- 2018-11-21FDA-DeviceAROA Biosurgery, LTD.Class IIOviTex 1S Reinforced BioScaffold 10x12cm, Part Number F10256-1012G
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
- 2018-11-21FDA-DeviceAROA Biosurgery, LTD.Class IIOviTex 1S Reinforced BioScaffold 6x10cm, Part Number F10256-0610G
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
- 2018-11-21FDA-DeviceAROA Biosurgery, LTD.Class IIOviTex 1S Reinforced BioScaffold 4x8cm, Part Number F10256-0408G
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
- 2018-11-21FDA-DeviceAROA Biosurgery, LTD.Class IIOviTex Reinforced BioScaffold 20x20cm, Part Number F10254-2020G
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
- 2018-11-21FDA-DeviceAROA Biosurgery, LTD.Class IIOviTex Reinforced BioScaffold 16x20cm, Part Number F10254-1620G
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
- 2018-11-21FDA-DeviceAROA Biosurgery, LTD.Class IIOviTex Reinforced BioScaffold 10x20cm, Part Number F10254-1020G
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
- 2018-11-21FDA-DeviceAROA Biosurgery, LTD.Class IIOviTex Reinforced BioScaffold 10x12cm, Part Number F10254-1012G
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
- 2018-11-21FDA-DeviceAROA Biosurgery, LTD.Class IIOviTex Reinforced BioScaffold 6x10cm, Part Number F10254-0610G
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
- 2018-11-21FDA-DeviceAROA Biosurgery, LTD.Class IIOviTex Reinforced BioScaffold 4x8cm, Part Number F10254-0408G
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
- 2018-11-21FDA-DeviceCollagen Matrix, Inc.Class IIIBioMend Extend, REF# 0142Z, Synthetic bone grafting material.
There is a possibility that the product was packaged in the wrong box. Specifically, the product may have been placed inside a BioMend box instead of a BioMend Extend box. The only difference between the two boxes is the product trade name. The label placed on the box and on the product packaged inside the box is correct. The Instructions for Use inside the box are correct.
- 2018-11-21FDA-DeviceGE Healthcare, LLCClass IITouchscreen display kit. Sold under the following product names: TOUCHSCREEN DISPLAY KIT 11.X ENGLISH FOR USA, TOUCHSCREEN DISPLAY KIT 11.X ENGLISH, TOUCHSCREEN DISPLAY KIT 11.X GERMAN, TOUCHSCREEN DISPLAY KIT 11.X DUTCH, TOUCHSCREEN DISPLAY KIT 11.X FRENCH, TOUCHSCREEN DISPLAY KIT 11.X DANISH, TOUCHSCREEN DISPLAY KIT 11.X ESTONIAN, TOUCHSCREEN DISPLAY KIT 11.X ITALIAN, TOUCHSCREEN DISPLAY KIT 11.X NORWEGIAN. Product Usage: The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide
Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.
- 2018-11-21FDA-DeviceGE Healthcare, LLCClass IIAisys CS2 Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH FOR USA, AISYS 10.X TO 11.X UPGRADE KIT FRENCH, AISYS 10.X TO 11.X UPGRADE KIT GERMAN, AISYS 10.X TO 11.X UPGRADE KIT JAPANESE. Product Usage: The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal,
Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.
- 2018-11-21FDA-DeviceGE Healthcare, LLCClass IIAisys Anesthesia System. Sold under the following product names: AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH FOR USA, AISYS 10.X TO 11.X UPGRADE KIT FRENCH, AISYS 10.X TO 11.X UPGRADE KIT GERMAN, AISYS 10.X TO 11.X UPGRADE KIT JAPANESE.
Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.
- 2018-11-21FDA-DeviceGE Healthcare, LLCClass IGE Healthcare Millennium MC
A detector can detach and fall.
- 2018-11-21FDA-DeviceGE Healthcare, LLCClass IGE Healthcare Millennium MG Multi-Geometry Digital CSE Dual Detector Gamma Camera H3000ZL Acquisition System
A detector can detach and fall.
- 2018-11-21FDA-DeviceGE Healthcare, LLCClass IGE Healthcare Millennium MyoSIGHT Nuclear Cardiology Imaging System H3000ZH Acquisition System
A detector can detach and fall.
- 2018-11-14FDA-DeviceR & D Systems, Inc.Class IIIR&D systems, a bio-techne brand Human Total 25-OH Vitamin D IVD ELISA Kit, REF RDKAP1971
The kit was provided without a copy of the lot-specific kit booklet and the Calibrator 0 vial 18D1/4 label incorrectly specified reconstitution volume of 2mL.
- 2018-11-14FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIDimension Thermal Chamber Motor Cable located in the following models: Model/UDI Number: Dimension¿ EXL 200 00630414593579; Dimension¿ EXL with LM 00630414593500; Dimension¿ EXL with LM with STM 00630414593593; Dimension¿ RxL Max 00630414949789; Dimension¿ RxL Max HM 00630414949796; Dimension¿ RxL Max HM STM 00630414949819; Dimension¿ Xpand¿ Plus 00630414949833; Dimension¿ Xpand¿ Plus HM 00630414949840; Dimension¿ RxL Max Refurbished 00630414592947; Dimension¿ RxL Max HM Refurbished 0
Thermal chamber motors have electrical cables with single insulation rather than the required double insulation. There is no exposed current due to the single insulation that could result in a potential electrical shock
- 2018-11-14FDA-DeviceAccuray IncorporatedClass IICyberKnife M6, Part Number 054000-004 The CyberKnife M6 Systems are indicated for treatment planning and image guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
A robotics supplier notified Accuray of two manufacturing variations in the fastening of the in-line wrist and the casting of the wrist for the robot. The variation may result in premature failure of the component.
- 2018-11-14FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIMedtronic CareLink Encore 29901 Programmer
Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.
- 2018-11-14FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIMedtronic CareLink 2090 Programmer
Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.
- 2018-11-14FDA-DeviceIntegrity Implants Inc.Class IIFlareHawk Interbody Fusion System (previously XCage Interbody Fusion System) Instructions for Use (IFU's) only. Product Usage: Indicated for spinal intervertebral body fusion with autogenous bone graft in skeletally mature individuals with degenerative disc disease.
Recently Integrity Implants uncovered a discrepancy in the cleaning parameters in several of their labeling material. The firm is initiating the field correction to ensure that updated information is provided to all accounts.
- 2018-11-14FDA-DeviceMedtronic Navigation, Inc.-LittletonClass IIMedtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS IMAGING 02 REFURB Model: BI70002000R
O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuit
- 2018-11-14FDA-DeviceDT MedTech, LLCClass IIDT Medi Tech Hintermann Series Talar Implant LEFT-SIZE 4 Part Number: 302114 components are prostheses for a mobile bearing ankle joint.
Incorrect package labeling identifying the device
- 2018-11-14FDA-DeviceDT MedTech, LLCClass IIDT Medical -Hintermann Series H3 Poly Inlay, Size 3 5mm Part Number: 300305 components are prostheses for a mobile bearing ankle joint.
Incorrect package labeling identifying the device
- 2018-11-14FDA-DeviceSafe N Simple, LLCClass IISimpurity HYDROGEL ABSORBENT SHEET WOUND DRESSING ADHESIVE BORDER, 4" X 5", 12 Pads, Catalogue number SNS58345
The sterilization process used was unqualified.
- 2018-11-14FDA-DeviceSafe N Simple, LLCClass IISimpurity HYDROGEL ABSORBENT SHEET WOUND DRESSING, 4" X 5", 10 Pads, Catalogue number SNS58315
The sterilization process used was unqualified.
- 2018-11-14FDA-DeviceZimmer Biomet, Inc.Class IIStageOne Hip, Models 431207 & 431209Usage: Product Usage: The single-use cement spacer molds are sterile disposables made of medical grade silicone with a 316L stainless steel reinforcement stem. They are designed to be filled with polymethylmethacrylate/gentamicin bone cement, or equivalent, by injecting with a dispenser/gun into the mold. After the cement cures, the temporary spacers are to be removed from the molds with the reinforcement remaining as the core of the spacer, and placed i
There is a potential for comingling of the 43MM and 51 MM molds
- 2018-11-14FDA-DeviceZimmer Biomet, Inc.Class IIBiomet Microfixation RapidFlap System 20MM RapidFlap SpinDown Clamp, 12 Pack Ref Number: 75-1040-12 Used for the re-attachment of the bone flap after a craniotomy.
Spin Down RapidFlap nut that interfaces with the outer plate was assembled incorrectly.
- 2018-11-14FDA-DeviceZimmer Biomet, Inc.Class IIBiomet Microfixation RapidFlap System 20MM RapidFlap SpinDown Clamp Ref Number: 75-1040 Used for the re-attachment of the bone flap after a craniotomy.
Spin Down RapidFlap nut that interfaces with the outer plate was assembled incorrectly.