Medical device recalls
39226 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2018-08-01FDA-DevicePhilips HealthcareClass IIUNIQ Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA,stent placing,atherectomies), pacemaker implantat ions, and electrophysiology (EP). Non-vascular interventions such as drainages,biopsies and vertebroplasties procedures.
The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.
- 2018-08-01FDA-DevicePhilips HealthcareClass IIAllura Xper Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA,stent placing,atherectomies), pacemaker implantat ions, and electrophysiology (EP). Non-vascular interventions such as drainages,biopsies and vertebroplasties procedures.
The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IITri-Staple 2.0 Curved Tip Intelligent Reload for use with Signia stapling system , Item Code EGIA30CTAV
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IITri-Staple 2.0 Curved Tip Intelligent Reload for use with Signia stapling system 45 mm Vascular/Medium, Item Code SIG45ACTAVM
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IITri-Staple 2.0 Curved Tip Intelligent Reload for use with Signia stapling system 30 mm Vascular/Medium, Item Code SIG30ACTAVM
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IITri-Staple 2.0 Intelligent Reload for use with Signia stapling system 30 mm Medium/Thick, Item Code SIG30AMT
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IIEndo GIA 45mm Curved Tip Articulating Vascular/Medium Reload with Tri-Staple Technology, Item Code EGIA45CTAVM
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IIEndo GIA 45 mm Articulating Vascular/Medium Loading Unit with Tri-Staple Technology, Item Code EGIA454AVM
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IIEndo GIA 45 mm Extra Thick Black Articulating Reload with Tri-Staple Technology, Item Code EGIA45AXT
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IIEndo GIA 45mm Gray Articulating Vascular Reload with Tri-Staple Technology, Item Code EGIA45AV
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IITri-Staple 2.0 Intelligent Reload for use with Signia stapling system 30 mm Vascular/Medium, Item Code SIG30AVM
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IITri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code SIG60AXT
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IIEndo GIA 60 mm Articulating Vascular/Medium Loading Unit with Tri-Staple Technology, Item Code EGIA60AVM
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IIEndo GIA 60 mm Extra Thick Black Articulating Reload with Tri-Staple Technology, Item Code EGIA60AXT
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IIEndo GIA 60 mm Articulating Medium/Thick Reload with Tri-StapleTechnology, Item Code EGIA60AMT
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceCOVIDIEN MEDTRONICClass IIEndo GIA 45 mm Articulating Medium/Thick Reload with Tri-Staple Technology, Item Code EGIA45AMT
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
- 2018-07-25FDA-DeviceMaquet SASClass IIVOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP; ARD568821961 VLT600DF AIM STP QL; ARD568811911 VLT600SF AIM STP; ARD568821911 VLT600SF AIM STP QL; ARD568821910 VLT600SF STP QL; ARD568811901 VLT600SF AIM STP Product Usage: The Maquet VOLISTA surgical lights are designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. StandOP VOLISTA is a system of ceiling-mounted surgical lights suited for installation in surgical suites,
Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.
- 2018-07-25FDA-DeviceMedtronic Navigation, Inc.Class IIVisualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase¿ Cooled Laser Applicator System 3mm Tip; b. Part Number 9735560 Visualase¿ Cooled Laser Applicator System 10mm Tip; c. Part Number 9735561 Visualase¿ Cooled Laser Applicator System 15mm Tip Product Usage: The Visualase Thermal Therapy System is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy under magnetic resonance imaging (MRI
Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.
- 2018-07-25FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation stand-alone software treatment planning system, RayStation 4.5, RayStation 4.7, RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2), RayStation 7 (RayPlan 7) and RayStation 8A (RayPlan 8A) Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer
If the beam model has a highly asymmetric primary source, it is not correctly taken into account in the calculation of DMLC fields when the collimator is rotated. This could lead to potentially significant overdosage at delivery. The user must be aware of the issue to avoid incorrect dose calculations during treatment planning.
- 2018-07-25FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIIngenuity Digital PET/CT System, Model 882446 with the following UDI's: (01)00884838059559(21)900004 (01)00884838059559(21)900005 (01)00884838059559(21)900006 (01)00884838059559(21)900007 (01)00884838059559(21)900008 (01)00884838059559(21)900009 (01)00884838059559(21)900010 (01)00884838059559(21)900011 (01)00884838059559(21)900012 (01)00884838059559(21)900013 (01)00884838059559(21)900014 (01)00884838059559(21)900015 (01)00884838059559(21)900016 (01)00884838059559(21)900017 (01)00
The block used to connect the patient pallet to the couch may be assembled incorrectly. A screw may be misaligned in the patient pallet.
- 2018-07-25FDA-DeviceCardinal Health 200, LLCClass IIHydroline Trumpet Valve, Trumpet Valve, No Probe, Double Spike, Ref ASU1221, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
The suction valve may not close properly which could cause continuous suction.
- 2018-07-25FDA-DeviceCardinal Health 200, LLCClass IIHydroline Trumpet Valve, Trumpet Valve, No Probe, Single Spike, Ref ASU1220, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
The suction valve may not close properly which could cause continuous suction.
- 2018-07-25FDA-DeviceCardinal Health 200, LLCClass IIHydroline Trumpet Valve, Trumpet Valve, 5mm x 28cm Probe, Double Spike, Ref ASU1211, packaged 1/box, 12 boxes/case, single use, RX, Sterile. Manufactured For Cardinal Health, Waukegan, IL.
The suction valve may not close properly which could cause continuous suction.
- 2018-07-25FDA-DeviceCardinal Health 200, LLCClass IIHydroline Trumpet Valve, Trumpet Valve, 5mm x 28cm Probe, Single Spike, Ref ASU1210, packaged 1/box, 12 boxes/case, single use, RX, Sterile. .
The suction valve may not close properly which could cause continuous suction.
- 2018-07-25FDA-DeviceCardinal Health 200, LLCClass IIHydroline Trumpet Valve, Trumpet Valve, 5mm x 33cm Probe, Double Spike, Ref ASU1201, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
The suction valve may not close properly which could cause continuous suction.
- 2018-07-25FDA-DeviceCardinal Health 200, LLCClass IIHydroline Trumpet Valve, Trumpet Valve, 5mm x 33cm Probe, Single Spike, Ref ASU1200, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
The suction valve may not close properly which could cause continuous suction.
- 2018-07-25FDA-DeviceCardinal Health 200, LLCClass IIHydroline Trumpet Valve with Pulse Wave Cassette, Trumpet Valve with Double Spike and no Probe, Ref ASC1221, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
The suction valve may not close properly which could cause continuous suction.
- 2018-07-25FDA-DeviceCardinal Health 200, LLCClass IIHydroline Trumpet Valve with Pulse Wave Cassette, Trumpet Valve with Single Spike and no Probe, Ref ASC1220, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
The suction valve may not close properly which could cause continuous suction.
- 2018-07-25FDA-DeviceCardinal Health 200, LLCClass IIHydroline Trumpet Valve with Pulse Wave Cassette, Trumpet Valve with Cassette, 5mm x 33cm Composite Probe, Double Spike, Ref ASC1201, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
The suction valve may not close properly which could cause continuous suction.
- 2018-07-25FDA-DeviceCardinal Health 200, LLCClass IIHydroline Trumpet Valve with Pulse Wave Cassette, Trumpet Valve with Cassette, 5mm x 33cm Composite Probe, Single Spike, Ref ASC1200, packaged 1/box, 12 boxes/case, single use, RX, Sterile.
The suction valve may not close properly which could cause continuous suction.