Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2019-12-18FDA-DeviceCook Inc.Class ICrossCath Support Catheter, RPN CXC3.4-2.2-18-135-P-NS-0, Order Number G51583 For diagnostic and interventional vascular procedures.
A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).
- 2019-12-18FDA-DeviceCook Inc.Class ICrossCath Support Catheter, RPN CXC3.4-2.2-18-90-P-NS-0, Order Number G51575 For diagnostic and interventional vascular procedures.
A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).
- 2019-12-18FDA-DeviceCook Inc.Class ICrossCath Support Catheter, RPN CXC3.0-1.9-14-90-P-NS-0, Order Number G51572 For diagnostic and interventional vascular procedures.
A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).
- 2019-12-11FDA-DeviceLens.comClass IIAlcon Air Optix Night & Day Aqua (lotrafilcon A), breathable contact lenses, PWR - 1.50 BC 8.6 DIA 13.8, Rx, Sterile. The firm name on the label is Alcon Laboratories, Inc., Fort Worth, TX.
The product may be counterfeit and could have microbial issues.
- 2019-12-11FDA-DevicePhilips North America, LLCClass IIHeartStart XL+ Defibrillator/Monitor, Model 861290
Device may fail to turn on or unexpectedly attempt to restart, rendering it unable to return to a ready for use state. Additionally, this issue may occur when the device is in standby mode, when attempting to power on to run a self-test. If this occurs, the device will indicate that it is not ready for use.
- 2019-12-11FDA-DeviceTheralase Inc.Class IIITLC-2000 Therapeutic Medical Laser System, Power Pack component TLC-2002 Used for the temporary relief of knee pain.
Device was shipped without a Unique Device Identification (UDI) label and a separate label identifies the power pack as CE Marked when the device has not yet received the CE Mark.
- 2019-12-11FDA-DeviceTheralase Inc.Class IIITLC-2000 Therapeutic Medical Laser System, used for the temporary relief of knee pain.
Device was shipped without a Unique Device Identification (UDI) label.
- 2019-12-11FDA-DeviceSteris CorporationClass IIEagle 3000 Sterilizer, small steam sterilizer. Product Usage: The Eagle 3000 Small Sterilizer is designed for sterilization of certain materials used in healthcare facilities, and are available in two configurations: Gravity designed for sterilization of non-porous, heat- and moisture stabile goods, and sterilization of items intended for immediate use. Prevacuum designed for fast, efficient sterilization of porous, heat- and moisture-stabile materials, in addition to the same sterilization
Potential for the incorrect pressure safety valve being included in the Preventive Maintenance (PM) kit for the 3000 Small Sterilizers.
- 2019-12-11FDA-DeviceMiami Device Solutions, LLC.Class IIProximal Humerus Plate, 8 holes, Right, REF MDS130108R - Product Usage: The Proximal Humerus Plating System consists of plates, screws, and locking caps intended to provide internal fixation of fractures of the proximal humerus. The polyaxial screws have a range of motion of +/- 20¿. The screws can be simply lagged or locked to the plate with the addition of a locking cap. The plates are side specific implants.
Right Proximal Humerus Plates. Product component parts MDS130108R, may be improperly identified as left.
- 2019-12-11FDA-DeviceCooperSurgical, Inc.Class IIQuantam 2000 Electrosurgical Generator, Part number 909075-05W.
Potential for the Quantum 2000 Electrosurgery Generator, part numbers 909075, 909075-05 and 909075-05W, to cease to either cut, coagulate, or blend during a procedure when the foot pedal is used for activation.
- 2019-12-11FDA-DeviceCooperSurgical, Inc.Class IIQuantam 2000 Electrosurgical Generator, Part number 909075-05.
Potential for the Quantum 2000 Electrosurgery Generator, part numbers 909075, 909075-05 and 909075-05W, to cease to either cut, coagulate, or blend during a procedure when the foot pedal is used for activation.
- 2019-12-11FDA-DeviceCooperSurgical, Inc.Class IIQuantam 2000 Electrosurgical Generator, Part number 909075.
Potential for the Quantum 2000 Electrosurgery Generator, part numbers 909075, 909075-05 and 909075-05W, to cease to either cut, coagulate, or blend during a procedure when the foot pedal is used for activation.
- 2019-12-11FDA-DeviceToby Orthopaedics, Inc.Class IIPANTERA Proximal Humerus Fracture Plate System
Potential suture clip breaking during a humerus fixation surgery.
- 2019-12-11FDA-DeviceAllergan PLCClass IIXEN Gel Stent, REF 5513-001(US Model), Sterile, RX Only, (01)10888628032439 ; Other Model Numbers: 5507-001(Global Model) and 5517-001(Australia model) Implant for glaucoma treatment.
Residual polishing compounds, that are used in the needle sleeve manufacturing process, may be present on the injector needle. The residual polishing compounds could transfer to the patient's eye during procedure possibly resulting in irritation, inflammation, local allergic reaction/ hypersensitivity, iritis, uveitis/sterile endophthalmitis or an intraocular foreign body.
- 2019-12-11FDA-DeviceNDDd Medical Technologies, Inc.Class IIEasyOne Pro LAB VOS Respiratory Analysis System Pulmonary Function Testing Device, Model 3100-1
The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lower electromagnetic shielding provided by the cable harness, the CO/CO2 sensor could have been influenced by high frequencies emitted by very close devices.
- 2019-12-11FDA-DeviceNDDd Medical Technologies, Inc.Class IIEasyOne Pro Respiratory Analysis System Pulmonary Function Testing Device, Model 3000-1
The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lower electromagnetic shielding provided by the cable harness, the CO/CO2 sensor could have been influenced by high frequencies emitted by very close devices.
- 2019-12-11FDA-DeviceZimmer Biomet, Inc.Class IIBiomet Orthopedics Intramedullary Bone Saw Cam Assembly 14MM- Orthopedic manual surgical instrument Item Number: 475665
Possible commingle of 13 MM and 14 MM Intramedullary Bone Saw Cam Assembly -an Intramedullary Bone Saw Cam Assembly was labeled and etched as a 14 MM but measured as a 13 MM cam
- 2019-12-11FDA-DeviceZimmer Biomet, Inc.Class IIBiomet Orthopedics Intramedullary Bone Saw Cam Assembly 13MM-Orthopedic manual surgical instrument Item Number: 475660
Possible commingle of 13 MM and 14 MM Intramedullary Bone Saw Cam Assembly -an Intramedullary Bone Saw Cam Assembly was labeled and etched as a 14 MM but measured as a 13 MM cam
- 2019-12-11FDA-DeviceStryker Instruments Div. of Stryker CorporationClass IIColor Cuff Disposable Tourniquet Cuff, 18 inch X 3 inch, 1BLA (Single Bladder), 2PRT(Dual Port) Quick Connect Connector, Product Number 5921-018-235 - Product Usage: The Stryker Disposable Tourniquet Cuff (cuff) is indicated for use in surgical procedures that require the temporary occlusion of blood flow in a patients extremities during surgical procedures to produce greater visualization of the operative field.
There is a potential for the cuff to leak air leading to a loss of vascular occlusion.
- 2019-12-11FDA-DeviceStryker Instruments Div. of Stryker CorporationClass IIColor Cuff Disposable Tourniquet Cuff, 18 inch X 3 inch, 1BLA (Single Bladder), 1PRT(Single Port) Quick Connect Connector, Product Number 5921-018-135 - Product Usage: The Stryker Disposable Tourniquet Cuff (cuff) is indicated for use in surgical procedures that require the temporary occlusion of blood flow in a patients extremities during surgical procedures to produce greater visualization of the operative field.
There is a potential for the cuff to leak air leading to a loss of vascular occlusion.
- 2019-12-11FDA-DeviceBrasseler USA I LpClass IIBRASSELER KM166-310-35 Twist Drill Quick Connect Diameter 3.5 mm, Total Length 112mm, Effective Length, 86mm - Product Usage: Twist drill intended use: to drill into bone in hip or knee surgeries.
One lot of 3.5 mm twist drills were incorrectly laser etched as 2.5 mm the shaft of the device.
- 2019-12-11FDA-DeviceSteris CorporationClass IIEvolution Loading Car and Transfer Carriage- Accessory to the Sterilizer, Steam
The locking pin of the front wheel asse mbly of the Evolution Transfer Carriages may not remain full y engaged with the transfer carriage and result in the transfer carriage and loading car unexpectedly falling forward
- 2019-12-11FDA-DeviceGE Healthcare, LLCClass IGiraffe Incubator infant warmer
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
- 2019-12-11FDA-DeviceGE Healthcare, LLCClass IGiraffe Incubator Carestation, Giraffe Incubator Carestation CS1 infant warmer
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
- 2019-12-11FDA-DeviceGE Healthcare, LLCClass IGiraffe OmniBed Carestation, Model Numbers: 2082844-001-XXXXXX; Giraffe OmniBed Carestation CS1, Model Numbers: 2082844-001-XXXXXXXX. infant warmer
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
- 2019-12-11FDA-DeviceGE Healthcare, LLCClass IGiraffe OmniBed infant warmer
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
- 2019-12-11FDA-DeviceMedtronic NeuromodulationClass ISynchromed II programmable pump, Model numbers 8637-20 and 8637-40
There is a potential for the presence of a foreign particle inside the pump motor assembly which could interfere with the motor gear rotation and lead to a permanent motor stall.
- 2019-12-04FDA-DeviceVygon MFG, Inc., dba/ Churchill Medical Systems, Inc.Class II4Fr x 20cm Single Lumen CT Midline Basic Tray, Product Code: VYML4S1001
Potential that 3Fr size catheter placed in the catheter tray instead of 4Fr for the 4Fr x 20cm Single Lumen CT Midline Basic Tray.
- 2019-12-04FDA-DeviceGN Hearing A/SClass IIIn-the-Ear hearing aids, Model Numbers: T9ITC-DW-LP, FT6ITC-D-MP, FT6ITE-D-MP, FT8CIC-MP, FT8ITC-W-HP, FT8ITC-W-LP, FT8ITC-W-MP, FT8ITE-DW-HP, FT8MIH-W-HP, FT8MIH-W-MP, LT5CIC-MP, LT5CIC-UP, LT5IIC-LP, LT5ITC-DW-MP, LT5ITC-DW-UP, LT5ITE-DW-HP, LT5ITE-DW-MP, LT5ITE-DW-UP, LT5MIH-W-HP, LT5MIH-W-MP, LT5MIH-W-UP, LT7CIC-HP, LT7CIC-LP, LT7CIC-MP, LT7CIC-W-MP, LT7IIC-LP, LT7ITC-DW-LP, LT7ITC-DW-MP, LT7ITC-MP, LT7ITC-W-HP, LT7ITC-W-MP, LT7ITE-DW-HP, LT7ITE-DW-MP, LT7ITE-DW-UP, LT7ITE-W-UP, LT7MIH-W-H
GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.
- 2019-12-04FDA-DeviceGN Hearing A/SClass IIBehind-the-Ear hearing aids, Model Numbers: Q962-DRW, T561-DRW, T562-DRW, T762-DRW, T777-DW, AMZ964-DRW, ET588-DWH, ET598-DW, ET788-DWH, ET798-DW, ET988-DWH, ET998-DW, FT861-DRW, FT862-DRW, RE561-DRWC, RE762-DRW, RE961-DRWC, RE962-DRW, PZ862-DRW, LT561-DRW, LT561-DRWZ, LT562-DRW, LT588-DW, LT761-DRW, LT761-DRWZ, LT762-DRW, LT777-DW, LT788-DW, LT961-DRW, LT961-DRWZ, LT962-DRW, LT967-DW, LT977-DW, LT988-DW, TST1763-DRW, TST1764-DRW, TST663-DRW, TST963-DRW, TST964-DRW, VA861-DRW, VA861-DRWZ,
GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.