Medical device recalls
39226 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2019-01-02FDA-DeviceAbbott Ireland Diagnostics DivisionClass IIAlinity i TSH Reagent Kit, List Number 07P4830
Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
- 2019-01-02FDA-DeviceICU Medical, Inc.Class IIIced Thermoset Closed-Loop Injectate Delivery System for cold Injectate. 1 unit per Cavity #2-4"x9" Cav/20 pieces per case R1-15645
Potential for leakage due to defect in the thermoset check valve component.
- 2019-01-02FDA-DeviceICU Medical, Inc.Class IIIced Thermoset Closed-Loop Injectate Delivery System for cold Injectate. 1 unit per Cavity #6x12x2.6 Cav/10 pieces per case R1-6090
Potential for leakage due to defect in the thermoset check valve component.
- 2019-01-02FDA-DeviceICU Medical, Inc.Class IIThermoset Room Temperature Closed Loop Injectate Delivery System. 1 unit per Cavity #3 Cavity Deep/20 pieces per case R1-6090
Potential for leakage due to defect in the thermoset check valve component.
- 2019-01-02FDA-DeviceRandox Laboratories Ltd.Class IISP CAL (LIQ), REF IT2692 This product is intended for in vitro diagnostic use with Randox Alpha-1-Antitrypsin, Alpha-1-Acid Glycoprotein, IgA, IgG and IgM assays that require sample predilution.
Calibrator value sheets have been modified for assigned value of Alpha-1-Acid-Glycoprotein (AGP), IgA and IgG to improve the alignment. IgG measurement reference has also been corrected.
- 2018-12-26FDA-DeviceKonica Minolta Healthcare, Americas, Inc.Class IIBucky Stand SU-4000 Input 2017.11.17. SU41711085 Date: 2017-11-17; Board Name; Board Version, Firmware Version Advanced Universal U-Arm 4000 System. The Positioner consists of the following associated subassemblies: Vertical Column, Swivel Arm with variable height, X-ray Tube, Automatic Exposure Control, Collimator, Touch Screen Control and Image Receptor Assembly. The Positioner and the Patient Mobile Table are associated equipment to the Generator X-ray Unit.
Unintentional U-Arm movement
- 2018-12-26FDA-DeviceFresenius Medical Care Renal Therapies Group, LLCClass IIOptiflux F160NR Capillary High Flux Dialyzer Single Use Only, Product Code 0500316E
Potential for external blood leaks from the dialyzer header
- 2018-12-26FDA-DeviceHowmedica Osteonics Corp.Class IIReliance LITe Nerve Hook Bayoneted, 90, Blunt Tip, REF # 48066103 The blunt probe can be used as a conventional means of achieving nerve retraction. The bend at the proximal tip of the ring the discectomy, as well as during the insertion of an IBD during a conventional TLIF procedure.
Instead of having a spherical diameter, the probe was observed to have 2 radii and a flat surface. Additionally, upon further investigation, it was observed that some parts exhibited an edge on the intersection between the radius and the diameter of the probe.
- 2018-12-26FDA-DeviceCOVIDIEN LLCClass II8Fr SAFE GRAD CAT W/Valve (Argyle Suction Catheter Kits with Chimney Valve), Model Number 36826
The product was released into commercial distribution while subject to importation refusal.
- 2018-12-26FDA-DeviceNidek Inc.Class IIIRT-5100 Refractor, a component of the Epic-5100 System.
The lens bonding was incomplete on the refractor.
- 2018-12-26FDA-DeviceBracco Injeneering S.A.Class IIEmpower FastLoad CT Syringe Pack -sterile disposable syringe kit for use with the EmpowerCT¿, EmpowerCTA¿ and EmpowerCTA¿+ Injector Systems SKU 017344/6720 Empower FastLoad" CT Syringe Pack is a sterile, disposable syringe kit designed for use with EmpowerCT¿, EmpowerCTA¿ and EmpowerCTA¿+ Injector Systems. It is used to intravenously facilitate the administration of contrast media and flushing solutions into the human vascular systems.
Insufficient seal on the sterile barrier of the device, compromising the sterility of the device
- 2018-12-26FDA-DeviceOrtho-Clinical DiagnosticsClass IIVITROS 4600 Chemistry System, Version 3.3.1 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain biological fluids (depending on the analyte) such as serum, plasma, urine and cerebrospinal fluid.
Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
- 2018-12-26FDA-DeviceOrtho-Clinical DiagnosticsClass IIVITROS¿ 3600 Immunodiagnostic System, Version 3.3.1 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain biological fluids (depending on the analyte) such as serum, plasma, urine and cerebrospinal fluid.
Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
- 2018-12-26FDA-DeviceOrtho-Clinical DiagnosticsClass IIVITROS 5,1 FS Chemistry System, Version 3.0 & below For use in the in vitro quantitative measurement of a variety of analytes of clinical interest in certain biological fluids (depending on the analyte) such as serum, plasma, urine and cerebrospinal fluid.
Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
- 2018-12-26FDA-DeviceOrtho-Clinical DiagnosticsClass IIVITROS 5600 Integrated System, Version 3.3.1 & below For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Product Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
- 2018-12-26FDA-DeviceFoundation Medicine, Inc.Class IIFoundationOne CDx Test Results
An incorrect test result for a single analyte may have been reported for up to 800 physician reports
- 2018-12-26FDA-DeviceDeerfield Imaging, Inc.Class IIIMRIS ORT 300, Removable Operating Room Table, Part Numbers: (a) 114093-000, (b) 114093-006 - Product Usage: The ORT200 and ORT300 are intended for use during diagnostic examinations, radiologic procedures, or surgical procedures to support and position a patient. The tables are indicated for imagining from head to sacrum with MR/CT imaging modalities. The ORT100 table is intended for use during diagnostic examination or surgical procedures to support and position the patient.
Table may drift in the roll position while in use
- 2018-12-26FDA-DeviceDeerfield Imaging, Inc.Class IIIMRIS ORT 200, Removable Operating Room Table, Part Numbers: (a) 113821-000, (b) 113821-600, (c) 114148-000, (d) 114148-600 - Product Usage: The ORT200 and ORT300 are intended for use during diagnostic examinations, radiologic procedures, or surgical procedures to support and position a patient. The tables are indicated for imagining from head to sacrum with MR/CT imaging modalities. The ORT100 table is intended for use during diagnostic examination or surgical procedures to support and positio
Table may drift in the roll position while in use
- 2018-12-26FDA-DeviceDeerfield Imaging, Inc.Class IIIMRIS ORT 100 table, Part Numbers: (a) 110470-000, (b) 110470-600 - Product Usage: The ORT200 and ORT300 are intended for use during diagnostic examinations, radiologic procedures, or surgical procedures to support and position a patient. The tables are indicated for imagining from head to sacrum with MR/CT imaging modalities. The ORT100 table is intended for use during diagnostic examination or surgical procedures to support and position the patient.
Table may drift in the roll position while in use
- 2018-12-26FDA-DeviceDeerfield Imaging, Inc.Class IIIMRIS T2X table, Part Numbers: (a)108000-000, (b)108000-600, (c)109682-000, (d) 109682-600, (e) 110470-600 - Product Usage: The ORT200 and ORT300 are intended for use during diagnostic examinations, radiologic procedures, or surgical procedures to support and position a patient. The tables are indicated for imagining from head to sacrum with MR/CT imaging modalities. The ORT100 table is intended for use during diagnostic examination or surgical procedures to support and position the patient.
Table may drift in the roll position while in use
- 2018-12-26FDA-DeviceZimmer GmbHClass IIAvenir¿ Muller Stem Standard, uncemented HA, 1, Taper 12/14 Item Number: 01.06010.001 - Product Usage: Avenir@ Muller Stems are intended to reduce pain and increase hip mobility through long term cementless fixation of total or hemi hip arthroplasty in the femur of patients with an adequate bone stock to support the component. A system consisting of a stem, a ball head and a cup is used for the treatment of degenerative diseases or trauma of the hip. All Avenir Muller Stems are intended to provi
Mislabeled Labeled Avenir M¿ller Stem size 1, the stem inside the packaging is an Avenir M¿ller Stem size 2 .
- 2018-12-26FDA-DeviceTheragenics CorporationClass IITheragenics Applicator Needle (18g x 20cm) The device is comprised four basic elements; a wire stylus with plastic handle, a plastic needle hub, a protective needle sleeve and the hollow steel needle.
Some of the pouches containing Brachytherapy Applicator Needles (18g x 20cm) were not sealed
- 2018-12-26FDA-DeviceZimmer Biomet, Inc.Class IIZimmer Compress Devices and Instruments: Item Number/Item Description 178350 Compress Device Anti-Rotation Spindle 178353 Compress Device Anti-Rotation Spindle 178356 Compress Device Anti-Rotation Spindle 178359 Compress Device Anti-Rotation Elliptical Spindle 178537 Compress Device Centering Sleeve 15mm 178541 Compress Device Centering Sleeve 19mm 178545 Compress Device Centering Sleeve 23mm 178544 Compress Device Centering Sleeve 22mm 178738 Compress Device Centering Sleeve 28mm 178
Correction to update the surgical technique for the Compress System. To make users aware of the changes that were made only to the selection criteria section on page 42 regarding which array to use in the placement of the anti-rotation pins.
- 2018-12-26FDA-DeviceGE Healthcare, LLCClass IICentricity Universal Viewer Zero Footprint Client, Model Number 2090258-001, Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2, SP9 and SP9.0.1 display o orientation markers can become is incorrect.
Issue #1 affects Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2. When an image is flipped horizontally and vertically, the orientation markers appear correctly on the image to which they were applied. However, once new image rendering happens on this image or series (with scrolling; playing cin¿; maximizing or changing viewports/layouts; or synchronizing viewports and resizing the browser), the new image orientation is correctly maintained but the display of orientation markers can become incorrect. Issue #2 affects Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a, SP8.0.2, SP9 and SP9.0.1. When an image in a series is flipped and then rotated in Zero Footprint Viewer (ZFP), the orientation markers appear correctly on the image to which they were applied. However, once you navigate to another image in the series and the image you navigated to has a different orientation than the prior image from which you navigated, the new image orientation is correctly maintained but the display of orientation markers can become incorrect.
- 2018-12-26FDA-DeviceAdvanced Instruments, LLCClass IIGloCyte RBC Reagent - Product Usage: The GloCyte Automated Cell Counter System is intended for use by trained healthcare professionals in clinical laboratories to provide quantitative determination of fluorescence labeled total nucleated cells and erythrocytes in cerebrospinal fluid collected from adult and pediatric patients. The GloCyte Low and High Level Controls are assayed hematology controls designed to monitor the performance of the GloCyte Automated Cell Counter System. Assayed parameter
RBC Reagent failure. High fluorescent counts during RBC Reagent + Diluent QC check that will result in failed QC check, preventing user from testing patient samples for RBC.
- 2018-12-26FDA-DevicebioMerieux, Inc.Class IIVITEK¿ 2 Systems Software Version 9.01 Update Kit.
The VITEK¿ 2 Systems Software Version 9.01 software may display incorrect organism identification information as well as the associated Antimicrobial Susceptibility Testing (AST) results when using the VITEK¿ 2 FLEXPREP" feature.
- 2018-12-26FDA-DeviceCardinal Health 200, LLCClass IIProtexis Neoprene Surgical Glove, Size 9.0 This powder-free sterile light brown colored surgeons gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Labeling issue in that the breakthrough (i.e., permeation) times for two chemotherapy drugs (Carmustine and Thiotepa) stated on the wallet (primary packaging) of the Protexis Neoprene surgical gloves are incorrect. Additionally, breakthrough times for Mitomycin were added to the same label in error.
- 2018-12-26FDA-DeviceCardinal Health 200, LLCClass IIProtexis Neoprene Surgical Glove, Size 8.5 This powder-free sterile light brown colored surgeons gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Labeling issue in that the breakthrough (i.e., permeation) times for two chemotherapy drugs (Carmustine and Thiotepa) stated on the wallet (primary packaging) of the Protexis Neoprene surgical gloves are incorrect. Additionally, breakthrough times for Mitomycin were added to the same label in error.
- 2018-12-26FDA-DeviceCardinal Health 200, LLCClass IIProtexis Neoprene Surgical Glove, Size 8.0 This powder-free sterile light brown colored surgeons gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Labeling issue in that the breakthrough (i.e., permeation) times for two chemotherapy drugs (Carmustine and Thiotepa) stated on the wallet (primary packaging) of the Protexis Neoprene surgical gloves are incorrect. Additionally, breakthrough times for Mitomycin were added to the same label in error.
- 2018-12-26FDA-DeviceCardinal Health 200, LLCClass IIProtexis Neoprene Surgical Glove, Size 7.5 This powder-free sterile light brown colored surgeons gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Labeling issue in that the breakthrough (i.e., permeation) times for two chemotherapy drugs (Carmustine and Thiotepa) stated on the wallet (primary packaging) of the Protexis Neoprene surgical gloves are incorrect. Additionally, breakthrough times for Mitomycin were added to the same label in error.