Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2019-08-14FDA-DeviceSenTec AGClass IISenTec Membrane Changer Insert [5 pcs - Product Usage: is indicated for continuous, non-invasive monitoring of carbon dioxide tension and oxygen tension as well as oxygen saturation and pulse rate in adult and pediatric patients. In neonatal patients the SDMS is indicated for carbon dioxide and oxygen tension monitoring only. Oxygen tension monitoring is contraindicated for patients under gas anesthesia.]
There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.
- 2019-08-14FDA-DeviceSenTec AGClass IISenTec Membrane Changer Set (1 charger plus 1 insert) - Product Usage: is indicated for continuous, non-invasive monitoring of carbon dioxide tension and oxygen tension as well as oxygen saturation and pulse rate in adult and pediatric patients. In neonatal patients the SDMS is indicated for carbon dioxide and oxygen tension monitoring only. Oxygen tension monitoring is contraindicated for patients under gas anesthesia.
There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.
- 2019-08-14FDA-DeviceSenTec AGClass IISenTec Neonatal Starter Set with and without Service Gas (2 Chargers each) - Product Usage: is indicated for continuous, non-invasive monitoring of carbon dioxide tension and oxygen tension as well as oxygen saturation and pulse rate in adult and pediatric patients. In neonatal patients the SDMS is indicated for carbon dioxide and oxygen tension monitoring only. Oxygen tension monitoring is contraindicated for patients under gas anesthesia.
There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.
- 2019-08-14FDA-DeviceSenTec AGClass IISenTec Standard Starter Set with and without Service Gas (2 changers each) - Product Usage: Is indicated for continuous, non-invasive monitoring of carbon dioxide tension and oxygen tension as well as oxygen saturation and pulse rate in adult and pediatric patients. In neonatal patients the SDMS is indicated for carbon dioxide and oxygen tension monitoring only. Oxygen tension monitoring is contraindicated for patients under gas anesthesia.
There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.
- 2019-08-14FDA-DeviceSenTec AGClass IISenTec Membrane Changer (reloadable) [9 pcs] - Product Usage: Is indicated for continuous, non-invasive monitoring of carbon dioxide tension and oxygen tension as well as oxygen saturation and pulse rate in adult and pediatric patients. In neonatal patients the SDMS is indicated for carbon dioxide and oxygen tension monitoring only. Oxygen tension monitoring is contraindicated for patients under gas anesthesia.
There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.
- 2019-08-14FDA-DeviceFresenius Medical Care Renal Therapies Group, LLCClass IILiberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal dialysis (APD) in conjunction with the Liberty cycler as follows: (1) LIBERTY Line Cycler Set with Stay-safe connectors (Dual Patient Connector) Part Number: 050-87212 (2) LIBERTY Line Cycler Set with Stay-safe connectors (Single Patient Connector) Part Number: 050-87215 (3) LIBERTY Integrated Cycler Set with single Stay-safe patient connector and extended drain line Part Number: 050-87216
Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for contamination.
- 2019-08-14FDA-DeviceRandox Laboratories, LimitedClass IIHuman Assayed Multi-Sera Level 2, Model NO. HN1530 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic assays. The Human Assayed Multi-sera is for the control of accuracy.
The RX Series control target and range value for ALT (Tris buffer without P5P 37C), Bicarbonate and Gamma-GT has been incorrectly assigned in specific lots of the Human Assayed Multi-Sera Controls.
- 2019-08-14FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus MAJ-209 Single Use Suction Valves, Model Number MAJ-209 Product Usage: This product has been designed to be attached to the suction cylinder of compatible endoscopes for aspirating fluid from the distal end of the endoscopes through the instrument channel. Compatible endoscopes include Olympus bronchoscopes, cystoscopes, rhino-laryngoscopes and pleurascopes models.
There is a potential for the single use suction valve to come apart and/or remain stuck inside the endoscope.
- 2019-08-14FDA-DeviceGeistlich Pharma North America, Inc.Class IIGeistlich Bio-Oss Pen, Sterile cancellous granules in an applicator, 1-2mm 0.5g, For single use only, Part Number 20126 - Product Usage: Augmentation or reconstructive treatment of the alveolar ridge; Filling of infrabony periodontal defects; Filling of defects after root resection, apicoectomy and cystectomy; Filling of extraction sockets to enhance preservation of the alveolar ridge; Elevation of the maxillary sinus floor, etc.
When unscrewing the green cap from the pen, a visible piece of green plastic was seen on the outer barrel of the pen.
- 2019-08-14FDA-DeviceGeistlich Pharma North America, Inc.Class IIGeistlich Bio-Oss Pen, Sterile cancellous granules in an applicator, 0.25-1.mm 0.5g, For single use only, Part Number 20116 - Product Usage: Augmentation or reconstructive treatment of the alveolar ridge; Filling of infrabony periodontal defects; Filling of defects after root resection, apicoectomy and cystectomy; Filling of extraction sockets to enhance preservation of the alveolar ridge; Elevation of the maxillary sinus floor, etc.
When unscrewing the green cap from the pen, a visible piece of green plastic was seen on the outer barrel of the pen.
- 2019-08-14FDA-DeviceGeistlich Pharma North America, Inc.Class IIGeistlich Bio-Oss Pen, Sterile cancellous granules in an applicator, 0.25-1.mm 0.25g, For single use only, Part Number 20115 - Product Usage: Augmentation or reconstructive treatment of the alveolar ridge; Filling of infrabony periodontal defects; Filling of defects after root resection, apicoectomy and cystectomy; Filling of extraction sockets to enhance preservation of the alveolar ridge; Elevation of the maxillary sinus floor, etc.
When unscrewing the green cap from the pen, a visible piece of green plastic was seen on the outer barrel of the pen.
- 2019-08-14FDA-DeviceBiomet 3i, LLCClass IICERTAIN(R) BELLATEK(R) TITANIUM ABUTMENT 3.4MM, Item Number ILDAT3 - Product Usage:BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible
During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.
- 2019-08-14FDA-DeviceBiomet 3i, LLCClass IICERTAIN(R) BELLATEK(R) TIN ABUTMENT 3.4MM, Item Number ILDAN3 - Product Usage: BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible
During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.
- 2019-08-14FDA-DeviceBiomet 3i, LLCClass IICERTAIN(R) BELLATEK(R) TITANIUM ABUTMENT 3.4MM, Item Number IIAT3 - Product Usage: BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible
During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.
- 2019-08-14FDA-DeviceBiomet 3i, LLCClass IICERTAIN(R) BELLATEK(R) TITANIUM ABUTMENT 3.4MM, Item Number IEDAT3 - Product Usage: BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible
During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.
- 2019-08-14FDA-DeviceBiomet 3i, LLCClass IICERTAIN(R) BELLATEK(R) TIN ABUTMENT 3.4MM, Item Number IEDAN3 - Product Usage: BIOMET3i Restorative products are intended for use as an accessory to endosseous dental implants for placement in the maxilla and mandible
During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.
- 2019-08-14FDA-DeviceSage Products IncClass IIToothete Oral Care Sodium Bicarbonate Mouthpaste, Net Wt. 1/3 oz/ 9g, included in Independent Care System. The system is an oral care convenience kit which includes: an Ultra-Soft Toothbrush, 0.5 oz. tube of Mouth Moisturizer, 0.33 oz. /9g tube of Sodium Bicarbonate Mouthpaste and 1.5fl.oz. bottle of Corinz oral rinse. Product Usage: The mouthpaste is used to clean teeth, gums, and the tongue.
A voluntary recall of a lot (67924) of Independent Care System due to the potential for damaged Sodium Bicarbonate Mouthpaste (cosmetic) tubes.
- 2019-08-14FDA-DeviceMED-EL Elektromedizinische Gereate, GmbhClass IIMED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sensations via electrical stimulation of the auditory pathways for severely to profoundly hearing impaired individuals who obtain little or no benefit from acoustic amplification in the best aided condition
Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.
- 2019-08-14FDA-DeviceMED-EL Elektromedizinische Gereate, GmbhClass IIMED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard Product Usage: The SONATATIi100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sensations via electrical stimulation of the auditory pathways for severely to p
Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.
- 2019-08-14FDA-DeviceMED-EL Elektromedizinische Gereate, GmbhClass IIMED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCERTO PIN +FLEXsoft; d. Mil000 CONCERTO +Standard; e. Mil000 CONCERTO +Medium; f. Mi1000 CONCERTO PIN +Standard; g. M11000 CONCERTO +Compressed Product Usage: The Mi1000 CONCERTO Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatibl
Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.
- 2019-08-14FDA-DeviceComet Technologies Usa IncClass IIYXLON Cougar Cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and quality control
Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door.
- 2019-08-14FDA-DeviceComet Technologies Usa IncClass IIYXLON Cheetah cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and quality control
Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door.
- 2019-08-14FDA-DeviceASTRA, LLCClass IILidar System for Erosion Flume Application Class 1 Laser Product Product Usage: Product is installed and fixed above a dynamic indoor flume designed for modeling water erosion in an indoor research environment with limited human access
Failure to comply event due to incorrect labeling of a class 1 laser product.
- 2019-08-14FDA-DeviceGk Photonics IncClass IIBurstberry Laser Projection Series
The User Remote Interlock Dongle provided with the E-Stop Remote is not sufficient to fully comply with the FDA's Manual Reset Mechanism requirement for laser projectors and could unintentionally expose the user and/or observers to high power laser radiation. It needs to be replaced for safety reasons.
- 2019-08-14FDA-DeviceGk Photonics IncClass IILogolas Laser Projection Series
The User Remote Interlock Dongle provided with the E-Stop Remote is not sufficient to fully comply with the FDA's Manual Reset Mechanism requirement for laser projectors and could unintentionally expose the user and/or observers to high power laser radiation. It needs to be replaced for safety reasons.
- 2019-08-14FDA-DeviceGk Photonics IncClass IIMaxim Laser Projection Series
The User Remote Interlock Dongle provided with the E-Stop Remote is not sufficient to fully comply with the FDA's Manual Reset Mechanism requirement for laser projectors and could unintentionally expose the user and/or observers to high power laser radiation. It needs to be replaced for safety reasons.
- 2019-08-14FDA-DeviceGk Photonics IncClass IISpectrum Laser Projection Series
The User Remote Interlock Dongle provided with the E-Stop Remote is not sufficient to fully comply with the FDA's Manual Reset Mechanism requirement for laser projectors and could unintentionally expose the user and/or observers to high power laser radiation. It needs to be replaced for safety reasons.
- 2019-08-14FDA-DeviceGk Photonics IncClass IIAtom Laser Projection Series
The User Remote Interlock Dongle provided with the E-Stop Remote is not sufficient to fully comply with the FDA's Manual Reset Mechanism requirement for laser projectors and could unintentionally expose the user and/or observers to high power laser radiation. It needs to be replaced for safety reasons.
- 2019-08-14FDA-DeviceGk Photonics IncClass IIClubMax Laser Projection Series
The User Remote Interlock Dongle provided with the E-Stop Remote is not sufficient to fully comply with the FDA's Manual Reset Mechanism requirement for laser projectors and could unintentionally expose the user and/or observers to high power laser radiation. It needs to be replaced for safety reasons.
- 2019-08-14FDA-DeviceSt Jude Medical Inc.Class IELLIPSE DR, Tiered-therapy cardioverter/defibrillator, REF: CD2411-36Q, UDI: 05414734507615
Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.