2020-04-01FDA-DeviceHill-Rom Inc.Class IINaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety) Software is compatible with the following beds: 1. VersaCare Revision A-J with upgrade kit Revision K with UCB 1.18 2. Centrella (all versions) 3. Progressa (firmware version 1.15.0.0)
There is a software issue which may result in failure to monitor or control patient's bed exit mechanism.
- 2020-04-01FDA-DeviceInsulet CorporationClass IIOmnipod DASH Personal Diabetes Manager (PDM), Catalog numbers US: 18239 M/D: PDM-USA1-D001-MG-USA1 EU: PT-000010 - M/D: INT1-D001-MG PT-000011 M/D: INT1-D001-MM UDI 10385082000009 The Omnipod DASH Personal Diabetes Manager (PDM) is a mobile, handheld Android device that is intended to communicate with the Omnipod DASH Pod (pump).
In certain scenarios, the Omnipod DASH PDM may suggest an inaccurate bolus amount based on a blood glucose value that is more than 10 minutes old when the user does not exit the bolus calculator as designed or when a system alarm interrupts a bolus calculation. If the user delivers the bolus, this may lead to hypoglycemia or hyperglycemia.
- 2020-04-01FDA-DeviceQUIDEL CARDIOVASCULAR INCClass IIQuidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com
Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.
- 2020-04-01FDA-DeviceCarroll-Baccari, Inc.Class IIWave Prep Tubes (1710-03) and single use cups (17--00-24)
Samples of 114gm tubes of Lemon Prep, collected during a Food and Drug Administration inspection on October 15, 2019 were tested and found to be contaminated with Burkholderia cepacia. Mavidon is recalling all products manufactured at their facility.
- 2020-04-01FDA-DeviceCarroll-Baccari, Inc.Class IIMVAP MEDICAL SUPPLIES INC, Mavidon CardioPrep Single, 24 use cups, Re order # MCP-24
Samples of 114gm tubes of Lemon Prep, collected during a Food and Drug Administration inspection on October 15, 2019 were tested and found to be contaminated with Burkholderia cepacia. Mavidon is recalling all products manufactured at their facility.
- 2020-04-01FDA-DeviceCarroll-Baccari, Inc.Class IIPediaPrep Tubes (MD0033-T) and Single use cups (MD0033-SUP)
Samples of 114gm tubes of Lemon Prep, collected during a Food and Drug Administration inspection on October 15, 2019 were tested and found to be contaminated with Burkholderia cepacia. Mavidon is recalling all products manufactured at their facility.
- 2020-04-01FDA-DeviceCarroll-Baccari, Inc.Class IILemonPrep Tubes (MD0019-T) and Single use cups (MD0019-SUP)
Samples of 114gm tubes of Lemon Prep, collected during a Food and Drug Administration inspection on October 15, 2019 were tested and found to be contaminated with Burkholderia cepacia. Mavidon is recalling all products manufactured at their facility.
- 2020-04-01FDA-DeviceRandox Laboratories Ltd.Class IIChemistry QC Premium Plus, Cat. No. LAL4213, GTIN 05055273209006 - Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic assays. The device is for the control of accuracy.
The control target and range value for Sodium using the ISE indirect method has been assigned incorrectly.
- 2020-04-01FDA-DeviceMedline Industries IncClass IIMedline Sterile 3mL Syringes with 21GA x 1 Inch Needle 80 each REF DYND3ML21G1A lot 18JBH655 (01) 1088438982153.
Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging.
- 2020-04-01FDA-DeviceMedline Industries IncClass IIMedline Winged Sponges Presaturated W/PVP REF DYND70289 DYND70289H Expiration 06/30/2022.
Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging.
- 2020-04-01FDA-DeviceMedline Industries IncClass IIMedline Stick Sponge Presaturated W/PVP REF DYND70288 DYND70288H 2/PK Expiration 06/30/2022.
Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging.
- 2020-04-01FDA-DeviceBiomet, Inc.Class IIComprehensive Shoulder System Mini Humeral Stem 9mm,, Model No. 113629 - Product Usage: The intended use of the comprehensive humeral stem mini is to act as the humeral component of the comprehensive total shoulder system.
A 15mm Shoulder Mini Humeral Stem was misidentified as a 9mm Shoulder Mini Humeral Stem.
- 2020-04-01FDA-DeviceT.A.G. MEDICAL PRODUCTS CORPORATION, LTD.Class IIFlipCutter III Drill
There is potential for the blister tray to be cracked causing the sterile barrier to be compromised and potential for particulate within the sterile barrier.
- 2020-04-01FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIDimension Vista¿ Total Bilirubin Flex¿ reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10445146 Product Usage: Dimension Vista TBIL method is an in vitro diagnostic test intended to quantitatively measure total bilirubin in human serum and plasma on the Dimension Vista¿ System. Measurements of total bilirubin are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder disease
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
- 2020-04-01FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIDimension Total Bilirubin Flex reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10444957 Product Usage: Dimension TBI method for the ¿ clinical chemistry syst6em is an in vitro diagnostic test intended to quantitatively measure total bilirubin in human serum and plasma. Dimension Measurements of total bilirubin are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder diseas
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
- 2020-04-01FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIADVIA Chemistry Total Bilirubin_2 assay- SMN: 10341115 (40 mL) & 10341113 (70 mL). Product Usage: ADVIA TBIL_2 is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates on the ADVIA¿ Chemistry systems. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disord
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
- 2020-04-01FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIADVIA Chemistry Direct Bilirubin_2 assay-SMN: 10316610 (20 mL) & 10341114 (70 mL) - Product Usage: ADVIA DBIL_ 2 is for in vitro diagnostic use in the quantitative determination of direct bilirubin in human serum and plasma on the ADVIA¿ Chemistry systems. Such measurements are useful in the screening of liver function disorders or in the diagnosis of jaundice. In vitro Diagnostic for the determination of direct bilirubin.
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
- 2020-04-01FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica CH Total Bilirubin_2 assay-In vitro Diagnostic for the determination of total bilirubin SMN: 11097531 - Product Usage: Atellica TBil_2 assay is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma (lithium heparin) of adults and neonates using the Atellica¿ CH Analyzer. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
- 2020-04-01FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica CH Direct Bilirubin_2 assay In vitro Diagnostic for the determination of direct bilirubin SMN: 11097532 - Product Usage: ADVIA DBIL_ 2 is for in vitro diagnostic use in the quantitative determination of direct bilirubin in human serum and plasma on the ADVIA¿ Chemistry systems. Such measurements are useful in the screening of liver function disorders or in the diagnosis of jaundice.
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
- 2020-04-01FDA-DeviceBeckman Coulter, Inc.Class IIiChemVELOCITY Strips, Catalog 800-7212
The Leukocyte and Compensation pads were switched on some vials of lot number 7212269M of the iChem VELOCITY Strips.
- 2020-04-01FDA-DevicebioMerieux, Inc.Class IIVITEK 2 card AST-N371 - Product Usage: The VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.
The top seal of some of the pouches was compromised which can allow moisture to enter that can impact some antibiotics on the card.