Medical device recalls
39226 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2018-08-29FDA-DeviceElekta, Inc.Class IIELEKTA Digital Accelerator under the following brand names: Elekta Synergy, Synergy S, Synergy Platform, Precise Digital Accelerator, Infinity, Axesse, and Versa HD
There is a potential for an uncontrolled extension of iViewGT / XVI detector arms.
- 2018-08-29FDA-DeviceHitachi Medical Systems America IncClass IIHitachi Scenaria CT system Product Usage: The SCENARIA CT system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the SCENARIA can be post processed by the SCENARIA to provide additional information. Post processing capabilities included in the SCENARIA software include CT angiography (CTA), Multiplanar reconstruct
The firm received a complaint where, during patient positioning, the technologist was adjusting the patient table height and the table failed to stop when the control was released and contacted the CT gantry. The problem could not be duplicated at the site, so to avoid further issues while a root cause investigation was pursued, Hitachi choose to add an optional touch sensor to the gantry to add a redundant level of protection. It is conceivable that a patient s extremity may be caught between the tabletop and the gantry surface during uncontrolled motion or that the table contact of the gantry might cause internal damage to the unit.
- 2018-08-29FDA-DeviceSiemens Medical Solutions USA, IncClass II3D V, model. no. 7042042 7042059 3070021 Product Usage: The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.
In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.
- 2018-08-29FDA-DeviceSiemens Medical Solutions USA, IncClass II3D TOP Ceiling Stand, model no. 3070039 3074486 3070013 3070021 Product Usage: The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.
In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.
- 2018-08-29FDA-DeviceLIEBEL-FLARSHEIM COMPANY LLCClass IILiebel-Flarsheim Direct Digital Imaging System (DDIS) Liebel-Flarsheim urology systems facilitate radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation, primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/ surgery.
Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.
- 2018-08-29FDA-DeviceLIEBEL-FLARSHEIM COMPANY LLCClass IILiebel-Flarsheim Hydra Vision Urology X-Ray System (DR) Liebel-Flarsheim urology systems facilitate radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation, primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/ surgery.
Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.
- 2018-08-29FDA-DeviceVyaire MedicalClass IThe Tri-Flo Subglottic Suction System
Vyaire Medical has discovered potential patient safety risk with associated with the Tri-Flow Subglottic Suction System. The distal soft tip of the catheter has been identified as having the potential to break off during patient use and result in aspiration of a solid foreign body.
- 2018-08-29FDA-DevicebioMerieux, Inc.Class IbioMerieux VITEK 2 Gram Positive Cefoxitin Screen, VITEK 2 Gram Positive AST for Oxacillin, VITEK SYSTEMS General Susceptibility Card Cefotoxin and AST for Oxacillin bioMerieux VITEK 2 AST-P631 bioMerieux VITEK 2 AST-GP67 Test Kit bioMerieux VITEK 2 AST-GP71 Test Kit bioMerieux VITEK 2 AST-GP75 Test Kit bioMerieux VITEK 2 AST-GP78 Test Kit Components for an automated invitro diagnostic system consisting of instruments, software and reagent cards designed for the ID and antimicro
Customer reports indicated an increase in the rate of non-detected MRSA in association with the VITEK 2 AST-P631 product.
- 2018-08-29FDA-DeviceHandicare Usa IncClass IIHandicare P600 Series Patient Lift Product Usage: A device used to lift and transport a patient from one place to another.
Premature strap wear and breakage at maximum weight conditions (625 lbs.).
- 2018-08-29FDA-DeviceHandicare Usa IncClass IIHandicare C Series Patient Lift Product Usage: A device used to lift and transport a patient from one place to another.
Premature strap wear and breakage at maximum weight conditions (625 lbs.).
- 2018-08-29FDA-DeviceCompass Health Brands (Corporate Office)Class ICPAP Mask Cushion, Model Nos. PB781S, PB781M, PB781L, 781S, 781M and 781L Product Usage: A CPAP mask is attached to an end user s face and a CPAP machine, to treat end users with sleep apnea. The replacement cushion is a non-durable component that requires replacement every 3 - 6 months. The cushion is used for both comfort and to provide a seal between the face and the mask.
A May 2015 design change in the cushion seal replacement part and accompanying elbow replacement part resulted in incompatibility of the new cushion seal with the previous design of the elbow. Use of the new cushion seal with the previous design elbow could lead to unacceptable CO2 porting. As there was no model number change for this design change, customers may not be aware of the incompatibility.
- 2018-08-22FDA-DeviceDRG Instruments GmbHClass IISalivary Estradiol ELISA, REF SLV-4188 in vitro diagnostic quantitative measurement of active free Estradiol, an estrogenic steroid, in saliva.
Lower recovery of salivary estradiol levels due to poor differentiation between kit Standard 0 and Standard 1.
- 2018-08-22FDA-DeviceIntuitive Surgical, Inc.Class IIThe da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
- 2018-08-22FDA-DeviceBeekley CorporationClass IITomoSPOT REF 782, 782S, 783, 783S, 784, 784S, 785 & 785S Adhere to skin for marking and identifying moles in mammography.
There is a potential for imaging artifact to mimic calcifications on patient's mammograms
- 2018-08-22FDA-DeviceBeekley CorporationClass IIS-SPOT REF 777, 777S Adhere to skin for marking and identifying moles in mammography.
There is a potential for imaging artifact to mimic calcifications on patient's mammograms
- 2018-08-22FDA-DeviceBeekley CorporationClass IIO-SPOT REF 791, 791S Adhere to skin for marking and identifying moles in mammography.
There is a potential for imaging artifact to mimic calcifications on patient's mammograms
- 2018-08-22FDA-DeviceWithings SasClass IINokia BPM+ Wireless Blood Pressure Monitor Product System, measurement, blood-pressure, non-invasive
Device could not meet the requirements for systolic pressure
- 2018-08-22FDA-DeviceCypress Medical Products LLCClass IIPowder Free Vinyl Exam Gloves, Item Nos. 14-116, 14-118 Product Usage: Medical exam glove for use by professional healthcare workers and/or consumers.
The firm is recalling one lot of McKesson Powder Free Vinyl Exam Gloves because a sample of product from that lot did not meet certain release specifications. It is possible that certain gloves from this lot may not be of medical exam quality and, therefore, may not provide an effective barrier to infectious material.
- 2018-08-22FDA-DeviceTosoh Bioscience IncClass IIITosoh Bioscience AIA Analyte Application Manual (AAM) CD Part Number: 997027 CD contains analyte application manuals for 47 analyte reagent packs for use with the Tosoh AIA (Automated Immuno-Assay) family of analyzers. Each analyte application manual includes a reference section on assay controls in the analyte application manual.
Distribution of unapproved product labeling
- 2018-08-22FDA-DeviceSorin CRM SASClass IISorin Platinium CRT-D 1741 DF-4, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator
Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrillation shock therapy.
- 2018-08-22FDA-DeviceSorin CRM SASClass IISorin Platinium CRT-D 1711 DF-1, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator
Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrillation shock therapy.
- 2018-08-22FDA-DeviceMerit Medical Systems, Inc.Class IIFlex-Neck Classic Peritoneal Dialysis Catheter (Pediatric) Product Usage: If the patient is a suitable candidate for peritoneal dialysis (PD) therapy, The Flex-Neck¿ PD Catheter can be implanted either surgically or peritoneoscopically. The only contraindication to implantation of the FlexNeck PD Catheter is if the patient is not a candidate for peritoneal dialysis. Numerous prior surgeries or suspected or documented intraperitoneal adhesions may be relative contraindications to PD. However,
Mix-up between adult and pediatrics PD catheter.
- 2018-08-22FDA-DeviceNico Corp.Class IINICO Shepherd' s Hook - Greenberg, sterile. An accessory to NICO s BrainPath device Part Number: NN-8024
Mislabeled hook pouch included a transport/storage label rather than the product label containing key product information such as part number, lot number, and expiration date, etc
- 2018-08-22FDA-DeviceZimmer Biomet, Inc.Class IIECHO Bi-Metric Hip System, Reduced Proximal Profile, Standard 135o neck, item number 192414. orthopedic hip prosthesis femoral stem
Two lots of the Echo BI-Metric Hip Stem and ARCOS Modular Revision Hip Stem may have been comingled. Potential health consequences include extension of surgery time to find a replacement component.
- 2018-08-22FDA-DeviceZimmer Biomet, Inc.Class IIARCOS Modular Revision Hip System, item number 11-300814 orthopedic hip prosthesis femoral stem.
Two lots of the Echo BI-Metric Hip Stem and ARCOS Modular Revision Hip Stem may have been comingled. Potential health consequences include extension of surgery time to find a replacement component.
- 2018-08-22FDA-DeviceArrow International IncClass IIHANDS-OFF Multi-Lumen Central Venous Catheter with Blue FlexTip for use only with Arrow One-Piece Percutaneous Sheath Introducer System, Material HO-14703 The central venous catheter permits venous access to the central circulation through an indwelling Arrow percutaneous sheath introducer.
The packaging may not be sealed. If the packaging is compromised in this manner, the sterility of the product cannot be guaranteed.
- 2018-08-22FDA-DeviceArrow International IncClass IIHANDS-OFF Multi-Lumen Central Venous Catheter with Blue FlexTip for use only with Arrow One-Piece Percutaneous Sheath Introducer System, Material HO-14702 The central venous catheter permits venous access to the central circulation through an indwelling Arrow percutaneous sheath introducer.
The packaging may not be sealed. If the packaging is compromised in this manner, the sterility of the product cannot be guaranteed.
- 2018-08-22FDA-DeviceICU Medical de Mexico, S.A. de C.V.Class II6" Smallbore Ext Set w/6-Port NanoClave Manifold, Check Valve, Clamp, Rotating Luer.
ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.
- 2018-08-22FDA-DeviceICU Medical de Mexico, S.A. de C.V.Class II10" Smallbore Ext Set w/6-Port NanoClave Manifold (Orange, Red, Blue, Purple, Yellow, Green Rings), NanoClave (Light Blue Ring), Check Valve, Clamp, Rotating Luer.
ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.
- 2018-08-22FDA-DeviceICU Medical de Mexico, S.A. de C.V.Class II11" Smallbore Ext Set w/NanoClave, 6-Port NanoClave Manifold, Check Valve, Clamp, Rotating Luer.
ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.