Medical device recalls
39226 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2018-07-18FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIMedtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D154VRC Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.
Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
- 2018-07-18FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIMedtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153VRC Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.
Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
- 2018-07-18FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIMedtronic EnTrust Dual Chamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D154ATG Product Usage: The device indicated for use in ICD patients with atrial tachyarrhythmias, or who are at significant risk of developing atrial tachyarrhythmias. The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhyth
Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
- 2018-07-18FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIMedtronic EnTrust Dual Chamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153ATG Product Usage: The device indicated for use in ICD patients with atrial tachyarrhythmias, or who are at significant risk of developing atrial tachyarrhythmias. The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhyth
Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
- 2018-07-18FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIPercepta Quad CRT-P MRI SureScan, REF W4TR04 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and regulate the patient s heart rate by providing single or dual chamber rate responsive bradycardia pacing, sequential biventricular pacing, and atrial tachyarrhythmia therapies. These devices are indicated for heart failure patients who may have atrial arrhythmias or AV block.
Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
- 2018-07-18FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIPercepta CRT-P MRI SureScan, REF W1TR04 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and regulate the patient s heart rate by providing single or dual chamber rate responsive bradycardia pacing, sequential biventricular pacing, and atrial tachyarrhythmia therapies. These devices are indicated for heart failure patients who may have atrial arrhythmias or AV block.
Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
- 2018-07-18FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIPercepta Quad CRT-P MRI SureScan, REF W4TR01 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and regulate the patient s heart rate by providing single or dual chamber rate responsive bradycardia pacing, sequential biventricular pacing, and atrial tachyarrhythmia therapies. These devices are indicated for heart failure patients who may have atrial arrhythmias or AV block.
Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
- 2018-07-18FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIPercepta CRT-P MRI SureScan, REF W1TR01 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and regulate the patient s heart rate by providing single or dual chamber rate responsive bradycardia pacing, sequential biventricular pacing, and atrial tachyarrhythmia therapies. These devices are indicated for heart failure patients who may have atrial arrhythmias or AV block.
Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
- 2018-07-18FDA-DeviceFort Defiance Industries, LLCClass IIFDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641 Product Usage: The Fort Defiance Industries Automated Field Steam Sterilizer Model P2131 is designed for sterilization of porous and non-porous, heat- and moisture-stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131 sterilizer is a transportable device designed for use in a variety of austere environments.
During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.
- 2018-07-18FDA-DeviceOhio Medical CorporationClass IIVacuum Regulator, Push-T-Set Intermittent Suction Unit (PTS-ISU) Product Usage: The vacuum regulator is intended to be used in a medical facility as a means to evacuate media (i.e. fluids) from the body.
Final quality control testing was not completed before the devices were distributed..
- 2018-07-18FDA-DeviceImmuno-Mycologics, IncClass IICryptococcal Antigen Lateral Flow Assay (CrAg LFA), Model # CR2003 Product Usage: The CrAg Lateral Flow Assay is an immunochromatographic test system for the qualitative or semi-quantitative detection of the capsular polysaccharide antigens of Cryptococcus species complex (Cryptococcus neoformans and Cryptococcus gattii) in serum and cerebral spinal fluid (CSF). The CrAg Lateral Flow Assay is a prescription-use laboratory assay which can aid in the diagnosis of cryptococcosis.
Potential for LFA strips to improperly wick which may cause false negative cryptococcosis test results
- 2018-07-18FDA-DeviceStryker GmbHClass IIAnkle Arthrodesis Nail, right T2 Ankle ¿11x300mm, Product Number 18191130S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIAnkle Arthrodesis Nail, right T2 Ankle ¿10x300mm, Product Number 18191030S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, RIGHT T2 GTN ¿14x460 mm, Product Number 18511446S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, RIGHT T2 GTN ¿9x320 mm, Product Number 18510932S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, LEFT T2 GTN ¿14x300 mm, Product Number 18501430S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, LEFT T2 GTN ¿12x360 mm, Product Number 18501236S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, LEFT T2 GTN ¿10x360 mm, Product Number 18501036S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, A/R, R1500 T2 Femur ¿12x380 mm, Product Number 18281238S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, A/R, R1500 T2 Femur ¿12x360 mm, Product Number 18281236S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, A/R, R1500 T2 Femur ¿12x300 mm, Product Number 18281230S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, A/R, R1500 T2 Femur ¿10x320 mm, Product Number 18281032S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, A/R T2 Femur ¿15x240 mm, Product Number 18251524S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, A/R T2 Femur ¿12x360 mm, Product Number 18251236S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, A/R T2 Femur ¿10x340 mm, Product Number 18251034S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IIFemoral Nail, A/R T2 Femur ¿10x320 mm, Product Number 18251032S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceStryker GmbHClass IITibial Nail, Standard T2 Tibia ¿12x315 mm, Product Number 18221231S
The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
- 2018-07-18FDA-DeviceSteris CorporationClass IIReliance 1227 Cart and Utensil Washer/Disinfector, FW03101, FW03102, FW03S003 Product Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theatre shoes and other miscellaneous reusable items used in the care of patients.
The firm has become aware that the Reliance 1227 Cart and Utensil Washer/Disinfector s Chemical Low Level alarm, intended to prevent the user from initiating a cycle when a low chemical level situation occurs, does not operate as intended. Currently, if a low chemical level situation occurs, the alarm will only be generated at the unit s next power-up or when accessing service mode. The failure of the Reliance 1227 to identify that the chemicals used for cleaning are low or empty could result in bedpans and urinals, basins, case carts, beds, theater shoes and other miscellaneous reusable patient care items not being properly cleaned before disinfection or reuse.
- 2018-07-18FDA-DeviceEkla CorporationClass IINOVAPLUS(R) Pregnancy Cassette Device with Droppers, Urine hCG Pregnancy Cassette, Cat. No. VPT1-CASS Product Usage: pregnancy test
Incorrect hCG results
- 2018-07-18FDA-DeviceRoche Diagnostics CorporationClass IIElecsys Vitamin D total II, Vitamin D Test System, Material number 07464215160 Product Usage: This assay is intended for the quantitative determination of total 25-hydroxyvitamin D in human serum and plasma. This assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The electrochemiluminescence binding assay is intended for use on the immunoassay analyzer.
The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by laboratory tests and adapted based on the result, a dosage reduction of vitamin D supplementation cannot be excluded, and as a consequence, vitamin D deficiency may develop over time.