Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2019-09-04FDA-DeviceBard Peripheral Vascular IncClass IIEQUISTREAM Long-Term Hemodialysis Catheter with Preloaded Styet, REF/Product code:5903230, 14.5F, Straight ,23cm length, BARD, UDI: 00801741013706
The action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
- 2019-09-04FDA-DeviceBard Peripheral Vascular IncClass IIEQUISTREAM Long-Term Hemodialysis Catheter with Preloaded Styet, REF/Product code:5903190, 14.5F, Straight ,19cm length, BARD, UDI# 00801741013690
The action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
- 2019-09-04FDA-DeviceBard Peripheral Vascular IncClass IIEQUISTREAM LONG-TERM HEMODIALYSIS CATHETER, REF/Product code:5905280, 14.5 Fr. Alphacurve Catheter with Standard Kit, 28cm length, BARD, UDI# 00801741013812
The action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
- 2019-09-04FDA-DeviceBard Peripheral Vascular IncClass IIEQUISTREAM LONG-TERM HEMODIALYSIS CATHETER, REF/Product code:5905240, 14.5 Fr. Alphacurve Catheter with Standard Kit, 24cm length, BARD, UDI# 00801741013805
The action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
- 2019-09-04FDA-DeviceBard Peripheral Vascular IncClass IIEQUISTREAM LONG-TERM HEMODIALYSIS CATHETER, REF/Product Code:5905190, 14.5 Fr. Alphacurve Catheter with Standard Kit, 19 cm length, BARD, UDI# 00801741013799
The action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
- 2019-09-04FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis zee and Artis Q/ Q.zen Interventional Fluoroscopic X-Ray System with patient tables as follows: Systems Material Number Artis zee biplane 10094141; Artis Q floor 10848280; Artis Q biplane 10848282; Artis Q.zen floor 10848353; Artis Q.zen biplane 10848355; Artis zee ceiling 10094137; Artis Q ceiling 10848281; Artis Q zeego 10848283 ArtisQ.zen ceiling 10848354 Product Usage: Artis is a family of dedic
Artis zee/Q systems patient tables may potentially have cracks in the table mainframe, and cause mechanical detachment of the upper part of the patient bed and lead to hazardous situations for patients and medical personnel due to falling parts
- 2019-09-04FDA-DeviceBausch & Lomb Inc IrbClass IILaseredge Knife 1.5mm X 1.7mm TRAPEZOID ANGLED, REF E7599 Product Usage: Available to surgeons for use in ophthalmic surgical procedures.
Possible dull knife edge
- 2019-09-04FDA-DeviceBausch & Lomb Inc IrbClass IILaseredge 1.1mm Arrow Knife Angled, REF E7592 Product Usage: Available to surgeons for use in ophthalmic surgical procedures.
Possible dull knife edge
- 2019-09-04FDA-DeviceCooperSurgical, Inc.Class IITransWarmer Warming Infant Transport Mattress, Model no. 20421 Product Usage: The TransWarmer¿ Warming Infant Transport Mattress is gel-filled, disposable mattress that provides heat for up to two hours when cold stress is a concern. The patented WarmGel heats up to 104 (when starting from 75 ) in 60 seconds or less. The product is sold in boxes of 6 Infant Transport Mattresses per box. The intended use of the product is to provide warmth during infant transport within the hospital or betwe
The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.
- 2019-09-04FDA-DeviceAbbott Gmbh & Co. KGClass IIAlinity ci -series System Control¿Module, Part Number 03R70 01 Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a clinical chemistry module and an immunoassay module, each performing all sample processing activities, and a system control module to provide a single user friendly interface. Each system can be customized by combining on
Abbott has identified an issue with all on market versions (v2.6.0 and v2.6.1) of Alinity ci series Software where incorrect results may occur after a system Stop due to the Alinity i re use of reaction vessels (RVs). This issue only occurs if the system is transitioned from Processing to Stopped to Idle.
- 2019-09-04FDA-DeviceRoche Diagnostics Operations, Inc.Class IIRoche Elecsys Anti-CCP Immunoassay Catalog # 05031656160
Performance issues with certain lots of the Elecsys Anti-CCP assay on the cobas e 411 analyzer; MODULAR ANALYTICS E 170 module; and cobas e 601, 602, and 801 modules with plasma samples.
- 2019-09-04FDA-DeviceRoche Diagnostics Operations, Inc.Class IIRoche Elecsys Anti-CCP Immunoassay Catalog # 07251670190
Performance issues with certain lots of the Elecsys Anti-CCP assay on the cobas e 411 analyzer; MODULAR ANALYTICS E 170 module; and cobas e 601, 602, and 801 modules with plasma samples.
- 2019-09-04FDA-DeviceLeventon S. A. U.Class IIDOSI-FUSER 12591 5-500D1 -USA, 12591 5-600D1 -USA Provide parenteral drug infusions at a constant flow without impeding patient mobility
The product lacks 510(k) clearance, and therefore is misbranded and must be removed from the U.S. market.
- 2019-09-04FDA-DeviceAgilent Technologies, Inc.Class IIMassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass.
When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.
- 2019-09-04FDA-DeviceMedline Industries IncClass IIMedline SYRINGE, PISTON,IRR TRAY,W/.9, REF DYNC2303. Sterile Piston Irrigation Syringe Tray with Saline
Samples in the identified lot were found to have incomplete or open packaging seals.
- 2019-09-04FDA-DeviceSpectrum Medical Ltd.Class IQuantum Pump Console
Potential for Quantum Pump Console, part of the Quantum Perfusion Systems, to unexpectedly shut down while in use.
- 2019-08-28FDA-DeviceCambridge Sensors LimitedClass IIMicrodot Xtra blood glucose test strips
The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.
- 2019-08-28FDA-DeviceMolded Products IncClass IILuer Lock Set - Catalog # MPC-125, for IV administration sets.
Mislabeling - Some unit of sale packages (bags of 100 sets of MPC-125) were found to be incorrectly labeled as Molded Products Male to Male Luer Adapters, Catalog # MPC-150.
- 2019-08-28FDA-DeviceSiemens Medical Solutions USA, Inc.Class IIACUSON NX3 Elite Diagnostic Ultrasound System, Model # 11235411, RX Only, Siemens Medical Solutions, containing software versions: VA10A, VA10B, VA10C, VA10D, VA10E, VA10F, VB20A, and VB20B; UDI # 04056869004303 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal a
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware
- 2019-08-28FDA-DeviceSiemens Medical Solutions USA, Inc.Class IIACUSON NX3 Diagnostic Ultrasound System, Model # 11235612, RX Only, Siemens Medical Solutions, containing software versions: VA10A, VA10B, VA10C, VA10D, VA10E, VA10F, VB20A, and VB20B; UDI # 04056869004297 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applica
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware
- 2019-08-28FDA-DeviceSiemens Medical Solutions USA, Inc.Class IIACUSON NX2 Elite Diagnostic Ultrasound System, Model # 11284500, RX Only, Siemens Medical Solutions, containing software version VA10A, VA10B, VA11A, VA11B; UDI # 04056869050775 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware
- 2019-08-28FDA-DeviceSiemens Medical Solutions USA, Inc.Class IIACUSON NX2 Diagnostic Ultrasound System, Model # 11284381, RX Only, Siemens Medical Solutions, containing software version VA10A, VA10B, VA11A, VA11B; UDI # 04056869030005 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for t
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware
- 2019-08-28FDA-DeviceThoratec Corp.Class IIThoratec Heartmate 3 Sealed Outflow Graft with Bend Relief, Sterile, REF 105581US/105581, RX only, Part of Left Ventricular Assist Device (LVAD) UDI: 00813024013266
Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.
- 2019-08-28FDA-DeviceThoratec Corp.Class IIThoratec HeartMate 3 LVAS Implant Kit, Sterile, REF 106524/106524US, RX only, Left Ventricular Assist System
Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.
- 2019-08-28FDA-DeviceAniara Diagnostica LLCClass IIZYMUTEST HIA MonoStrip, IgGAM ELISA kit, Model No. RK041D, for in vitro diagnostic use.
Residual crystallization of the microplate and recurrent negative controls out of range.
- 2019-08-28FDA-DeviceAniara Diagnostica LLCClass IIZYMUTEST HIA MonoStrip, IgG ELISA kit, Model No. RK041A, for in vitro diagnostic use.
Residual crystallization of the microplate and recurrent negative controls out of range.
- 2019-08-28FDA-DeviceSiemens Medical Solutions USA, IncClass IIMAGNETOM Prisma magnetic resonance diagnostic devices Model #10849582
Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.
- 2019-08-28FDA-DeviceSiemens Medical Solutions USA, IncClass IIMAGNETOM Spectra magnetic resonance diagnostic devices Model # 10837643
Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.
- 2019-08-28FDA-DeviceSiemens Medical Solutions USA, IncClass IIMAGNETOM Skyra magnetic resonance diagnostic devices Model # 10432915
Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.
- 2019-08-28FDA-DeviceSiemens Medical Solutions USA, IncClass IIMAGNETOM Aera magnetic resonance diagnostic devices Model # 10432914
Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.