Medical device recalls
39226 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2018-12-05FDA-DeviceEUROTROL INCClass IIHemoTrol Level 2- In Vitro Diagnostic Hematology quality control mixture Catalogue number: 171.002.002 Eurotrol HemoLin is an assayed hemoglobin control intended for professional use in the verification of the linearity of the HemoCue¿ Hb 201 systems
Incorrect measurement results caused by microbial contamination of the product, results in values lower than assigned values
- 2018-12-05FDA-DevicePerkinElmer Health Sciences Canada, Inc.Class IIPerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds in human samples by ionizing the compound and separating the resulting ions. Part Number: BC004872
Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled.
- 2018-12-05FDA-DeviceSmiths Medical ASD Inc.Class IIMedex High Pressure Stopcock used to control the direction of IV fluid flow. This device is intended for single use only and is provided sterile and fluid path non-pyrogenic unless package is damaged or opened.
Certain lots of Medex" High Pressure Stopcocks were manufactured with an inner body defect that may result in a leak. Model Number - Lot Number MX4301L - 3538569 MX4331L - 3562377 & 3546826 MX4331R - 3546924
- 2018-12-05FDA-DeviceNumed IncClass IIMulti-Track Angiographic Catheter (MMTA) Product Code: MMTA06100 Product Usage: Recommended for use in catheterization for angiography of cardiovascular vessels and I or chambers. It can be used for injection of contrast medium and pressure measurement in any chamber or vessel.
Instructions for Use booklets were not included on the outer pouch
- 2018-12-05FDA-DeviceNumed IncClass IIMulti-Track Angiographic Catheter (MMTA) Product Code: MMTA05100 Product Usage: Recommended for use in catheterization for angiography of cardiovascular vessels and I or chambers. It can be used for injection of contrast medium and pressure measurement in any chamber or vessel.
Instructions for Use booklets were not included on the outer pouch
- 2018-12-05FDA-DeviceTosoh Bioscience IncClass IIAIA-900 Automated Immunoassay Analyzer, product code 022930 /022930R Product Usage: The AIA-900 is capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step immunoenzymetric immunoassay. An antigen-antibody reaction begins by combining a patient sample, control, or calibrator with a diluent in an immunoreaction test cup from the AIA-PACK reagent series.
Tosoh Bioscience has become aware of a potential issue with the axis base of the turntable rotation drive motor on the AIA-900 Analyzer. If the tension of the timing belt between the turntable rotation pulley and drive motor is not adjusted correctly, it is possible to increase the lateral load on the motor axis resulting in the inclination and damage of the motor axis. Since January 12, 2017, Tosoh has been inspecting the tension of the timing belt during scheduled preventative maintenance visit and correcting the analyzers where this issue has been observed. If the drive motor fails, the AIA-900 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received five (5) complaints related to this issue with no serious injuries reported.
- 2018-12-05FDA-DeviceWOM World of Medicine AGClass IIPNEUMOCLEAR Heated High-Flow Tube Set Product Usage: The Heated High-Flow Tube Set is a single use sterile tube set used in combination with the PNEUMOCLEAR C02 insufflator as a separate accessory. The device PNEUMOCLEAR is a C02 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The Standard, High Flow/Bariatric, Pediatric and Advanced Flow operating mode of the device are indicated to fill and distend a peri
It is possible that the packaging of the product can be damaged by the prongs on the tube set.
- 2018-12-05FDA-DeviceNxStage Medical, Inc.Class IINxStage PureFlow B Solution-RFP-RFP-456, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the eyes or skin
- 2018-12-05FDA-DeviceNxStage Medical, Inc.Class IINxStage PureFlow B Solution-RFP-RFP-454, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the eyes or skin
- 2018-12-05FDA-DeviceNxStage Medical, Inc.Class IINxStage PureFlow B Solution-RFP-406, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the eyes or skin
- 2018-12-05FDA-DeviceNxStage Medical, Inc.Class IINxStage PureFlow B Solution-RFP-404, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the eyes or skin
- 2018-12-05FDA-DeviceNxStage Medical, Inc.Class IINxStage PureFlow B Solution-RFP-403, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the eyes or skin
- 2018-12-05FDA-DeviceNxStage Medical, Inc.Class IINxStage PureFlow B Solution-RFP-402, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the eyes or skin
- 2018-12-05FDA-DeviceNxStage Medical, Inc.Class IINxStage PureFlow B Solution-RFP-401, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the eyes or skin
- 2018-12-05FDA-DeviceNxStage Medical, Inc.Class IINxStage PureFlow B Solution-RFP-400, Premixed Dialysate for Hemodialysis. NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the eyes or skin
- 2018-12-05FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIDimension Vista Lactate Dehydrogenase (LDI) Flex reagent cartridge, an in vitro diagnostic test for the quantitative measurement of lactate dehydrogenase in human serum and plasma on the Dimension Vista¿ System.Cat. No. K2054, Mat. No. 10464323
Siemens Healthcare Diagnostics has confirmed that Dimension Vista Lactate Dehydrogenase (LDI) may exhibit an erroneously elevated result when the preceding assay in the reaction cuvette is Lactic Acid (LA). The magnitude of the elevation may be variable. Both quality control and patient samples may be impacted. Incomplete removal of residual lactate dehydrogenase, a component of the LA reagent, has been identified as the cause of the potentially elevated result. A falsely elevated result for lactate dehydrogenase may lead to additional investigations to determine the etiology of tissue damage. Results would be correlated with clinical history and presentation in addition to other diagnostic laboratory testing.
- 2018-12-05FDA-DeviceStryker Sustainability SolutionsClass IIReprocessed Xcel Bladeless Trocar (H12LP) Model # H12LP Reprocessed Endoscopic Trocars12x100mm - Product Usage: Physical Product Description: The Reprocessed Blunt Tip Trocar consists of a radiolucent cannula and obturator sized 12mm in diameter (refer to Figure 1). The Reprocessed Blunt Tip Trocar has an obturator that has a blunt plastic tip that gently moves aside any internal viscera that may be adjacent to the abdominal or thoracic wall. The trocar cannula consists of two seals, an outer in
Through investigation, it was determined that H12LP Trocars in scope of a previous H12LP recall were distributed to facilities within the distribution chain following the recall execution activities initially performed by distributor.
- 2018-12-05FDA-DeviceSynaptive Medical IncClass IBrightMatter Guide, Product Number SYN-0026, Versions: 1.5.0, 1.5.1 and 1.5.2 Product Usage: BrightMatter Guide can navigate a third-party port (NICO BrainPath device) by attaching Synaptive s custom Tracking Array to the port directly. As the BrainPath is navigable in two configurations (sheath only / sheath and obturator), the tip of the obturator is represented in the software with a tip extension beyond the bottom of the sheath. Note sheath and cannula are used interchangeably in this do
This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure.
- 2018-12-05FDA-DeviceMaquet Cardiovascular Us Sales, LlcClass IQUADROX-i Neonatal with integrated arterial filter and VHK 11000 (VKMO 11000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055207 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.
There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.
- 2018-12-05FDA-DeviceMaquet Cardiovascular Us Sales, LlcClass IQUADROX-i Neonatal without integrated arterial filter and VHK 11000 (VKMO 10000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055120 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.
There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.
- 2018-11-28FDA-DeviceZimmer Biomet, Inc.Class IIVanguard Knee System -PS Open Box Femoral Right, 65 mm Item Number:183108
Label on the outer carton or the patient labels does not match label on the inner sterile packaging
- 2018-11-28FDA-DeviceZimmer Biomet, Inc.Class IIVanguard Knee System -PS Open Box Femoral Left, 62.5 mm Item Number: 183126
Label on the outer carton or the patient labels does not match label on the inner sterile packaging
- 2018-11-28FDA-DeviceBecton Dickinson & CompanyClass IIKIT URIN CUP 16X100 8.0 CONI UA YEL, Cat. No. 364989 BD Vacutainer Urine Cup Kit with UA Tube for Midstream Specimens
BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
- 2018-11-28FDA-DeviceBecton Dickinson & CompanyClass IIKIT URIN CUP PLC 16X100 10.0 UA YEL, Cat. No. 364981: Urine Collection Kit for collection and transport of urine specimens for urinalysis
BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
- 2018-11-28FDA-DeviceBecton Dickinson & CompanyClass IIBD Vacutainer Specimen Collection Cups, Cat. No. 364975
BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
- 2018-11-28FDA-DeviceBecton Dickinson & CompanyClass IIBD Vacutainer Urine Collection Cups, Cat. No. 364941
BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
- 2018-11-28FDA-DeviceMedex Cardio-Pulmonary Inc., d.b.a. Smiths Medical CompanyClass IIIrrisept OR, Finished Bottle Assembly 450 ml STEP 1
Lack of sterility assurance: leaking containers which could lead to exposure to infectious agents.
- 2018-11-28FDA-DeviceMerck Millipore Ltd.Class IIMinicon Clinical Sample Concentrators for In Vitro Diagnostic Use 9031 (Model: Minicon B-15) & 9051 (Model: Minicon CS-15) Minicon concentrators are non-sterile, disposable, multiwall ultrafiltration devices. Minicon¿ clinical sample concentrators are for in vitro diagnostic use and are intended for concentrating serum, urine, cerebrospinal fluid, and other body fluids prior to analysis.
In vitro diagnostic devices, used greater than 6 months after date of manufacture, may have increased concentration times of greater than the specification of 3 hours and up to 24 hours, which may lead to delays in carrying out the subsequent analysis tests.
- 2018-11-28FDA-DeviceSedecal S.A.Class IISedecal NOVA FA DR System. for diagnostic radiography.
A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient, user, or third parties.
- 2018-11-28FDA-DeviceSVS LLCClass IIPremierPro Aluminum Adjustable Walker
This recall is due to a mandated order from the FDA for stock recovery and return of the imported item.