- 2019-10-16FDA-DeviceBecton Dickinson & CompanyClass IIBD Microtainer Tubes with FE (Sodium Fluoride / Disodium EDTA) Glycolytic Inhibitor Part/Catalog Number: 365992
Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.
- 2019-10-16FDA-DeviceBecton Dickinson & CompanyClass IIBD Microtainer PST Tubewith LH (Lithium Heparin) - Amber Part/Catalog Number: 365987
Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.
- 2019-10-16FDA-DeviceBecton Dickinson & CompanyClass IIBD Microtainer PSTTubes with LH (Lithium Heparin) Part/Catalog Number:365985
Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.
- 2019-10-16FDA-DeviceBecton Dickinson & CompanyClass IIBD Microtainer SST- Amber Part/Catalog Number: 365978
Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.
- 2019-10-16FDA-DeviceBecton Dickinson & CompanyClass IIBD MicrotainerTubes with K2E (K2EDTA) Catalog Number: 365974
Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.
- 2019-10-16FDA-DeviceBecton Dickinson & CompanyClass IIBD Microtainer SST Part/Catalog No.365967
Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.
- 2019-10-16FDA-DeviceBecton Dickinson & CompanyClass IIBD Microtainer Tubes with LH (Lithium Heparin) Part/Catalog No.365965
Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.
- 2019-10-16FDA-DeviceBecton Dickinson & CompanyClass IIBD Microtainer Z (No Additive Tubes) Part/Catalog No.365963
Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.
- 2019-10-16FDA-DeviceArrow International IncClass IIPediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, REF AK-25502 Product Usage; Provide short-term (< 30 days) central venous access for treatment of diseases or conditions requiring central venous access
Lidstock contains a labeling error. The lidstock shows the catheter cross section of the gauge (GA) values in a reversed manner.
- 2019-10-16FDA-DeviceK2M, IncClass IIK2M Special Connector Instruments (Special Connector, Special Adaptor, and Quick Adaptor) Product Usage: Special Connector Instruments include an additional feature on their shafts that enable surgeon users to combine them with non-K2M navigation systems as part of their standard surgical technique.
Instruments were developed without adequate consideration given to the need for a new regulatory filing. This event was reported to FDA as part of Stryker's retrospective review of K2M product field actions; recall actions were completed in 2017.
- 2019-10-16FDA-DeviceRepro-Med Systems, Inc.Class II20" 2-Needle 26-Gauge 6mm HlgH-Flo Subcutaneous Safety Needle Sets" Product # RMS22606 - Product Usage: RMS HlgH-Flo Subcutaneous Safety Needle Sets are intended for the delivery of medication to the subcutaneous tissue.
Mislabeled needle lengths
- 2019-10-16FDA-DeviceBaxter Healthcare CorporationClass IIIEXACTAMED Oral Dispensers, 5mL, AMBER: (a) PHARMACY PACK, REF H9388105, x 100; (b) PHARMACY PACK, REF H9388505, x 500; (c) COST CONTAINMENT, REF H93884205, x 1500 - Product Usage: Oral Liquid Dispensers are intended for the accurate dose measurement and clean intermittent administration of liquid medication and nutrition to humans through the patient s mouth.
Exactamed 5 mL Oral Dispenser Pharmacy Packs may contain both 5 mL and 3 mL oral dispensers.
- 2019-10-16FDA-DeviceBecton Dickinson & CompanyClass IICareFusion MaxPlus Clear Needleless Connector. Intended for Positive Displacement Connector for needleless access to the IV line and/or IV catheter during IV therapy Cat. No. MP1000-C
After disconnection of a luer, the valve of the connector remains recessed; not fully closing, either temporarily or permanently. The risks of blood loss and air embolism exist if the valve remains permanently open and goes undetected, especially in the neonate population.
- 2019-10-16FDA-DeviceEthicon, Inc.Class IISTRATAFIX Spiral PDS Plus Violet 27 (70cm) CTX Needle Product code: SXPP1B401 suture
The product does not meet certain internal strength testing specifications.
- 2019-10-16FDA-DeviceEthicon, Inc.Class IISTRATAFIX Spiral PDS Plus Violet 18 (45cm) CT-1 Needle Product code: SXPP1B430 suture
The product does not meet certain internal strength testing specifications.
- 2019-10-16FDA-DeviceLeica Biosystems Imaging, Inc.Class IIAperio CS2 CE IVD scanner instrument, scanner part #23CS100CE.
The incorrect sensor was used in the assembly of the camera in the scanner instrument which can potentially generate whole slide images that impact analysis algorithm results, as well as potentially impact the visual appearance of whole slide images.
- 2019-10-16FDA-DeviceHelena Laboratories, Corp.Class IISPIFE ImmunoFix-6 Kit, Cat. No. 3401, 3401T, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX.
The plates do not have wells punched for the controls in the patient 1 area.
- 2019-10-16FDA-DeviceRoche Diagnostics Operations, Inc.Class IIAssayTip/AssayCup Tray-used on the cobas e 801 module which is a part of the cobas c 8000 MODULAR Analyzer Series (IVD for in vitro determination of analytes in human body fluids) Catalog Number: 5694302001
AssayTips part of the AssayTip/AssayCup Tray Tips used on the cobas e 801 module- may have an Abnormal Internal Structure and lead to Incorrect patient results
- 2019-10-16FDA-DeviceRoche Diagnostics Operations, Inc.Class IIOnline TDM Vancomycin Gen.3 ( 200 tests) Catalog # 06779344190 Analyzer/Module: cobas c 311 and cobas c 501/502
Inaccurately Low Sample Results
- 2019-10-16FDA-DeviceRoche Diagnostics Operations, Inc.Class IIOnline TDM Vancomycin Gen.3 (100 tests) Catalog # 06779336190 Analyzer/Module: cobas c 311 and cobas c 501/502
Inaccurately Low Sample Results
- 2019-10-16FDA-DeviceOhmeda MedicalClass IIGiraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers pr
GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.
- 2019-10-16FDA-DeviceAZIYO BIOLOGICSClass IICanGaroo Envelopes labeled as the following: 1. CanGaroo Envelope (Large) (single pack); 2. CanGaroo Envelope (Medium) (single pack); 3. CanGaroo Envelope (Extra Large) (single pack); 4. CanGaroo Envelope (Extra Extra Large) (single pack); 5. CanGaroo Envelope (Large) (5 pack); 6. CanGaroo Envelope (Medium) (5 pack); 7. CanGaroo Envelope (Extra Large) (5 pack); 8. CanGaroo Envelope (Extra Extra Large) (5 pack) Product Usage: The CanGaroo Envelope is in
Possible suture brittleness or breakage along the Envelope seam before the expiration date of the device.
- 2019-10-16FDA-DeviceAZIYO BIOLOGICSClass IIBioEnvelope, surgical mesh envelope. Labeled as the following: 1. BioEnvelope (Medium) (single pack); 2. BioEnvelope (Large) (single pack); 3. BioEnvelope (Extra Large) (single pack); 4. BioEnvelope (Extra Extra Large) (single pack); 5. BioEnvelope (Large) (5-pack); 6. BioEnvelope (Medium) (5-pack); 7. BioEnvelope (Extra Extra Large) (5-pack); 8. BioEnvelope (Extra Large) (5-pack) Product Usage: The BioEnvelope is intended to securely hold a cardiac im
Possible suture brittleness or breakage along the Envelope seam before the expiration date of the device.
- 2019-10-16FDA-DeviceVital Scientific N.V.Class IIV-Twin, Catalog # 6002-800, UDI: 03661540600180 - Product Usage: An in vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.
The action is being initiated because the cooling indicator board was improperly installed. This was first noticed by a service employee during assembly. If for some reason the cooling of the reagent rotor fails, the reagent cooling indicator light will not illuminate and the acoustic beeper will not go off. As a consequence the operator will not be warned and the reagent storage conditions may fall out of specification.
- 2019-10-09FDA-DeviceWOM World of Medicine AGClass IIAquilex Fluid Control System component: Bag deflector, REF AQL-100CBS Product Usage: Distends uterus with fluid during diagnostic and operative hysteroscopies.
The deficit displayed by the pump can differ from the real deficit, leading potentially to patient risk.
- 2019-10-09FDA-DeviceRoche Diabetes Care, Inc.Class IIAccu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the
Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
- 2019-10-09FDA-DeviceRoche Diabetes Care, Inc.Class IIAccu-Chek Aviva II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the bo
Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
- 2019-10-09FDA-DeviceELITech Group B.V.Class IIIViva-ProE Systems, Model # 6003-670, UDI: 03661540600357 - Product Usage: The ELITech Clinical Systems Selectra ProM is an automated clinical chemistry system intended for use in clinical laboratories. It is intended to be used for a variety of assay methods that have been applied to spectrophotometric and electrochemical techniques. The system has two core modules: one consisting of a spectrophoto-metric system for measurement of analytes using spectrophotometric techniques, such as end point,
Software malfunction; The action being taken due to the device becoming inoperable due to software lockup.
- 2019-10-09FDA-DeviceBeckman Coulter Inc.Class IIThe Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.
Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).
- 2019-10-09FDA-DeviceZimmer GmbHClass IISidus Stem-Free Shoulder Humeral Head 50-18 Item number: 01.04555.500 orthopedic implant - Product Usage: The Sidus Stem-Free Shoulder is designed for long-term implantation into the human shoulder joint as a total shoulder arthroplasty in combination with Zimmer or Biomet s glenoid components, or as a hemi shoulder arthroplasty.
Incorrect labeling on the products package. The label on the package shows size 50-18 whereas the implant inside the packaging size 40-14.